Arthrex® AR-1595TCRecalls & reported problems
Arthrex, Inc. · DRL PIN,ACL T-ROPE CLOSED EYELET,4MM · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00888867019904), received between May 21, 2024 and Mar 20, 2026 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Break” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break11
- Defective Device1
- Dull, Blunt1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information5
- Foreign Body In Patient1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2026Break
ON 27-FEB-2026, AN FDA MEDWATCH NOTIFICATION WAS RECEIVED VIA EMAIL. A FACILITY REPRESENTATIVE REPORTED THAT AN AR-1595TC DRILL PIN BROKE OFF DURING A PROCEDURE ON (B)(6) 2026. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24649053 ↗MalfunctionFeb 20, 2026Break
ON 02-FEB -2026, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA (B)(4) THAT AN AR-1595TC PIN BROKE WHEN IT WAS PRESSED AGAINST THE BONE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24399978 ↗MalfunctionJan 22, 2026Break
ON 07-JAN-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (QTY. 2) OF AN AR-1595TC DRL PIN, ACL T-ROPE CLOSED EYELET,4MM SPADE TIP PINS BROKE IN THE FEMUR. THIS WAS DISCOVERED DURING AN ACL RECONSTRUCTION AND THE FRAGMENTS WERE RETRIEVED.
Patient problems: Insufficient Information
MAUDE report 24144750 ↗MalfunctionJan 22, 2026Break
ON (B)(6) 2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (QTY. 2) OF AN AR-1595TC DRL PIN, ACL T-ROPE CLOSED EYELET,4MM SPADE TIP PINS BROKE IN THE FEMUR. THIS WAS DISCOVERED DURING AN ACL RECONSTRUCTION AND THE FRAGMENTS WERE RETRIEVED.
Patient problems: Insufficient Information
MAUDE report 24144311 ↗MalfunctionDec 24, 2025Break; Break
IT WAS REPORTED THAT DURING A KNEE SURGERY THE TIPS OF BOTH PINS AR-1595TC (LOT: 15458338 AND 15461606) BROKE DURING USE. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23888417 ↗MalfunctionDec 24, 2025Break; Break
IT WAS REPORTED THAT DURING A KNEE SURGERY THE TIPS OF BOTH PINS AR-1595TC (LOT: 15458338 AND 15461606) BROKE DURING USE. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23888340 ↗InjuryAug 19, 2025Break; Break
ON (B)(6) 2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-1595TC, DRILL PIN, ACL TIGHTROPE®, CLOSED EYELET, 4 MM, THE SPADE TIP BROKE OFF WHEN DRILLING THROUGH FEMORAL CONDYLE. THIS WAS DETECTED DURING USE IN AN ACL REVISION PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS THE BROKEN TIP WAS RETRIEVED AND A 2.4 DRILL PIN WAS USED TO CORRECT IT. ON (B)(6) 2025 - THE COMPLAINT INITIATOR REPLIED THAT THE PURPOSE OF THE REVISION PROCEDURE WAS BECAUSE THE PATIENT'S INITIAL PATELLAR TENDON ACL RECONSTRUCTION HAD FAILED WHEN THE PATIENT RETURNED TO SPORT TOO EARLY WITHOUT COMPLETING AN ADEQUATE RETURN TO PLAY/ REHABILITATION PROGRAM. ON (B)(6) 2025 - THE COMPLAINT INITIATOR RECEIVED UPDATES THAT ALTHOUGH INITIALLY IT WAS BELIEVED THAT ALL OF THE BROKEN MATERIAL WAS REMOVED FROM THE PATIENT DURING THE SURGERY. SUBSEQUENTLY ON A PRECAUTIONARY X-RAY THERE WAS STILL METAL LEFT IN THE PATIENT. THIS PATIENT HAD A SUBSEQUENT FOLLOW UP SURGERY YESTERDAY FOR THE FOREIGN MATERIAL TO BE REMOVED WHICH WAS PART OF THE SPADETIP DRILL PIN. NO REPORTED ¿PATIENT HARM¿ APART FROM HAVING TO HAVE A FOLLOW-UP SURGERY.
