Arthrex® AR-6068-25TLRecalls & reported problems
Arthrex, Inc. · 25 DEG TIGHT CURVE LEFT, QUICKPASS LASSO · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00888867038899), received between Apr 20, 2021 and Dec 24, 2025 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Failure to Advance2
- Defective Device1
- Material Frayed1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Insufficient Information4
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 24, 2025Break; Break
ON (B)(6) 2025, IT WAS REPORTED BY A CUSTOMER VIA EMAIL THAT FOR AN AR-6068-25TL, 25°, TIGHT CURVE, LEFT, QUICKPASS LASSO, SUTURE PASSER CURVED END SNAPPED OFF THE SHAFT DURING USE INSIDE THE PATIENT. THIS WAS DETECTED DURING USE IN A LABRAL REPAIR PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS THE BROKEN PIECE WAS RETRIEVED, AND THEY CONTINUED THE SURGERY.
Patient problems: Insufficient Information
MAUDE report 23888223 ↗MalfunctionDec 24, 2025Break; Break
ON 12/03/2025, A FACILITY REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-6068-25TL 25-DEGREE TIGHT LEFT-CURVED QUICKPASS LASSO WAS BEING USED TO PASS THROUGH THE LABRUM. AFTER PASSING THROUGH, THERE WAS SCRAPING. THE DEVICE WAS TAKEN OUT AND INSPECTED. WHEN IT WENT BACK IN, IT SNAPPED. PART OF THE ITEM WAS REMOVED, BUT THE TIP REMAINED INSIDE THE PATIENT'S CAPSULE. THE BROKEN PIECE TOOK OVER TWO HOURS TO RETRIEVE FROM THE PATIENT, WHO SEEMS TO BE NORMAL AS OF NOW.
Patient problems: Insufficient Information
MAUDE report 23887388 ↗MalfunctionSep 16, 2025Material Frayed; Material Frayed
ON 08/26/2025, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA (B)(4) THAT AN AR-6068-25TL QUICKPASS LASSO CABLES ARE FRAYED. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23063068 ↗MalfunctionOct 3, 2024Break; Break
ON 09/13/2024, A SALES REPRESENTATIVE REPORTED VIA SEMS-(B)(4) THAT AN AR-6068-25TL 25° TIGHT CURVE, LEFT, QUICKPASS LASSO HAD THE TIP BREAK OFF IN THE PATIENT'S SHOULDER. THE SURGEON RETRIEVED THE BROKEN PIECE. THE CASE WAS COMPLETED BY FINISHING REPAIRING THE LABRUM AS INTENDED. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2024, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 09/18/2024: THE CASE WAS DELAYED UNTIL THEY OPENED A NEW QUICK PASS AND WAS NOT LONG. THIS WAS DISCOVERED DURING A ANTERIOR LABRAL REPAIR PROCEDURE ON (B)(6) 2024.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20358938 ↗MalfunctionSep 9, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC STABILISATION OF GLENOHUMERAL JOINT THE TIP OF THE LASSO BROKE OFF AS THE SURGEON WAS TRYING TO PASS THE DEVICE THROUGH THE LABRAL TISSUE. THE BREAK WAS WHERE THE TIP IS WELDED TO THE SHAFT OF THE LASSO. SURGEON ADMITTED HE MAY HAVE PUT TOO MUCH PRESSURE ON THE DEVICE WHEN IT WAS CONTACTING THE BONE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM ANOTHER MANUFACTURER (SMITH & NEPHEW ACCUPASS). IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20169624 ↗MalfunctionJun 21, 2024Failure to Advance; Failure to Advance
ON 05/23/2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA (B)(4) THAT AN AR-6068-25TL 25°, TIGHT CURVE, LEFT, QUICKPASS, HAD A LASSO THAT WAS STUCK OR CAUGHT ON SOMETHING ON THE INSIDE. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCEDURE, WITH NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19583784 ↗MalfunctionMay 16, 2024Failure to Advance; Failure to Advance
IT WAS REPORTED ON 10/24/2022 BY A ARTHREX EMPLOYEE VIA (B)(4) THAT AN AR-6068-25TL SUTURE LASSO WIRE STOPPED TURNING THE ROLLER. THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19332154 ↗MalfunctionNov 21, 2023Use of Device Problem; Use of Device Problem
IT WAS REPORT THAT DURING AN ARTHROSCOPIC SUTURING OF THE MENISCUS, THE DEVICE BLOCKED AND THE SURGEON WAS NOT ABLE TO PULL THE SUTURE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18180689 ↗MalfunctionOct 17, 2023Defective Device; Defective Device
ON 25TH SEPTEMBER 2023, IT WAS REPORTED BY ARTHREX EMPLOYEE VIA EMAIL THAT 2 PIECES OF AR-6068-25TR, 25°, TIGHT CURVE, RIGHT, QUICK PASS LASSO AND 2 PIECES OF AR-6068-25TL, 25°, TIGHT CURVE, LEFT, QUICK PASS LASSO WERE NOT FEEDING WIRE. THESE WERE DETECTED AT AN UNSPECIFIED TIME ON (B)(6) 2023.
Patient problems: Insufficient Information
MAUDE report 17951824 ↗MalfunctionApr 28, 2023Break; Break
IT WAS REPORTED ON 03/31/2023 BY A SALES REPRESENTATIVE VIA SEMS, THAT AN AR-6068-25TL SUTURE LASSO TIP BROKE OFF, JUST PROXIMAL TO THE BEND AS THE SURGEON WAS TRYING TO PASS THE LASSO THROUGH THE SHOULDER CAPSULAR TISSUE. CASE INVOLVEMENT, DEVICE BROKE DURING USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16838935 ↗MalfunctionApr 20, 2021Break
IT WAS REPORTED THAT A QUANTITY TWO AR-6068-25TL (LOT: 000399887) BROKE RIGHT AT THE END OF THE SHAFT.
Patient problems: Insufficient Information
MAUDE report 11692665 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page