Arthrex® AR-4068CRecalls & reported problems
Arthrex, Inc. · SUTURELASSO, CRESCENT · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867031555), received between Jun 14, 2018 and Oct 23, 2018 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
- Bent1
- Detachment Of Device Component1
Patient problems
- No Code Available1
- No Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryOct 23, 2018Break
IT WAS REPORTED THAT, WHILE PIERCING THROUGH THE LABRUM WITH THE AR-4068C IN A SHOULDER LABRAL REPAIR PROCEDURE, THE TIP OF THE SUTURE LASSO BROKE OFF IN THE LABRUM NEAR THE BASE OF THE BICEPS. THIS OCCURRED WHILE THE SURGEON WAS ATTEMPTING TO PASS THE LAST SUTURE. THE SURGEON MADE AN ADDITIONAL PORTAL AND WAS ABLE TO EFFICIENTLY AND FULLY REMOVE THE BROKEN PIECE. THE PROCEDURE WAS COMPLETED, AND AN ADDITIONAL SURGERY IS NOT REQUIRED. THIS OCCURRED DURING THE INITIAL/FIRST TIME USE OF THE DEVICE.
Patient problems: No Code Available
MAUDE report 7994467 ↗InjuryJun 14, 2018Bent; Break
IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY CASE, AFTER THE CRESCENT SUTURELASSO, AR-4068C, HAD BEEN BENT MULTIPLE TIMES BY THE SURGEON, THE TIP BROKE OFF IN THE GLENOHUMERAL JOINT OF THE PATIENT AND NEEDED TO BE RETRIEVED. THE BROKEN PIECE OF THE SUTURELASSO WAS APPROXIMATELY AN INCH LONG. THE SURGEON ATTEMPTED TO RETRIEVE THE PIECE USING FLUOROSCOPY DURING THE ORIGINAL SURGERY BUT WAS UNABLE TO RETRIEVE THE PIECE ARTHROSCOPICALLY. A SECOND SURGERY WAS SCHEDULED AT A DIFFERENT HOSPITAL THE SAME DAY FOR THE REMOVAL OF THE SUTURELASSO FRAGMENT BY THE SAME SURGEON. IT WAS CONFIRMED THAT THE SECOND SURGERY TOOK PLACE AND THE SUTURELASSO FRAGMENT WAS RETRIEVED.
Patient problems: No Information
MAUDE report 7600209 ↗InjuryJun 14, 2018Break; Detachment Of Device Component
IT WAS REPORTED THAT DURING A CASE, AFTER THE CRESCENT SUTURELASSO, AR-4068C, HAD BEEN BENT MULTIPLE TIMES BY THE SURGEON, THE TIP BROKE OFF IN THE GLENOHUMERAL JOINT OF THE PATIENT AND NEEDED TO BE RETRIEVED. PRIOR TO BREAKAGE, THE SURGEON FORCED THE LASSO TO BEND AT THE TIP MULTIPLE TIMES BY PUSHING IT INTO A METAL INSTRUMENT TRAY AND ALSO USING A NEEDLE DRIVER TO MANIPULATE THE TIP OF IT. THE BROKEN PIECE OF THE SUTURELASSO WAS APPROXIMATELY AN INCH LONG. THE SURGEON ATTEMPTED TO RETRIEVE THE PIECE USING FLUOROSCOPY BUT WAS UNABLE TO RETRIEVE THE PIECE ARTHROSCOPICALLY. THE SURGERY LASTED APPROXIMATELY 4 - 4 1/2 HOURS. A SECOND SURGERY WAS SCHEDULED AT A DIFFERENT HOSPITAL THE SAME DAY FOR THE REMOVAL OF THE SUTURELASSO FRAGMENT BY THE SAME SURGEON (ARTHREX REFERENCE (B)(4)). THE NEEDLE DRIVER (PART AND LOT NUMBER UNKNOWN) THAT WAS USED TO TRY AND BEND THE SUTURELASSO BROKE DURING THE ATTEMPT. THIS OCCURRED OUTSIDE THE PATIENT FIELD AND THE NEEDLE DRIVER BROKEN PIECE WAS ACCOUNTED FOR. FOLLOW-UP INVESTIGATION: TYPE OF PROCEDURE BEING PERFORMED WAS A SHOULDER ARTHROSCOPY WITH CAPSULORRAPHY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7600202 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page