Patient problems: Foreign Body In Patient
MAUDE report 22842751 ↗MalfunctionApr 25, 2025Break; Break
ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT FOR AN AR-1595TC, DRILL PIN, ACL TIGHTROPE®, CLOSED EYELET, 4 MM, AFTER MICRO FRACTURING THE LATERAL CONDYLE, THE 4MM SPADE TIP WAS INSERTED INTO THE MICROFRACTURE HOLE BY HAND, THE POWER TOOL WAS ATTACHED AND ACTIVATED ON THE DRILL, THE HEAD OF THE SPADE TIP PIN BROKE OFF AND WAS IMBEDDED IN THE LATERAL FEMORAL CONDYLE. A CURETTE WAS USED TO REMOVE THE METAL FRAGMENT AND THEY PROCEEDED WITH THE OPERATION. THIS WAS DETECTED ON (B)(6) 2025 DURING USE IN AN ACLBTB + LET + LATERAL MENISCAL REPAIR PROCEDURE. THE PROCEDURE WAS COMPLETED AS ANOTHER 4MM SPADE TIP DRILL PIN WAS USED.
Patient problems: Insufficient Information
MAUDE report 21910144 ↗MalfunctionApr 4, 2025Break; Break
ON 17 MARCH 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-1595TC DRL PIN, ACL T-ROPE CLOSED EYELET,4MM WAS REPORTED TO HAVE SNAPPED OFF AT THE TIP. THIS OCCURRED ON (B)(6) 2025 DURING USE IN AN UNKNOWN PROCEDURE. THERE WAS CASE INVOLVEMENT.
Patient problems: Insufficient Information
MAUDE report 21771997 ↗MalfunctionFeb 26, 2025Dull, Blunt; Defective Device; Dull, Blunt; Defective Device
ADDITIONAL INFORMATION WAS RECEIVED ON 4/21/2025: DURING AN ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2025, IT WAS IDENTIFIED THAT THE AR-1595TC DRILL PIN HAD AN INCONSISTENT DIAMETER ALONG ITS LENGTH. THIS ISSUE PREVENTED THE PROPER ADVANCEMENT OF THE CANNULATED REAMER OVER THE PIN. THE DRILL PIN WAS NOT STUCK, BUT ITS IRREGULAR DIMENSIONS INTERFERED WITH THE REAMING PROCESS. NO PIECES OF THE INSTRUMENT BROKE INSIDE THE PATIENT, AND NO CASE DELAY WAS REPORTED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY OPENING A NEW 4MM DRILL PIN (AR-1595TC). THE PATIENT'S BONE QUALITY WAS ASSESSED AS GOOD, AND NO ADDITIONAL ANESTHESIA WAS REQUIRED. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21468924 ↗MalfunctionFeb 24, 2025Break; Break
ON 02/11/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-1595TC DRILL PIN IS DAMAGED. THE TIP OF THE PIN BROKE OFF WHILE BEING DRILLED INTO A FEMUR AND BECAME STUCK. IT WAS RETRIEVED WITH A GRASPER AND NO METAL WAS LEFT IN THE PATIENT. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21447897 ↗MalfunctionMay 21, 2024Break; Break
ON 04/23/2024, IT WAS REPORTED BY A DISTRIBUTOR VIA (B)(4) THAT AN AR-1595TC DRILL PINS SHADE TIP BROKE OFF. THIS OCCURRED DURING AN ACL RECONSTRUCTION ON (B)(6) 2024 WHEN INSERTED INTO THE BONE, THE DEVICE BROKE IT TOOK A COUPLE OF MINUTES TO RETRIEVE THE FRAGMENT. THE CASE WAS COMPLETED USING ANOTHER PIN
Patient problems: Insufficient Information
MAUDE report 19357693 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page