Arthrex® AR-8724-10Recalls & reported problems
Arthrex, Inc. · LOW PROFILE SCREW, 2.4 X 10MM, CORTEX · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867045071), received between Jan 7, 2020 and Dec 2, 2020 — classified as 3 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Difficult to Advance1
- Positioning Failure1
Patient problems
- Failure of Implant2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 2, 2020Positioning Failure; Difficult to Advance
IT WAS REPORTED CONSTRUCTIVE FEEDBACK WAS RECEIVED FROM A SURGEON REGARDING THE 2.4 CORTICAL SCREW (AR-8724) IN THE CFS SET. THE SURGEON WAS TREATING A FRACTURE ACROSS THE MIDFOOT AND SECOND METATARSAL, AND WHAT THEY THOUGHT WOULD BE A 20 MIN CASE, TURNED INTO TWO HOURS AND THE SURGEON HAVING TO BAIL TO ANOTHER SYSTEM. THE AMOUNT OF AXIAL LOAD AND FORCE TO GET THE 2.4 CORTICAL SCREWS TO ADVANCE WAS SIGNIFICANT, AND THE SURGEON COULD NOT GET THE 2.4 CORTICAL SCREWS TO ADVANCE WITHOUT LOSING THEIR REDUCTION AS A RESULT OF HAVING TO REALLY LEAN INTO THE SCREWDRIVER TO ADVANCE THE SCREW. THE SURGEON TRIED TO MAKE IT WORK AND WANTED TO STICK WITH ARTHREX FOR THEIR IMPLANT, BUT HAD TO BAIL. ONCE THE SURGEON WENT TO ANOTHER MANUFACTURER'S SYSTEM, THE SCREWS WENT IN WITHOUT ANY ISSUE. THE SURGEON MENTIONED THEY NOTICED THE SCREW TIPS HAVE MORE OF A SNUB NOSE AND THEY ARE NOT SURE IF THE SELF-TAPPING FLUTES WHERE GRABBING. IT WAS CONFIRMED THAT THE CORRECT DRILL BIT WAS USED. ADDITIONAL INFORMATION OBTAINED 11/11/20: DATE OF THE SURGERY WAS (B)(6) 2020. THE PART NUMBERS OF THE ARTHREX 2.4 SCREWS ATTEMPTED WERE: AR-8724-08, AR-8724-10 AND AR-8724-12. THE SPECIFIC LOT NUMBERS OF THE SCREWS AR
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant
MAUDE report 10934397 ↗InjuryJan 9, 2020Device Dislodged or Dislocated
IT WAS REPORTED THAT THE SURGEON HAD TO REMOVE THE WRIST SPANNING PLATE, AR-8916SPN, BECAUSE THE SCREWS WERE NOT LOCKED INTO THE DISTAL HOLES. NO FURTHER INFORMATION, ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED 12/18/2019: CASE DATE OF IMPLANTATION: (B)(6) 2019, CASE DATE OF EXPLANTATION: NOT YET DETERMINED. THE FOLLOWING SCREWS WERE ALSO EXPLANTED DURING THE REVISION: AR-8724-10 LOT: 1058931, AR-8724V-10 LOT: W654331, AR-8724V-16 LOT: W630941, AR-8935-12 LOT: 10220124, AR-8935L-12 LOT: 10248943, AR-8935L-12 LOT: 10257929. NOTE: THE SPECIFIC DATE OF THE SECOND SURGERY (EXPLANT) IS UNKNOWN, HOWEVER IT IS KNOWN THAT THE DATE WAS PRIOR TO (B)(6) 2019. FOR REPORTING PURPOSES (B)(6) 2019 HAS BEEN ENTERED AS THE DATE OF EVENT.
Patient problems: Failure of Implant
MAUDE report 9570004 ↗InjuryJan 7, 2020Device Dislodged or Dislocated
IT WAS REPORTED THAT THE SURGEON HAD TO REMOVE THE WRIST SPANNING PLATE, AR-8916SPN, BECAUSE THE SCREWS WERE BACKING OUT OF THE DISTAL HOLES. NO FURTHER INFORMATION, ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED 12/18/19: CASE DATE OF IMPLANTATION: (B)(6) 2019, CASE DATE OF EXPLANTATION: (B)(6) 2019. THE FOLLOWING SCREWS WERE ALSO EXPLANTED DURING THE REVISION: AR-8916SPN LOT: 105641930, AR-8724-10 LOT: W646301, AR-8724V-10 LOT: 2610171, AR-8724V-12 LOT: W657601, AR-8935-14 LOT: 10279504, AR-8935L-12 LOT: 10257929, AR-8935L-14 LOT: 10277889.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9557183 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page