Arthrex® AR-1409LPRecalls & reported problems

Arthrex, Inc. · REAMER, LOW PROFILE · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867016767), received between Nov 21, 2017 and Feb 1, 2024 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Entrapment of Device1
  • Material Fragmentation1

Patient problems

  • Device Embedded In Tissue or Plaque2
  • Foreign Body In Patient1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 1, 2024Break; Break

    ON 01/11/2023, AN FDA MEDWATCH NOTIFICATION WAS RECEIVED VIA EMAIL. A FACILITY REPRESENTATIVE REPORTED THAT AN AR-1409LP LOW PROFILE REAMER TIP BROKE OFF. THIS OCCURRED DURING A RIGHT KNEE ARTHROSCOPY, ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION WITH PATELLA TENDON AUTOGRAFT AND MEDIAL MENISCAL REPAIR WHERE ALL BROKEN FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED WITHOUT FURTHER ISSUES.

    Patient problems: Insufficient Information

    MAUDE report 18624784 ↗
  • MalfunctionJan 11, 2024

    SURGICAL ATTENDING INFORMED CIRCULATING RN THAT THE ARTHREX REAMER DRILL BIT TIP BROKE OFF - PIECES WERE COLLECTED AND NO HARM TO THE PATIENT. THE BROKEN ITEM WAS AN ARTHREX LOW PROFILE REAMER 9MM. ITEM # AR-1409LP, LOT # 15125284, AND EXPIRATION DATE OF 07/31/2028.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18491051 ↗
  • MalfunctionApr 18, 2022Break

    ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-1409LP REAMER, LOW PROFILE BROKE WHEN REAMING THE TIBIA. HARD BONE CAUSED THE WINGS OF THE REAMER TO BREAK OFF WHEN REAMER IMPACTED TIBIAL BONE WHEN REAMING OVER GUIDE WIRE. WINGS COULD NOT BE LOCATED. POST OP X-RAYS SHOWED METAL PIECES IN PATIENTS LEG MEDIAL TO TIBIAL TUNNEL. ON 04/06/2022, THE SALES REPRESENTATIVE STATED THE FOLLOWING INFORMATION VIA EMAIL: THIS EVENT OCCURRED DURING AN ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2022. ANOTHER REAMER WAS USED TO COMPLETE THE CASE. THE PATIENT IS FINE TO DATE AND HAS NO EFFECT RECORDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 14135718 ↗
  • InjuryNov 21, 2019Break; Entrapment of Device

    IT WAS REPORTED THAT A BONE TENDON BONE SURGERY WAS PERFORMED. DURING THE FEMORAL DRILLING OVER THE DRILL PIN, THE TIP OF THE LOW PROFILE DRILL HAS BROKEN OFF. THE DRILL PIN WAS THEN SLIGHTLY BENT. THE SURGEON WANTED TO RETRIEVE THE BROKEN FRAGMENT OF THE DRILL FROM THE KNEE, BUT COULD NOT FIND IT. THE PATIENT WAS X-RAYED DURING THE SURGERY TO FIND THE BROKEN FRAGMENT, BUT THE SURGEON COULD NOT LOCATE IT ON THE IMAGE CONVERTER. THE BROKEN PART REMAINS IN THE PATIENT'S KNEE. ACCORDING TO THE SURGEON, HE DOES NOT WANT TO PERFORM A SECOND SURGERY. OVERALL, IT WAS A VERY DIFFICULT SURGERY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH ARTHREX DEVICE AR-1409.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9353297 ↗
  • InjuryNov 21, 2017Material Fragmentation

    IT WAS REPORTED THAT AS THE SURGEON WAS PERFORMING AN ARTHROSCOPIC LEFT KNEE ACL RECONSTRUCTION PROCEDURE, THE HEAD OF THE REAMER BROKE AS IT WAS BEING STARTED. THE HEAD OF THE REAMER WAS IN CONTACT WITH THE PATIENT'S BONE AND BROKE INTO THREE PIECES - SHAFT AND TWO BLADES. ONE OF THE BLADES WAS EMBEDDED INTO THE BONE AND THE OTHER WAS LOST FROM VIEW. THE SALES REP SUGGESTED TO CHECK THE POSTERIOR COMPARTMENT AND A BASIC EXAMINATION WAS CARRIED OUT. THE SURGEON WAS UNCONCERNED, HOWEVER, REQUESTED AN X-RAY BE TAKEN WHILST IN RECOVERY. FOLLOW-UP INVESTIGATION: DEVICE IS NOT AVAILABLE FOR RETURN AS IT WAS KEPT BY THE HOSPITAL IN A SHARPS CONTAINER AND CANNOT BE RETRIEVED. THE THIRD PIECE OF THE DEVICE WAS LEFT IN THE POSTERIOR COMPARTMENT OF THE PATIENT'S JOINT. AN IMAGE INTENSIFIER WAS USED TO LOCATE THE FRAGMENTS AND THEREFORE A PICTURE IS UNABLE TO BE PROVIDED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7048070 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →