Arthrex® AR-6475Recalls & reported problems
Arthrex, Inc. · CWIII ARTHROSCOPY PUMP · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Arthrex AR-6475 Continuous Wave III arthroscopy pump is used to pump and control fluid pressure during joint surgery. Most of the 54 reports describe the pump building up too much pressure or losing suction, with some units not sounding alarms, screens going blank, buttons not responding, or shutting off mid-case; several mention fluid leaking into the patient's joint with swelling and pain, and 14 injuries were reported. No deaths were reported and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 54
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 00888867039346), received between Dec 9, 2016 and Aug 11, 2026 — classified as 14 injuries, 40 malfunctions.
The most frequently reported specific device problem is “Pressure Problem” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Pressure Problem13
- Increase in Pressure8
- Defective Device5
- No Display/Image3
- Defective Component2
- Device Sensing Problem2
- Infusion or Flow Problem2
- Key or Button Unresponsive/not Working2
Patient problems
- No Clinical Signs, Symptoms or Conditions30
- Insufficient Information6
- Swelling6
- Swelling/ Edema5
- Extravasation3
- Pain3
- No Code Available2
- Unspecified Infection2
Recent adverse event reports
The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2026Electrical Power Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50207293 ↗MalfunctionMar 10, 2025Increase in Pressure; Pressure Problem; Increase in Pressure; Pressure Problem
ON 02/19/2025, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA (B)(4) THAT AN AR-6410 MAIN PUMP TUBING EXPERIENCED AN OVERPRESSURE EVENT. THIS WAS DISCOVERED DURING USE IN A KNEE SURGERY CASE. THE PUMP USED IN THE SURGERY WAS AN AR-6475 CONTINUOUS WAVE III. DURING SURGERY, THE AMOUNT OF WATER INCREASED ABNORMALLY. THE PATIENT'S KNEE BECAME SWOLLEN, AND THE PATIENT WAS UNABLE TO BEND HIS LEG. FINALLY, THE SURGERY WAS CALLED OFF. ADDITIONAL SURGERY IS BEING CONSIDERED AT A LATER DATE. ADDITIONAL INFORMATION REQUESTED.
Patient problems: Swelling/ Edema
MAUDE report 21558095 ↗MalfunctionJan 22, 2025Increase in Pressure; Increase in Pressure
ON 12/27/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-6475 ARTHROSCOPY PUMP DID NOT SET OFF AN ALARM WHEN IT WAS TRYING TO OVERPRESSURE WHEN CLAMPED. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21209481 ↗MalfunctionNov 22, 2024Suction Failure; Suction Failure
ON (B)(6)2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6475 ARTHROSCOPY PUMP WAS NOT GETTING ANY SUCTION. THIS OCCURRED DURING A CASE, WITH NO EFFECT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20759101 ↗MalfunctionJun 19, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE PUMP WAS REPORTED TO BE FAULTY AND WON'T STOP. THE USER REPORTING THAT TURNING IT ON WAS OKAY, HOWEVER THE STOP BUTTON IS NOT WORKING. INTERMITTENT FAILURES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19567539 ↗MalfunctionJun 12, 2024Defective Device; Defective Device
IT WAS REPORTED THAT THE PLIXIGLASS COVER IS DEFECTIVE. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19518789 ↗MalfunctionJun 11, 2024Mechanical Problem; Mechanical Problem
ON 04/25/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-6475 ARTHROSCOPY PUMP ROLLER ASSEMBLY SPINNING EXCESSIVELY WHEN THE PRESSURE IS SET TO 40. 1.1 NO ERROR MESSAGE ON THE SCREEN, BUT THE FACILITY COULD NOT GET THE ROLLER TO STOP SPINNING UNLESS THE PRESSURE WAS SET TO 20. 1.2 THIS WAS DURING A CASE, HOWEVER, THE PATIENT WAS NOT ADVERSELY AFFECTED. THEY FINISHED THE CASE SUCCESSFULLY BY KEEPING THE PRESSURE SET TO 20. THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19511403 ↗MalfunctionJun 10, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE PRESSURE AND FLOW CANNOT BE ADJUSTED AS THE CONTROL PANEL DOES NOT RESPOND. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19498196 ↗MalfunctionMay 23, 2024Intermittent Loss of Power; Intermittent Loss of Power
IT WAS REPORTED ON 09/26/2022 BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-6475 PUMP IS CUTTING OUT IN THE MIDDLE OF SURGERY. THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19387714 ↗MalfunctionMay 15, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE TOUCHPAD/SCREEN OF THE PUMP NO LONGER DISPLAYS ANYTHING. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: Insufficient Information
MAUDE report 19321903 ↗MalfunctionMay 15, 2024Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE PUMP DOES NOT STOP PUMPING. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19321564 ↗MalfunctionMay 6, 2024Pressure Problem; Pressure Problem
ON 01/26/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT A AR-6475 ARTHROSCOPY PUMP IS CAUSING SWELLING, THE PRESSURE READINGS ARE NOT ACCURATE. THIS OCCURRED IN AN UNKNOWN CASE, WITH NO PATIENT EFFECT REPORTED. ADDITIONAL INFORMATION PROVIDED BY THE SALES REPRESENTATIVE VIA PHONE ON 02/01/2022 : THIS EVENT OCCURRED IN A SHOULDER SCOPE. ARTHREX TUBING WAS USED WITH THE PUMP. THERE WAS LITTLE SWELLING THAT WENT DOWN NATURALLY ON ITS OWN.
