Arthrex® AR-1529Recalls & reported problems
Arthrex, Inc. · SUTURE TENSIONER WITH TENSIOMETER · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867018556), received between Feb 14, 2022 and Feb 14, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Markings/Labelling Problem1
- Improper or Incorrect Procedure or Method1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryFeb 14, 2022Improper or Incorrect Procedure or Method; Device Markings/Labelling Problem
IT WAS REPORTED THAT DURING AN ARTHRO ASSISTED LATARJET SURGERY THE PATIENT¿S CORACOID WAS TORN THROUGH BY THE FIBERTAPE CERCLAGE SYSTEM AND HALF OF THE GLENOID WAS TORN THROUGH DUE TO A MISUNDERSTANDING REGARDING THE USED TENSIONER AND THE SCALE, WHICH LED TO A FALSE ADJUSTMENT OF THE TENSIONER OF ABOVE 350 NEWTONS. TWO TENSIONERS WERE SENT OUT FOR THIS CASE WHICH WAS AN ARTHROSCOPIC ASSISTED LATARJET CERCLAGE. THE TENSIONER LISTED (AR-7800 AND AR-7801) PLUS THE BLUE ACL TENSIONER AR-1529. AR-1529 IS THE TENSIONER THAT WAS PREVIOUSLY ALWAYS USED BY THE ARTHREX EMPLOYEE, WHICH MAX¿S OUT AT JUST OVER 100 NEWTONS WITH THE UNIT OF MEASURE LISTED ON THERE IN LBF AND NEWTONS. THE CASE WENT REALLY WELL UNTIL TENSIONING WHERE THE ARTHREX EMPLOYEE DIDN¿T REALISE THAT THE SURGEON WAS USING THE BLACK TENSIONER (AR- 7800 & AR-7801). THIS TENSIONER DOESN¿T HAVE THE UNIT OF MEASURE LISTED ON IT, AND THE SCALE IS IN LBF. THE SURGEON ASKED HOW MUCH TO TENSION TO AND THE ARTHREX EMPLOYEE REPLIED THAT THE SURGEON CAN MAX THE TENSIONER OUT AS 100 NEWTONS ARE DESIRED, IT MAX¿S OUT AT THAT POINT. THE SURGEON THEN TENSIONED TO AROUND 80 IN LBF SCALE, WHICH WAS OVER 350 NEWTONS. THE PATIENT¿S CORACOID W
Patient problems: Insufficient Information
MAUDE report 13526856 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page