Skip to content

Arthrex® AR-1922BCMRecalls & reported problems

Arthrex, Inc. · BIOCOMPOSITE PUSHLOCK SP 4.5MM X 28MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00888867022898), received between Aug 15, 2024 and Sep 9, 2025 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 3 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 4 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Device Dislodged or Dislocated3
  • Positioning Failure3
  • Failure to Advance2
  • Separation Failure2
  • Detachment of Device or Device Component1
  • Mechanical Jam1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • Insufficient Information1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2025Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY WHEN THE PUSHLOCK WAS INSERTED THE TIP FELL OFF. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23009719 ↗
  • MalfunctionSep 3, 2025Failure to Advance; Separation Failure; Failure to Advance; Separation Failure

    IT WAS REPORTED DURING A SURGERY THAT THE ANCHOR DID NOT SEPARATE FROM THE DRIVER AS IT WAS HIT INTO THE BONE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22953155 ↗
  • MalfunctionSep 3, 2025Failure to Advance; Separation Failure; Failure to Advance; Separation Failure

    IT WAS REPORTED DURING A SURGERY THAT THE ANCHOR DID NOT SEPARATE FROM THE DRIVER AS IT WAS HIT INTO THE BONE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22953144 ↗
  • MalfunctionApr 17, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A CUFF SURGERY THE ANCHOR CAME ALONG WITH THE SHAFT WHEN THE SHAFT WAS REMOVED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21856832 ↗
  • MalfunctionFeb 25, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT DURING A CUFF SURGERY WHILE DRILLING PREHOAL TO GET THE ANCHOR DOWN, TURNING COUNTERCLOCKWISE AS INSTRUCTIONS SAYS BUT WHEN PULLING UP THE ANCHOR FOLLOWS ALONG. TWO ANCHORS FAILED AND THE THIRD ONE THEY DID NOT WANT TO KEEP IT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21460460 ↗
  • MalfunctionFeb 25, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT DURING A CUFF SURGERY WHILE DRILLING PREHOAL TO GET THE ANCHOR DOWN, TURNING COUNTERCLOCKWISE AS INSTRUCTIONS SAYS BUT WHEN PULLING UP THE ANCHOR FOLLOWS ALONG. TWO ANCHORS FAILED AND THE THIRD ONE THEY DID NOT WANT TO KEEP IT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21460322 ↗
  • MalfunctionFeb 24, 2025Positioning Failure; Device Dislodged or Dislocated; Positioning Failure; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A CUFF SURGERY WHILE DRILLING PREHOAL TO GET THE ANCHOR DOWN, TURNING COUNTERCLOCKWISE AS INSTRUCTIONS SAYS BUT WHEN PULLING UP THE ANCHOR FOLLOWS ALONG. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21445757 ↗
  • MalfunctionFeb 12, 2025Mechanical Jam; Physical Resistance/Sticking; Mechanical Jam; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING A BICEPS TENDON SURGERY THE ANCHOR DID NOT SEPARATE FROM THE DRIVER. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21363799 ↗
  • MalfunctionJan 16, 2025Break; Break

    IT WAS REPORTED THAT DURING AN AUTOGRAFT CUFF SURGERY THE TIP DID BREAK OFF, IMPOSSIBLE TO USE. THE BROKEN OFF PIECE HAS BEEN RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21170542 ↗
  • MalfunctionJan 3, 2025Break; Break

    ON 12/10/2024 IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA (B)(4) THAT AN AR-1922BCM BIOCOMPOSITE PUSHLOCK TIP BROKE DURING INSERTION. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6)2024. ADDITIONAL INFORMATION RECEIVED ON 12/17/2024: THIS WAS DISCOVERED DURING AN RCR PROCEDURE. THE ANCHOR NEVER MADE IT INTO THE BONE AND WAS ABLE TO BE REMOVED FROM THE PATIENT. THE CASE WAS DELAYED FIVE MINUTES WITHOUT ADDITIONAL ANESTHESIA. THE CASE WAS COMPLETED USING AN AR-2324BCSP SWIVELOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21067909 ↗
  • MalfunctionOct 2, 2024Break; Break

    IT WAS REPORTED THAT DURING AN INSTABILITY REFIXATION LABRUM SURGERY THE TIP OF THE DEVICE BROKE OFF. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20352761 ↗
  • MalfunctionSep 20, 2024Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE ANCHOR WAS DEFECTIVE. THE TIP WAS NOT STABLE AND IS BENT DURING INSERTION. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20268458 ↗
  • MalfunctionSep 20, 2024Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE ANCHOR WAS DEFECTIVE. THE TIP WAS NOT STABLE AND IS BENT DURING INSERTION. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20268366 ↗
  • MalfunctionSep 20, 2024Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE ANCHOR WAS DEFECTIVE. THE TIP WAS NOT STABLE AND IS BENT DURING INSERTION. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20268313 ↗
  • MalfunctionAug 15, 2024Break; Break

    ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-1922BCM SUTURE ANCHORS PILOT SNAPPED OFF. THIS OCCURRED DURING A ROTATOR CUFF REPAIR ON (B)(6) 2024, THE PILOT SNAPPED OFF AND COULDN'T BE THREADED BACK ON. ALL FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED USING ANOTHER AR-1922BCM. NO ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: Insufficient Information

    MAUDE report 19991830 ↗

Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices →

Sources and updates