Arthrex® AR-5028C-08Recalls & reported problems
Arthrex, Inc. · BIO-COMP INTER-FERENCE SCRW · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00888867035324), received between Apr 25, 2017 and Jun 27, 2024 — classified as 3 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Fitting Problem1
- Loose or Intermittent Connection1
- Material Twisted/Bent1
- Melted1
- Positioning Failure1
Patient problems
- Foreign Body In Patient4
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information1
- Pain1
- Purulent Discharge1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 27, 2024Positioning Failure; Positioning Failure
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY THE SCREW DID NOT GRIP AND COULD NOT BE INSERTED TIBIALLY. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19626401 ↗MalfunctionJun 3, 2024Melted; Melted
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY WHILE INSERTING THE IMPLANT MELTED. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. *** 02-SEP-2021 UPDATE DW FURTHER INFORMATION WAS PROVIDED THAT THE REPORTED SCREW WAS JUST DEFORMED AND DID NOT MELT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19446930 ↗MalfunctionAug 15, 2022Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR SURGERY THE DEVICE BROKE. ACCORDING TO THE SURGEON THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15223780 ↗MalfunctionApr 25, 2022Material Twisted/Bent
IT WAS REPORTED THAT DURING ANTERIOR CRUCIATE LIGAMENT SURGERY DURING THE FIRST TURNS OF THE SCREW, WHEN INSERTING, THE SCREW THREAD ON THE SCREW HEAD TWISTED OFF DIRECTLY. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH PART NUMBER AR-5035TC-09. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14192646 ↗MalfunctionJan 4, 2022Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY THE DEVICE DID NOT FIT WELL IN THE SCREW DRIVER. A PIECE OF THE DEVICE BROKE OF INSIDE THE PATIENT, BUT WAS RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: Foreign Body In Patient
MAUDE report 13152518 ↗MalfunctionNov 30, 2021Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT, POSTERIOR CRUCIATE LIGAMENT AND POSTEROLATERAL CORNER RECONSTRUCTION SURGERY THE SCREW DRIVER DIDN'T FIT IN THE DEVICES. ONLY A VERY SHORT PART OF THE DRIVER GOT INSERTED IN THE DEVICES WHICH LEAD TO A BREAKAGE OF THE DEVICES WHILE PLACING THEM. EACH BROKEN SCREW HAD TO BE REPLACED WITH A PEEK INTERFERENCE SCREW OF ONE SIZE LARGER TO ACHIEVE PROPER FIXATION OF EACH GRAFT. THE BROKEN PARTS WERE NOT RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH DIFFERENT DEVICES. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: Foreign Body In Patient
MAUDE report 12897103 ↗MalfunctionJun 28, 2021Break; Fitting Problem
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY THE DEVICE DIDN`T FIT IN THE SCREWDRIVER. AFTER THE SCREWDRIVER WAS FORCED INTO THE DEVICE IT WAS NOT POSSIBLE TO IMPLANT THE DEVICE PROPERLY IN THE TIBIAL TUNNEL. THE DEVICE BROKE INSIDE THE PATIENT AND A CLAMP WAS USED TO SECURE THE IMPLANT IN PLACE. THE BROKEN PART REMAINED INSIDE THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. IT WAS NOT NECESSARY TO DO A SECOND SURGERY.
Patient problems: Insufficient Information
MAUDE report 12076284 ↗InjuryFeb 15, 2019Break
IT WAS REPORTED THAT DURING A AN ANTERIOR CRUCIATE LIGAMENT SURGERY WHEN INSERTING THE BIOCOMPOSITE SCREW THE TIP BROKE OFF AND WAS LEFT IN THE JOINT. THE SURGEON WAS NOT ABLE TO RETRIEVE THE BROKEN PIECE. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE 29-JAN-2019: THE BROKEN FRAGMENT IS SCREWED TIGHT IN THE JOINT. IT CANNOT MOVE.
Patient problems: Foreign Body In Patient
MAUDE report 8344600 ↗InjuryMay 3, 2018Loose or Intermittent Connection
IT WAS REPORTED THAT THE INITIAL SURGERY WAS ON (B)(6) 2017. PLACE OF THE TUNNEL WAS TIBIALLY SCORED. SCREW LOOSENED AND WAS PULLED OUT BY HAND 3 MONTHS LATER. (NO INFECTION PRESENT) FOLLOW-UP INVESTIGATION: IN (B)(6) 2018 THE PATIENT COMPLAINED ABOUT PAIN. THE SCREW WAS NOTICEABLE UNDER THE SKIN. ACCORDING TO THE SURGEON THE PATIENT WAS COMPLIANT TO THE POST-OP GUIDELINES. THERE WAS AN UNPLANNED SECOND SURGERY WHICH TOOK PLACE ON (B)(6) 2018. THE PATIENT HAD STERILE PUS WHICH WAS TREATED WITH ANTIBIOTICS. ACCORDING TO THE SURGEON THE CULTURES WERE TESTED AND REVEALED NO BACTERIAL INFECTION.
Patient problems: Pain; Purulent Discharge
MAUDE report 7483532 ↗InjuryApr 25, 2017Break
IT WAS REPORTED THAT THE SCREW BROKE DURING INSERTION INTO THE TIBIA. ABOUT 1/3 OF THE SCREW WAS INSERTED WHEN IT BROKE. SURGEON DECIDED NOT TO USE A NEW SCREW BUT TO LEAVE THE BROKEN PART OF SCREW IN THE BONE FOR FIXATION OF THE TENDON. IT WAS REPORTED THAT PATIENT MUST NOT PUT LOAD ON THE LEG FOR 6 WEEKS POST-OP IN ORDER TO ENABLE THE IMPLANT TO INCORPORATE TO THE BONE.
Patient problems: Foreign Body In Patient
MAUDE report 6520185 ↗
Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page