Arthrex® AR-5035TC-09Recalls & reported problems
Arthrex, Inc. · BIO-COMP INTERFSCRW CAN-DELTATPR 9X 35MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867035508), received between Aug 23, 2016 and Jun 19, 2024 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Defective Device1
- Difficult to Insert1
- Failure to Advance1
- Material Separation1
- Patient-Device Incompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Burning Sensation1
- Hypersensitivity/Allergic reaction1
- Rash1
- Swelling1
- Unspecified Infection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 19, 2024Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY THE SCREW THREAD PEELED OFF DURING THE FIRST FEW TURNS AND IT WAS THEREFORE NO LONGER POSSIBLE TO IMPLANT THE SCREW. NO PART OF THE DEVICE BROKE OFF. THE DIAMETER OF THE BONE TUNNEL WAS 8 MM AND THE DIMENSIONS OF THE DEVICE ARE 9 MM X 35 MM. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE (8X28 MM; AR-5028C-08). IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19566915 ↗MalfunctionApr 9, 2024Defective Device; Defective Device
ON 4/19/2022, IT WAS REPORTED BY A DISTRIBUTOR VIA EMAIL THAT AN AR-5035TC-09 BIOCOMPOSITE INTERFERENCE SCREW HAD AN ISSUE. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2022. ADDITIONAL INFORMATION REQUESTED
Patient problems: Insufficient Information
MAUDE report 19071948 ↗MalfunctionApr 8, 2024Failure to Advance; Failure to Advance
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY DURING INSERTION OF THE SCREW THE FIRST THREADS WERE TURNED OFF AND THE INSERTION WAS NO LONGER POSSIBLE. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE AVOE 20-MAR-2024: IT WAS CONFIRMED THAT ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19067308 ↗MalfunctionMar 13, 2023Break; Break
ADDITIONAL INFORMATION RECEIVED ON 2/23/2023: THIS WAS DISCOVERED DURING AN ACL PROCEDURE AND COMPLETED USING ANOTHER AR-5035TC-09 CANNULATED DELTA TAPERED BIOCOMPOSITE INTERFERENCE SCREW. ALL THE BROKEN FRAGMENTS WERE RETRIEVED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information
MAUDE report 16528578 ↗MalfunctionJul 11, 2022Break; Difficult to Insert
IT WAS REPORTED THAT DURING A SURGERY DURING THE FIRST TWO TURNS, THE SCREW THREAD CAME LOOSE AND A CORRECT INSERTION OF THE SCREW WAS NO LONGER POSSIBLE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH DIFFERENT PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14976841 ↗InjuryAug 23, 2016Patient-Device Incompatibility
IT WAS REPORTED BY THE PATIENT ON DAY OF THE REVISION PROCEDURE, (B)(6) 2016, THAT SHE HAD UNDERGONE AN ACL RECONSTRUCTION ON (B)(6) 2016 ON HER RIGHT LEG. ON (B)(6) 2016, THE PATIENT BEGAN TO NOTICE EXTREME SWELLING, KNEE WAS HOT TO THE TOUCH, SHE HAD A SKIN RASH WITH WELTS AND BURNING IN AND AROUND THE INCISION. THE REACTION ALSO SPREAD TO THE LATERAL SIDE OF THE FEMUR AND TIBIA. THE PATIENT HAD AN ER VISIT ON (B)(6) 2016 WHERE BLOODWORK WAS TAKEN. RESULTS WERE NEGATIVE FOR INFECTION. SURGEON DECIDED TO BRING PATIENT IN FOR SECOND SURGERY AND POSSIBLE REMOVAL OF THE IMPLANTS. PATIENT STATES THAT PRIOR TO THE (B)(6) 2016 PROCEDURE, SHE HAD ALLERGY TESTING WHICH SHOWED THAT SHE WAS ALLERGIC TO NICKEL. THE FOLLOWING PARTS WERE IMPLANTED DURING THE ORIGINAL (B)(6) 2016 PROCEDURE: AR-1934BCF LOT 1240957 ((B)(6)), AR-1934BCF LOT 1240946 ((B)(6)), AR-2324BCC LOT 10025466 ((B)(6)) AND AR-5035TC-09 LOT 1314851 ((B)(6)). ON (B)(4) 2016, THE SURGEON REPORTED THAT DURING THE (B)(6) 2016 SECOND SURGERY, NO ARTHREX PRODUCTS WERE EXPLANTED. THE SURGEON PERFORMED A SYNOVECTOMY WITH IRRIGATION & DEBRIDEMENT. SURGEON STATED THAT THE REPAIR LOOKED GOOD. SURGEON STATES THAT THE PATIENT SYMPTOMS APPE
Patient problems: Burning Sensation; Hypersensitivity/Allergic reaction; Rash; Swelling; Unspecified Infection
MAUDE report 5897903 ↗
Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page