Skip to content

Arthrex® AR-300-B001Recalls & reported problems

Arthrex, Inc. · BURR, STRAIGHT, 13MM DIAM. 2.0 MM · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867028630), received between Nov 14, 2024 and Jul 23, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent2
  • Break1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 23, 2025Material Twisted/Bent; Physical Resistance/Sticking; Material Twisted/Bent; Physical Resistance/Sticking

    ON (B)(6) 2025, A SALES REPRESENTATIVE REPORTED VIA SEMS-06959817 THAT AN AR-300-B001 BURR WAS BENT. THIS OCCURRED DURING A CASE AS THEY WENT TO START AND NOTICED THE BURR WAS SLIGHTLY BENT. THEY WERE ABLE TO REMOVE IT, DISCARD IT, AND REPLACE IT WITH ANOTHER. THERE WERE NO ADVERSE EFFECTS ON THE PATIENT OR PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22578092 ↗
  • MalfunctionJul 22, 2025Material Twisted/Bent; Material Twisted/Bent

    ON 06/30/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-300-B001 BURR STRAIGHT WAS BENT. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22568401 ↗
  • MalfunctionNov 14, 2024Break; Break

    IT WAS REPORTED THAT DURING A MINIMALINVASIVE MICA SURGERY THE ATTACHMENT BROKE OFF. THE END REMAINED ATTACHED INSIDE THE ADAPTOR. THE PART BROKE OFF OUTSIDE THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. ***UPDATE DW (B)(6) 2024: FURTHER INFORMATION WAS PROVIDED BY THE CUSTOMER THAT THE INITIAL REPORTED EVENT WAS INCORRECT. IT WAS CONFIRMED THAT THE ATTACHMENT HAD NO ISSUE BUT AN ARTHREX DRILL BIT HAS BROKEN OFF AND A PART REMAINED STUCK INSIDE THE ATTACHMENT. THE CUSTOMER WAS NOT ABLE TO PROVIDE FURTHER DETAILS WHICH DRILL BIT HAS BROKEN OFF HOWEVER DURING THE EVALUATION A PICTURE OF THE END OF DRILL BIT HAS SHOWN THAT IT HAS A DIAMETER OF 2MM AND THE CUSTOMER ONLY PURCHASED ONE AR-300-B001 WITH THE BATCH 037417 WHICH FITS THIS SPECIFICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20685244 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page