Patient problems: Insufficient Information
MAUDE report 19248735 ↗MalfunctionMay 2, 2024Pressure Problem; Pressure Problem
ON 04/21/2022, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-6475 CWIII ARTHROSCOPY PUMP HAD AN ISSUE, OPEN TUBING, WHEN CLAMPED DID NOT GIVE PRESSURE FAULT. THE ISSUE OCCURRED IN A CASE AND WAS VERIFIED BY BIO AS AN ISSUE. THIS WAS DISCOVERED DURING USE WITH NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19231003 ↗MalfunctionApr 30, 2024Defective Device; Defective Device
IT WAS REPORTED THAT THE PUMP LOCKING LATCH IS LOOSE, BENT AND NOT CLOSING PROPERLY. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19213448 ↗MalfunctionApr 29, 2024Pressure Problem; Pressure Problem
IT WAS REPORTED THAT DURING A SHOULDER SURGERY THE DEVICE OFTEN DID NOT SWITCH OFF AND CONTINUED TO RUN. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE AVOE 29-SEP-2021: IT WAS CONFIRMED THAT THE ERROR WAS IMMEDIATELY NOTICED AND THEREFORE NO PATIENT INJURY OCCURRED. IT WAS FURTHER CONFIRMED THAT NO CLAMP TEST WAS PERFORMED IN ADVANCE AND THE NEOPRENE SLEEVE WAS COMPRESSED. THE TUBING AR-6420 (LOT: Z1006) WAS USED WITH THE REPORTED PUMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19200385 ↗MalfunctionApr 26, 2024Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE PUMP IS WORKING CONTINOUSLY. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19193260 ↗MalfunctionApr 26, 2024Pressure Problem; Pressure Problem
IT WAS REPORTED THAT THE DEVICE ONLY RUNS AT A PRESSURE OF 60 OR HIGHER. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19190557 ↗MalfunctionApr 26, 2024Increased Pump Speed; Increased Pump Speed
ON 04/03/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-6475 ARTHROSCOPY PUMP WAS RUNNING AT A FASTER SPEED THAN INTENDED. THIS OCCURRED DURING A CASE WHERE THERE WERE NO ERROR MESSAGES CORRESPONDING TO THE PROBLEM. THE SPEED OF THE PUMP WAS DESCRIBED AS "REALLY FAST," EVEN WITH THE CLAMP OFF, AND IT WAS CONTINUOUSLY RUNNING.
Patient problems: Insufficient Information
MAUDE report 19190436 ↗MalfunctionFeb 7, 2024Pressure Problem; Pressure Problem
IT WAS REPORTED THAT THE DEVICE IS DEFECTIVE, IT IS SPINNING PERMANENTLY. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED. UPDATE, 04-MAR-2021 -SAC. RECEIVED FURTHER INFORMATION THAT THE PROBLEM WAS NOTICED AFTER THE SURGERY. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18657549 ↗MalfunctionJan 16, 2024Increase in Pressure; Increase in Pressure
ON 6/22/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-6475 WAVE III PUMP, DESPITE BEING SET TO NORMAL, THE FLUID WAS BEING PUMPED INTO THE JOINT AT AN EXPONENTIAL RATE AND COULD NOT BE REDUCED. THIS WAS DISCOVERED DURING A KNEE ARTHROSCOPY PROCEDURE. ADDITIONAL INFORMATION REQUESTED.
Patient problems: Swelling/ Edema
MAUDE report 18516772 ↗MalfunctionJan 10, 2024Pressure Problem; Pressure Problem
ON 3/27/2023, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS THAT AN AR-6475 PUMP HAD LESS THAN ADEQUATE PRESSURE. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18486712 ↗MalfunctionDec 29, 2023No Display/Image; No Display/Image
ON 02/23/2022 IT WAS REPORTED BY A DISTRIBUTOR VIA SEMS THAT AN AR-6475 PUMP SCREEN DOES NOT WORK. THIS WAS DISCOVERED DURING A PROCEDURE, WITH NO PATIENT HARM. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18418439 ↗MalfunctionDec 29, 2023Pressure Problem; Pressure Problem
ON 01/19/2022 IT WAS REPORTED BY A ARTHREX REPRESENTATIVE VIA SEMS THAT AN AR-6475 PUMP DISPLAYS PRESSURE FAULT WHEN IN USE. THIS WAS DISCOVERED DURING USE IN A PROCEDURE. NO PATIENT HARM. TS: CHECKED WITH A MANOMETER AND NOTICED THAT THE PRESSURE INDICATED AND THE MEASURED PRESSURE IS OFF. REQUESTED ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18418178 ↗MalfunctionDec 28, 2023Pressure Problem; Pressure Problem
ON 12/20/2021, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-6475 PUMP IS EXPERIENCING PRESSURE ISSUES. THIS WAS DISCOVERED DURING A PROCEDURE.
Patient problems: Insufficient Information
MAUDE report 18413411 ↗MalfunctionDec 11, 2023Increase in Pressure; Increase in Pressure
IT WAS REPORTED DURING A SHOULDER REPAIR THE PUMP STARTED RUNNING REALLY FAST WITHOUT ANY CHANGE TO THE PUMP. THE REP STOPPED THE UNIT BEFORE ANY INJURY OCCURRED. IT WAS REPLACED AND THE CASE WAS COMPLETED WITH NO FURTHER ISSUES AND THE PATIENT IS FINE TO DATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18300383 ↗
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.