Arthrex® AR-1927BCF-45Recalls & reported problems

Arthrex, Inc. · BIO-COMPSI CRKSCRW FT 4.5X 14MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Arthrex Bio-Composite Corkscrew anchor is an absorbable anchor used to reattach soft tissue to bone. Most reports describe the anchor breaking during insertion, some during rotator cuff repairs. The 156 reports include no deaths and 6 injuries. No recalls name this device, and reports fell from 59 in the prior 12 months to 33 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
164
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 164 adverse event reports that list this exact device (DI 00888867023376), received between Aug 9, 2016 and Sep 29, 2026 — classified as 7 injuries, 157 malfunctions.

The most frequently reported specific device problem is “Break” (158 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 3 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 3 reportsJun 2024: 1 reportsJul 2024: 14 reportsAug 2024: 2 reportsSep 2024: 6 reportsOct 2024: 7 reportsNov 2024: 4 reportsDec 2024: 1 reportsJan 2025: 5 reportsFeb 2025: 4 reportsMar 2025: 1 reportsApr 2025: 7 reportsMay 2025: 7 reportsJun 2025: 5 reportsJul 2025: 6 reportsAug 2025: 8 reportsSep 2025: 4 reportsOct 2025: 7 reportsNov 2025: 0 reportsDec 2025: 9 reportsJan 2026: 2 reportsFeb 2026: 5 reportsMar 2026: 11 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 4 reportsAug 2026: 1 reportsSep 2026: 7 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break158
  • Patient Device Interaction Problem3
  • Fitting Problem2
  • Material Deformation2
  • Detachment Of Device Component1
  • Difficult to Insert1
  • Difficult to Remove1
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions144
  • Insufficient Information9
  • Foreign Body In Patient5
  • Failure of Implant3
  • No Known Impact Or Consequence To Patient3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Pain2
  • Bacterial Infection1

Recent adverse event reports

The 25 most recent of 164 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933468 ↗
  • MalfunctionSep 29, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933275 ↗
  • MalfunctionSep 15, 2026Break

    On 1st September 2026, it was reported by an Arthrex's employee via an Email that for 1 unit of AR-1927BCF-45, 4.5 mm BioComposite-Corkscrew® FT Suture Anchor with one #2 FiberWire® and one TigerWire®, its anchor broke during insertion. This was detected during an arthroscopic rotator cuff repair surgery of the shoulder on (b)(6) 2026. The procedure was completed successfully by using another new AR-1927BCF-45. There were no patient harm and no secondary procedure needed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50774414 ↗
  • MalfunctionSep 15, 2026Break

    On 1st September 2026, it was reported by an Arthrex's employee via an Email that for 1 unit of AR-1927BCF-45, 4.5 mm BioComposite-Corkscrew® FT Suture Anchor with one #2 FiberWire® and one TigerWire®, its anchor broke during insertion.This was detected during a rotator cuff repair surgery on (b)(6) 2026. The procedure was completed successfully by using another new AR-1927BCF-45. There were no patient harm and no secondary procedure needed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50772164 ↗
  • MalfunctionSep 15, 2026Break

    On 1st September 2026, it was reported by an Arthrex's employee via an Email that for 2 units of AR-1927BCF-45, 4.5 mm BioComposite-Corkscrew® FT Suture Anchor with one #2 FiberWire® and one TigerWire®, both its anchor broke off during insertion.The 1st anchor broke during insertion, and a 2nd AR-1927BCF-45 was changed to but the same issue happened again.This was detected during a rotator cuff repair surgery on (b)(6) 2026. The procedure was completed successfully by using another new AR-1927BCF-45. There were no patient harm and no secondary procedure needed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50771874 ↗
  • MalfunctionSep 15, 2026Break

    On 1st September 2026, it was reported by an Arthrex's employee via an Email that for 2 units of AR-1927BCF-45, 4.5 mm BioComposite-Corkscrew® FT Suture Anchor with one #2 FiberWire® and one TigerWire®, both its anchor broke off during insertion.The 1st anchor broke during insertion, and a 2nd AR-1927BCF-45 was changed to but the same issue happened again.This was detected during a rotator cuff repair surgery on (b)(6) 2026. The procedure was completed successfully by using another new AR-1927BCF-45. There were no patient harm and no secondary procedure needed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50771667 ↗
  • MalfunctionSep 10, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50712640 ↗
  • InjuryAug 10, 2026Patient Device Interaction Problem

