Skip to content

Arthrex® AR-8724-20Recalls & reported problems

Arthrex, Inc. · LOW PROFILE SCREW, 2.4 X 20MM, CORTEX · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867045170), received between May 13, 2024 and May 13, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Installation-Related Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Installation-Related Problem1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2024Installation-Related Problem; Installation-Related Problem

    ON 1/5/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8724-20 LOW PROFILE SCREW DID NOT LOCK AND WENT RIGHT THROUGH AN AR-8952MT-03 LOW PROFILE T-PLATE DURING INSERTION. SURGEON REMOVED THE SCREW INTACT, AS IT DID NOT BREAK. TWO 2.4 X 20 KREULOCKS SCREWS WERE IMPLANTED PROXIMALLY, A 3.0MM CORTICAL SCREW IN THE OBLONG SO THAT IT WOULD GET PURCHASE IN THE PLATE AND A 4TH SCREW DISTAL WERE USED TO COMPLETE THE CASE. THIS WAS DISCOVERED DURING AN ISOLATED 2ND TMT FUSION PROCEDURE ON (B)(6) 2022. ADDITIONAL INFORMATION RECEIVED ON 1/6/2023: PER THE SALES REPRESENTATIVE, ONCE THE AR-8724-20 LOW PROFILE SCREW WENT RIGHT THROUGH THE PLATE, NOT LOCKING, THE SURGEON REMOVED IT AND IMPLANTED IT IN THE MOST DISTAL, FOURTH HOLE.

    Patient problems: Insufficient Information

    MAUDE report 19296822 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page