Arthrex® AR-6425Recalls & reported problems
Arthrex, Inc. · PATIENT EXT TBG W/LUERLOCK CON · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867039315), received between Apr 24, 2025 and Sep 9, 2026 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Excess Flow or Over-Infusion” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Excess Flow or Over-Infusion1
- Increase in Pressure1
- Pressure Problem1
Patient problems
- Compartment syndrome1
- No Clinical Signs, Symptoms or Conditions1
- Swelling/ Edema1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2026Pressure Problem
It was reported that the during a rotator cuff surgery device pumps too much liquid. There was no harm for patient, operator or third party reported. The surgery was finished successfully with the the same pump. Update (b)(6) 2026:Further information has been provided that the surgery was continued with the same tubing and same pump. Aclamp/pressure test has been performed prior to tubing use. The neoprene sleeve did not flatten/compress.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50689807 ↗InjuryDec 12, 2025Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
IT WAS REPORTED THAT DURING LATERAL CLAVICULAR RESECTION AN ERROR MESSAGE "PRESSURE FALL" WAS RECEIVED AND THE PATIENT'S SHOULDER JOINT IS BECAME FULL OF WATER (THE SURGEON WAS NOT ABLE TO SEES ANYTHING IN THE JOINT). THE SUCTION THROUGH THE SHAVER DOES NOT EXTRACT FLUID, WHICH MEANS THAT WATER ONLY GOES INTO THE SHOULDER AND NOT OUT AGAIN. THE STAFF TRIED TO RESOLVE THE PROBLEM, BUT WAS UNSUCCESSFUL. THE SHOULDER BECAME VERY SWOLLEN BECAUSE OF THIS ISSUE AND THIS PART OF THE PROCEDURE WAS TERMINATED. THE REMAINING PART OF THE PROCEDURE (LATERAL CLAVICULAR RESECTION) WAS PERFORMED AS AN OPEN PROCEDURE AND WAS COMPLETED AS PLANNED. ***UPDATE DW 10-DEC-2025: FURTHER INFORMATION WAS PROVIDED THAT AN AR-6420, AR-6425 AND AN AR-6430 TUBING WAS USED DURING THE PROCEDURE BUT THEN DISCARDED. IT WAS FURTHER REPORTED THAT NO ADDITIONAL TREATMENT WAS PERFORMED.
Patient problems: Swelling/ Edema
MAUDE report 23792527 ↗InjuryApr 24, 2025Increase in Pressure; Increase in Pressure
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY THE PUMP AR-6480 (SN: (B)(6) WAS USED WITH THE TUBINGS AR-6420 (LOT: Z4053) AND AR-6425 (LOT: X4080). A CLAMPING TEST CARRIED OUT AFTER FILLING THE TUBE SET. AFTER APPROX. 5 MINUTES WITHOUT ANY ABNORMALITIES IN THE PUMP, THE PUMP SUDDENLY PUMPED AT MAXIMUM FLOW RATE (1200 ML/MIN) AND THEN STOPPED WITH A WARNING. THE SURGICAL ASSISTANT RESTARTED THE PUMP, AND THE PUMP AGAIN PUMPED AT MAXIMUM FLOW RATE WITHOUT STOPPING.. THE TUBE WAS THEN UNHOOKED FROM THE PUMP, THE PRESSURE SENSOR WAS DISCONNECTED AND RECONNECTED, THE SAME TUBE WAS REINSERTED, AND THE PUMP WAS THEN RESTARTED.. THE SURGERY COULDN`T BE COMPLETED SUCCESSFULLY DUE TO A COMPARTMENT SYNDROME WITH SPLITTING OF THE LOWER LEG. A SECOND PROCEDURE IS/WAS REQUIRED. ***UPDATE AVOE 10-APR-2025 IT WAS FURTHER REPORTED THAT THE COMPARTMENT SYNDROME WAS CURRENTLY TREATED WITH RELIEF INCISIONS ON THE LOWER LEG. THE PATIENT IS CURRENTLY BEING TREATED AS AN INPATIENT AND THE SECOND OPERATION HAS NOT YET TAKEN PLACE.
Patient problems: Compartment syndrome
MAUDE report 21902393 ↗
Other Arthroscope devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page