Arthrex® AR-14244Recalls & reported problems
Arthrex, Inc. · FRAC PLT SCRW LOC. CANC. 4.0X44MM TI · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867017412), received between May 3, 2024 and Dec 26, 2025 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Loosening of Implant Not Related to Bone-Ingrowth2
- Physical Resistance/Sticking1
Patient problems
- Failure of Implant2
- Nonunion/Delayed-union Bone Fracture2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 26, 2025Device Dislodged or Dislocated; Loosening of Implant Not Related to Bone-Ingrowth; Device Dislodged or Dislocated; Loosening of Implant Not Related to Bone-Ingrowth
ADDITIONAL INFORMATION WAS RECEIVED ON 12-MAR-2026:ON (b)(6) 2025, THE FAILED PROXIMAL HUMERAL PLATE WAS INITIALLY IDENTIFIED DURING A CLINIC VISIT. A REVISION SURGERY WAS PERFORMED, DURING WHICH THE PREVIOUSLY IMPLANTED DEVICES WERE REMOVED. PER COMMUNICATION FROM THE OPERATIVE SURGEON WHO PERFORMED THE REVISION PROCEDURE, THE BUSHINGS WERE OBSERVED TO HAVE DISENGAGED FROM THE PLATE. ADDITIONAL INFORMATION WAS RECEIVED ON 20-APR-2026:ALL IMPLANTED ARTHREX DEVICES FROM THE REVISION PROXIMAL HUMERUS ORIF PERFORMED ON (b)(6) 2025 WERE EXPLANTED DURING A SUBSEQUENT REVISION PROXIMAL HUMERUS PROCEDURE WITH PLACEMENT OF A SPACER ON (b)(6) 2026. NO ARTHREX IMPLANTS WERE IMPLANTED DURING THE (b)(6) 2026 PROCEDURE. THE REVISION PROCEDURE WAS PERFORMED AT THE SAME FACILITY AS THE PRIOR REVISION PROXIMAL HUMERUS ORIF CONDUCTED ON (b)(6) 2025; HOWEVER, THE PROCEDURE WAS PERFORMED BY A DIFFERENT SURGEON. THE PATIENT¿S CURRENT CONDITION WAS REPORTED AS STABLE. THE SALES REPRESENTATIVE WAS NOT PRESENT FOR THE (b)(6) 2026 SURGERY AND REPORTED BEING INFORMED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT IS SCHEDULED TO UNDERGO A TOTAL SHOULDER REPLACEMENT IN (b)(6).ADDITIONAL INFORMAT
Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture
MAUDE report 23897424 ↗InjuryDec 26, 2025Device Dislodged or Dislocated; Loosening of Implant Not Related to Bone-Ingrowth; Device Dislodged or Dislocated; Loosening of Implant Not Related to Bone-Ingrowth
ADDITIONAL INFORMATION WAS RECEIVED ON (b)(6)2026:ON (b)(6)2025, THE FAILED PROXIMAL HUMERAL PLATE WAS INITIALLY IDENTIFIED DURING A CLINIC VISIT. A REVISION SURGERY WAS PERFORMED, DURING WHICH THE PREVIOUSLY IMPLANTED DEVICES WERE REMOVED. PER COMMUNICATION FROM THE OPERATIVE SURGEON WHO PERFORMED THE REVISION PROCEDURE, THE BUSHINGS WERE OBSERVED TO HAVE DISENGAGED FROM THE PLATE. ADDITIONAL INFORMATION WAS RECEIVED ON (b)(6)2026:ALL IMPLANTED ARTHREX DEVICES FROM THE REVISION PROXIMAL HUMERUS ORIF PERFORMED ON (b)(6)2025 WERE EXPLANTED DURING A SUBSEQUENT REVISION PROXIMAL HUMERUS PROCEDURE WITH PLACEMENT OF A SPACER ON(b)(6)2026. NO ARTHREX IMPLANTS WERE IMPLANTED DURING THE (b)(6)2026 PROCEDURE. THE REVISION PROCEDURE WAS PERFORMED AT THE SAME FACILITY AS THE PRIOR REVISION PROXIMAL HUMERUS ORIF CONDUCTED ON (b)(6)2025; HOWEVER, THE PROCEDURE WAS PERFORMED BY A DIFFERENT SURGEON. THE PATIENT¿S CURRENT CONDITION WAS REPORTED AS STABLE. THE SALES REPRESENTATIVE WAS NOT PRESENT FOR THE (b)(6) 2026 SURGERY AND REPORTED BEING INFORMED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT IS SCHEDULED TO UNDERGO A TOTAL SHOULDER REPLACEMENT IN MAY.ADDITIONAL INFORMATION WAS REC
Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture
MAUDE report 23897328 ↗MalfunctionMay 3, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING AN ORIF PROXIMAL HUMERUS SURGERY THE SCREW AR-14244 GOT STUCK IN THE HUMERAL PLATE AR- 14003 (LOT NUMBER: 15103696). THE SURGEON HAD TO REMOVE THE PLATE TOGETHER WITH THE SCREW AND A NEW PLATE HAD TO BE USED. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19238024 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page