Skip to content

Arthrex® AR-8724V-28Recalls & reported problems

Arthrex, Inc. · 2.4 X 28MM VAL SCREW, Ti · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867045668), received between Jun 24, 2025 and Jun 24, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Positioning Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 24, 2025Positioning Failure; Positioning Failure

    ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT TWO 2.4 MM VAL SCREWS WENT THROUGH AN AR-8916VSR-05 VOLAR DISTAL RADIUS PLATE DURING INSERTION. THE ENTIRE CONSTRUCT HAD TO BE REMOVED IN ORDER TO CHANGE PLATES. THE PLATES WERE CHANGED TO A WIDE PLATE AND THE SURGEON RE-DRILLED AND RE-SIZED THE SCREWS. THIS WAS DISCOVERED DURING A DISTAL RADIUS FRACTURE PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION WAS RECEIVED ON 4/15/2025: NOTHING BROKE INSIDE THE PATIENT. THE CASE WAS SUCCESSFULLY COMPLETED WITH A NEW PLATE AND SCREWS. HOWEVER, THE PROCEDURE WAS DELAYED BY APPROXIMATELY FORTY-FIVE MINUTES, REQUIRING THE ADMINISTRATION OF ADDITIONAL ANESTHESIA. ADDITIONAL INFORMATION RECEIVED ON 6/3/2025: A TOTAL OF FIVE PRODUCTS WERE RECEIVED. AN AR-8916VSR-05 VOLAR DISTAL RADIUS PLATE, AN AR-8724V-20 LOW PROFILE VA LOCKING SCREW, AN AR-8724V-22 LOW PROFILE VA LOCKING SCREW, AN AR-8724V-28 LOW PROFILE VA LOCKING SCREW, AND AN AR-8935-18 LOW PROFILE NON-LOCKING SCREW.

    Patient problems: Insufficient Information

    MAUDE report 22318686 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page