Arthrex® AR-8724-12Recalls & reported problems
Arthrex, Inc. · LOW PROFILE SCREW, 2.4 X 12MM, CORTEX · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867045095), received between Mar 22, 2017 and Dec 2, 2020 — classified as 3 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Break1
- Difficult to Advance1
Patient problems
- Failure of Implant1
- No Clinical Signs, Symptoms or Conditions1
- Pain1
- Swelling1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 2, 2020Difficult to Advance
IT WAS REPORTED CONSTRUCTIVE FEEDBACK WAS RECEIVED FROM A SURGEON REGARDING THE 2.4 CORTICAL SCREW (AR-8724) IN THE CFS SET. THE SURGEON WAS TREATING A FRACTURE ACROSS THE MIDFOOT AND SECOND METATARSAL, AND WHAT THEY THOUGHT WOULD BE A 20 MIN CASE, TURNED INTO TWO HOURS AND THE SURGEON HAVING TO BAIL TO ANOTHER SYSTEM. THE AMOUNT OF AXIAL LOAD AND FORCE TO GET THE 2.4 CORTICAL SCREWS TO ADVANCE WAS SIGNIFICANT, AND THE SURGEON COULD NOT GET THE 2.4 CORTICAL SCREWS TO ADVANCE WITHOUT LOSING THEIR REDUCTION AS A RESULT OF HAVING TO REALLY LEAN INTO THE SCREWDRIVER TO ADVANCE THE SCREW. THE SURGEON TRIED TO MAKE IT WORK AND WANTED TO STICK WITH ARTHREX FOR THEIR IMPLANT, BUT HAD TO BAIL. ONCE THE SURGEON WENT TO ANOTHER MANUFACTURER'S SYSTEM, THE SCREWS WENT IN WITHOUT ANY ISSUE. THE SURGEON MENTIONED THEY NOTICED THE SCREW TIPS HAVE MORE OF A SNUB NOSE AND THEY ARE NOT SURE IF THE SELF-TAPPING FLUTES WHERE GRABBING. IT WAS CONFIRMED THAT THE CORRECT DRILL BIT WAS USED. ADDITIONAL INFORMATION OBTAINED 11/11/20: DATE OF THE SURGERY WAS (B)(6) 2020. THE PART NUMBERS OF THE ARTHREX 2.4 SCREWS ATTEMPTED WERE: AR-8724-08, AR-8724-10 AND AR-8724-12. THE SPECIFIC LOT NUMBERS OF THE SCREWS AR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10934510 ↗InjuryJan 9, 2020Device Dislodged or Dislocated
IT WAS REPORTED THAT THE SURGEON HAD TO REMOVE THE WRIST SPANNING PLATE, AR-8916SPN, BECAUSE THE SCREWS WERE BACKING OUT OF THE DISTAL HOLES. NO FURTHER INFORMATION, ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED (B)(6) 2019: CASE DATE OF IMPLANTATION: (B)(6) 2019. CASE DATE OF EXPLANTATION: NOT YET DETERMINED THE FOLLOWING SCREWS WERE ALSO EXPLANTED DURING THE REVISION: AR-8724-12 LOT: 2542021. AR-8724V-12 LOT: 2605841. AR-8724V-18 LOT: 2605831. AR-8935-14 LOT: 10176201. AR-8935-14 LOT:10249431. THE SPECIFIC DATE OF THE SECOND SURGERY (EXPLANT) IS UNKNOWN, HOWEVER IT IS KNOWN THAT THE DATE WAS PRIOR TO (B)(6) 2019. FOR REPORTING PURPOSES (B)(6) 2019 HAS BEEN ENTERED AS THE DATE OF EVENT.
Patient problems: Failure of Implant
MAUDE report 9569883 ↗InjuryMar 22, 2017Break; Device Dislodged or Dislocated
IT WAS REPORTED ON (B)(6) 2016 BY A PATIENT VIA THE ARTHREX WEBSITE THAT THE PATIENT HAD UNDERGONE FOOT SURGERY ON (B)(6) 2015. SURGERY WAS LEFT PROXIMAL BUNIONECTOMY IN CONJUNCTION WITH 2ND MTP RECONSTRUCTION FOLLOWING DISLOCATION AND PRIOR PODIATRIC MANAGEMENT. DURING THE SURGERY, THE FOLLOWING PARTS WERE IMPLANTED: AR13200M-055R ((B)(6)), AR-8724-16 ((B)(6)), AR-8724-12 ((B)(6)), AR-8724-14 ((B)(6)), AR-8724-18 ((B)(6)), AR-8930-12 ((B)(6)) AND AR-8724-12PT ((B)(6)). ON (B)(6) 2016, PATIENT FOLLOWED UP WITH SURGEON DUE TO RECENT ONSET OF LEFT MID-FOOT PAIN AND SWELLING. MEDICAL RECORDS NOTED THAT UPON EXAMINATION OF THE LEFT FOOT THERE WAS PAIN AT THE 1ST TMT AND PROXIMAL OSTEOTOMY SITE WITH VISIBLE SWELLING. THERE WAS NO ERYTHEMA, DRAINAGE OR OTHER EVIDENCE OF INFECTION. X-RAY RESULTS NOTED THERE IS BROKEN HARDWARE (PROXIMAL BUNION PLATE - AR-13200M-055R) WITH QUESTION OF NON-UNION. SURGEON HAS RECOMMENDED A PROXIMAL LEFT BUNION HARDWARE REMOVAL WITH A PLAN TO PROCEED WITH A LAPIDUS FUSION IF IN FACT THE OSTEOTOMY IS INCOMPLETE. PATIENT REVISION IS TENTATIVELY SCHEDULED FOR EARLY 2017 DUE TO PATIENT LACK OF AVAILABLE TIME OFF FROM WORK. ADDITIONAL INFORMATION RECEIVED 2/28/17:
Patient problems: Pain; Swelling
MAUDE report 6424833 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page