    Patient problems: Bacterial Infection; Pain; Purulent Discharge; Swelling/ Edema

    MAUDE report 50191619 ↗
  • MalfunctionJul 31, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50098945 ↗
  • MalfunctionJul 27, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50036093 ↗
  • MalfunctionJul 27, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50035040 ↗
  • MalfunctionJul 27, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50008410 ↗
  • MalfunctionApr 1, 2026Break

    ON (B)(6) 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 2 UNITS OF AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, BOTH THE ANCHORS BROKE DURING INSERTION. THE 1ST AR-1927BCF-45 BROKE DURING INSERTION, AND ANOTHER AR-1927BCF-45 WAS CHANGED TO, BUT THE SAME ISSUE HAPPENED AGAIN. THIS WAS DETECTED DURING AN ARTHROSCOPIC SHOULDER SURGERY FOR ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING A THIRD AR-1927BCF-45. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24752801 ↗
  • MalfunctionApr 1, 2026Break

    ON (B)(6) 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 2 UNITS OF AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, BOTH THE ANCHORS BROKE DURING INSERTION. THE 1ST AR-1927BCF-45 BROKE DURING INSERTION, AND ANOTHER AR-1927BCF-45 WAS CHANGED TO, BUT THE SAME ISSUE HAPPENED AGAIN. THIS WAS DETECTED DURING AN ARTHROSCOPIC SHOULDER SURGERY FOR ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING A THIRD AR-1927BCF-45. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24752643 ↗
  • MalfunctionMar 26, 2026Break

    ON (B)(6) 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1927BCF-45. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698178 ↗
  • MalfunctionMar 26, 2026Break

    ON (B)(6) 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1927BCF-45. THERE WERE NO PATIENT HAM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24697983 ↗
  • MalfunctionMar 26, 2026Break

    ON 6TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1927BCF-45. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24697834 ↗
  • MalfunctionMar 26, 2026Break

    ON 6TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, IT'S ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1927BCF-45. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24697774 ↗
  • MalfunctionMar 25, 2026Break

    ON 5TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, THE ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS ANOTHER NEW AR-1927BCF-45 WAS USED. AR-1922P AND AR-1927PTB-45 WERE USED TO PREPARE THE BONE SOCKET. THE BONE QUALITY OF PATIENT WAS HARD, AND NO FRAGMENTS WERE LEFT IN THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 24687653 ↗
  • MalfunctionMar 25, 2026Break

    ON 5TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, THE ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING USE IN A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS ANOTHER NEW AR-1927BCF-45 WAS USED. AR-1922P AND AR-1927PTB-45 WERE USED TO PREPARE THE BONE SOCKET. THE BONE QUALITY OF PATIENT WAS HARD, AND NO FRAGMENTS WERE LEFT IN THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 24686375 ↗
  • MalfunctionMar 20, 2026Break

    ON 2ND MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 2 UNITS OF AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, BOTH ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF AND LABRUM INJURY SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1927BCF-45. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649132 ↗
  • MalfunctionMar 20, 2026Break

    ON 2ND MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 2 UNITS OF AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, BOTH ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF AND LABRUM INJURY SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1927BCF-45. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649126 ↗
  • MalfunctionMar 18, 2026Break

    ON 27TH FEB 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, THE HEAD END OF THE ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1927BCF-45. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24634004 ↗
  • MalfunctionMar 18, 2026Break

    ON 27TH FEB 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A PATELLAR LIGAMENT RECONSTRUCTION SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1927BCF-45. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24630598 ↗
  • MalfunctionMar 18, 2026Break

    ON 27TH FEB 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1927BCF-45, 4.5 MM BIOCOMPOSITE-CORKSCREW® FT SUTURE ANCHOR WITH ONE #2 FIBERWIRE® AND ONE TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1927BCF-45. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24627665 ↗

Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices →