Arthrex, Inc. · SCRWDRV,CANN.,PINLOCK II,3.5MM HEX · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867022423), received between Sep 12, 2016 and Jul 26, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Detachment of Device or Device Component1
  • Difficult to Remove1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1
  • Insufficient Information1
  • No Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 26, 2024Break; Break

    ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-1896 SCRWDRV,CANN.,PINLOCK II,3.5MM HEX SNAPPED DURING USE. THIS OCCURRED ON (B)(6) 2024 IN (B)(6) HOSPITAL DURING AN ACL REVISION. IT WAS REPORTED THAT A METAL INTERFERENCE SCREW WAS BEING INSERTED INTO A FEMORAL TUNNEL. THE FEMORAL TUNNEL DIAMETER WAS 10MM, BTB GRAFT WAS 10MM AND THE SURGEON SELECTED A 9MM SCREW. ON INSERTION THE SURGEON NOTED THE "TOUGH" INSERTION AND WITH ROUGHLY 2 TURNS TO GO THE SCREW INSERTER SNAPPED AT THE HEX DRIVER POINT (PICTURED). THE BROKEN FRAGMENT WAS LEFT IN THE INSERTION POINT OF THE SCREW. THE NITIOL WIRE WAS LEFT IN PLACE, SO THAT THE FRAGMENT WAS NOT LOST IN THE JOINT SPACE AND A GRASPER WAS USED TO REMOVE THE FRAGMENT. THE CASE WAS COMPLETED SUCCESSFULLY USING AN 8MM PEEK INTERFERENCE SCREW WHICH HAD BEEN TAPPED BY THE METAL SCREW. THERE WAS CASE INVOLVEMENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 19839481 ↗
  • MalfunctionMay 10, 2021Break

    IT WAS REPORTED, THAT DURING AN ACL REVISION PROCEDURE, WHEN THE SURGEON WAS REMOVING THE SCREW THE TIP OF THE AR-1896, BROKE. IT IS UNKNOWN AT THIS TIME IF THE SCREW REMOVED IS AN ARTHREX DEVICE. ALL PIECES WERE REMOVED. THE CASE WAS COMPETED USING A COMPETITOR DEVICE. THE AR-1896 WILL BE RETURNED.

    Patient problems: Insufficient Information

    MAUDE report 11802291 ↗
  • InjuryDec 16, 2019Break; Difficult to Remove; Detachment of Device or Device Component

    IT WAS REPORTED DURING AN ACL PROCEDURE THE TIP OF THE DRIVER BROKE OFF IN THE PATIENT. THE DRIVER TIP WAS STUCK IN THE SCREW AND COULD NOT BE RETRIEVED. REPORTER STATED THE SCREW WAS SEATED AND ALL IS SECURE. PER REPORTER, THE PATIENT IS FINE TO DATE AND THE DOCTOR FEELS THERE IS NO NEED FOR A REVISION.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9473440 ↗
  • InjurySep 12, 2016Break

    IT WAS REPORTED VIA A FACILITY MEDWATCH ((B)(4)) THAT ON (B)(6) 2016 DURING AN ACL RECONSTRUCTION PROCEDURE, THE SURGEON WAS USING THE AR-1896 SCREWDRIVER TO PLACE A SCREW AND THE TIP OF THE SCREWDRIVER BROKE-OFF IN THE HEAD OF THE SCREW. SEVERAL ATTEMPTS HAVE BEEN MADE TO GATHER ADDITIONAL INFORMATION FROM THE REPORTER OF MEDWATCH (B)(4). TO DATE NO RESPONSE HAS BEEN RECEIVED.

    Patient problems: No Information

    MAUDE report 5943432 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →