Skip to content

Arthrex® AR-8825BRecalls & reported problems

Arthrex, Inc. · SUTURE ANCHOR, MINI BIO-PUSHLOCK · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00888867046641), received between Dec 15, 2023 and Dec 23, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Material Twisted/Bent1
  • Physical Resistance/Sticking1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Insufficient Information2

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 23, 2025Break; Break

    ON 12/04/2025, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA (B)(4) THAT AN AR-8825B MINI BIO-PUSHLOCK'S ANCHOR BROKE DURING INSERTION. THIS WAS DISCOVERED DURING A TFCC PROCEDURE WITH NO PATIENT HARM. THE CASE WAS COMPLETED WITH ANOTHER ANCHOR. THE PATIENT¿S BONE QUALITY WAS GOOD, AND NO FRAGMENTS REMAINED IN THE PATIENT. ADDITIONAL INFORMATION PROVIDED 11-DEC-2025: IT WAS FURTHER REPORTED THAT THE CASE WAS COMPLETED WITH A NEW ANCHOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23881700 ↗
  • MalfunctionAug 1, 2025Break; Break

    ON (B)(6) 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA EMAIL THAT FOR AN AR-8825B, SUTURE ANCHOR, MINI BIO-PUSHLOCK¿ 2.5 MM X 8 MM, WHEN THE SURGEON INSERTED THE ANCHOR, THE SCREW BROKE OFF EASILY AND THE EYELET FELL OUT EASILY. THIS WAS DETECTED DURING A TFCC REPAIR ON (B)(6) 2025. THE CASE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER AR-8825B. THERE WAS NO PATIENT HARM. ADDITIONAL INFORMATION: THERE WAS A DELAY OF ABOUT 10 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22690000 ↗
  • MalfunctionJul 17, 2025Break; Break

    ON 24TH JUNE 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA EMAIL THAT FOR AN AR-8825B, SUTURE ANCHOR, MINI BIO-PUSHLOCK¿ 2.5 MM X 8 MM, THE ANCHOR'S TIP WAS BROKEN DURING INSERTION. THIS WAS DETECTED DURING A TFCC REPAIR PROCEDURE ON (B)(6) 2025. THE CASE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER AR-8825B. THERE WAS NO PATIENT HARM. ADDITIONAL INFORMATION: BEFORE INSERTING THE ANCHOR, THE SURGEON SUSPECTED THAT THE ANCHOR TIP WAS LOOSELY ATTACHED TO THE METAL PART. THERE WAS A DELAY OF ABOUT 10 MINUTES FOR THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22535248 ↗
  • MalfunctionMay 9, 2025Break; Material Twisted/Bent; Break; Material Twisted/Bent

    ADDITIONAL INFORMATION RECEIVED ON 5/13/2025: THE ANCHOR TIP BENT, BUT NOTHING BROKE IN THE PATIENT. THE CASE WAS DELAYED FIVE MINUTES WITHOUT ADDITIONAL ANESTHESIA.

    Patient problems: Insufficient Information

    MAUDE report 21996010 ↗
  • MalfunctionAug 8, 2024Break; Break

    ON 7/26/2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA EMAIL THAT AN AR-8825B MINI BIO-PUSHLOCK EYELET CAME OFF THE INSERTER. THE CASE WAS COMPLETED USING A NEW AR-8825B MINI BIO-PUSHLOCK. THIS WAS DISCOVERED DURING A TFCC REPAIR PROCEDURE ON (B)(6) 2024. THE CASE WAS NOT DELAYED, AND THE PATIENT SUFFERED NO HARM. ADDITIONAL INFORMATION RECEIVED ON 8/1/2024: PUSHLOCK EYELET BROKE OFF IN THE PATIENT AND IT WAS SUCCESSFULLY RETRIEVED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19937791 ↗
  • MalfunctionFeb 1, 2024Break; Break

    ON 01/11/2024, IT WAS REPORTED BY A DISTRIBUTOR VIA EMAIL THAT AN AR-8825B MINI PUSH LOCK BROKE DURING INSERTION. ALL FRAGMENTS REMOVED. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18624535 ↗
  • MalfunctionFeb 1, 2024Break; Break

    ON 01/11/2024, IT WAS REPORTED BY A DISTRIBUTOR VIA EMAIL THAT AN AR-8825B MINI PUSH LOCK BROKE DURING INSERTION. ALL FRAGMENTS REMOVED. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18624517 ↗
  • MalfunctionJan 25, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 01/02/2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA EMAIL THAT AN AR-8825B SUTURE ANCHOR WOULD NOT SEPERATE FROM THE SCREW ONCE INSERTED. THE CASE WAS COMPLETED USING ANOTHER AR-8825B WITH UNKNOWN LOT NUMBERS.

    Patient problems: Insufficient Information

    MAUDE report 18581148 ↗
  • MalfunctionDec 15, 2023Separation Failure; Separation Failure

    IT WAS REPORTED THAT DURING A HAND SURGERY THE ANCHOR COULD NOT BE DETACHED FROM THE DRIVER. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18340857 ↗

Other Fastener, fixation, biodegradable, soft tissue devices

All Fastener, fixation, biodegradable, soft tissue devices →