Arthrex® AR-1662BCC-7Recalls & reported problems
Arthrex, Inc. · SWVLK TENO BIO-COMP 7X 19.1MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867020603), received between Nov 27, 2018 and May 21, 2019 — classified as 2 injuries.
The most frequently reported specific device problem is “Loose or Intermittent Connection” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Loose or Intermittent Connection2
- Break1
- Device Slipped1
- Material Fragmentation1
Patient problems
- Pain2
- Adhesion(s)1
- Foreign Body In Patient1
- Loss of Range of Motion1
- Muscle Spasm(s)1
- Weakness1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 21, 2019Loose or Intermittent Connection
IT WAS REPORTED DURING AN INITIAL LABRAL AND SLAP REPAIR PROCEDURE, THE FOLLOWING ARTHREX PARTS WERE IMPLANTED ON (B)(6) 2017: AR-1662BCC-7 // LOT: 10153102. AR-2324BCC-2 // LOT: 10140067. AR-2923BC // LOT: 10140969. APPROXIMATELY A YEAR AFTER THE INITIAL PROCEDURE, THE PATIENT STARTED EXPERIENCING SEVERE PAIN AND LIMITED RANGE OF MOTION IN THEIR LEFT SHOULDER. THE SURGEON PERFORMED A REVISION LEFT SHOULDER SCOPE, LYSIS OF ADHESIONS AND REMOVAL OF LOOSE BODY PROCEDURE ON (B)(6) 2019. THE LOOSENED BICEP ANCHOR (AR-1662BCC-7 / LOT: 10153102) WAS RETRIEVED SHATTERED IN MULTIPLE SMALL PIECES, WHICH HAD DEBRIDED ADHESION'S ONTO THE PATIENT'S ROTATOR CUFF. THE PRIMARY AND REVISION PROCEDURE WERE PERFORMED AT THE SAME FACILITY BY THE SAME SURGEON. NO ADDITIONAL ARTHREX PARTS WERE IMPLANTED DURING THE REVISION PROCEDURE. THE FOLLOWING ARTHREX PARTS WERE EXPLANTED DURING THE REVISION PROCEDURE: AR-1662BCC-7 // LOT: 10153102. ADDITIONAL INFORMATION PROVIDED VIA MEDICAL RECORDS ON MAY 6, 2019: ACCORDING TO THE MEDICAL RECORDS PROVIDED THE PATIENT HAD CONTINUED PAIN FROM (B)(6) 2017 THROUGH (B)(6) 2018. THE PHYSICAL THERAPY RECORDS DATED (B)(6) 2017 STATED "SHOULDER IS A LITTLE SORE ON ARRIVAL
Patient problems: Adhesion(s); Loss of Range of Motion; Pain
MAUDE report 8627538 ↗InjuryNov 27, 2018Break; Material Fragmentation; Loose or Intermittent Connection; Device Slipped
IT WAS REPORTED VIA FDA MEDWATCH LETTER (MW5080256) THAT A PATIENT UNDERWENT A TENODESIS SURGERY ON THEIR RIGHT SHOULDER ON (B)(6) 2017. THE FIBERLINK SUTURE WAS PASSED THROUGH THE TORN BICEPS TENDON AND THEN FED THROUGH THE EYELET OF AN ARTHREX SWIVELOCK ANCHOR AND SEATED INTO THE PILOT HOLE ON THE PROXIMAL HUMERUS. ELEVEN MONTHS POST-OP, THE PATIENT HAD A SUDDEN PAIN/SPASM IN THEIR RIGHT SHOULDER. IMAGING TESTS REVEALED NUMEROUS "LOOSE BODIES" IN THE FRONT AND BACK OF THE PATIENT'S SHOULDER JOINT. ON (B)(6) 2018, AN ARTHROSCOPIC PROCEDURE WAS PERFORMED FOR REMOVAL OF THE FOREIGN BODIES. THE BROKEN FRAGMENTS OF THE SCREW WERE REMOVED VIA THREE HOLES IN THE PATIENT'S SHOULDER. THE BICEPS TENDON WAS NOTED TO BE ABSENT. THE SCREW WAS UNSTABLE AND DEFECTIVE. SEVEN WEEKS POST-OP FROM THE SECOND SURGERY, THE PATIENT IS STILL EXPERIENCING SEVERE PAIN AND SPASMS WITH GREATLY DECREASED STRENGTH IN THE RIGHT ARM. NO FUNCTION TO BICEPS LONG HEAD DEFORMITY AND PERMANENT DEFORMITY. ADDITIONAL INFORMATION PROVIDED 11/15/2018: THE PATIENT HAS PROVIDED SCANNED ARTHROSCOPIC PHOTOS. THE SCANS SHOW THE ¿FOREIGN BODIES¿ FLOATING IN HER SHOULDER. THIS FAILED DEVICE HAS CAUSED HER BICEPS TO SLIP OFF TH
Patient problems: Foreign Body In Patient; Muscle Spasm(s); Pain; Weakness
MAUDE report 8107707 ↗
Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page