Skip to content

Arthrex® AR-8827-18Recalls & reported problems

Arthrex, Inc. · LO-PRO SCRW TM SS 2.7X 18MMCORTEX · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867046771), received between Sep 6, 2024 and Apr 22, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Defective Device1

Patient problems

  • Pain2
  • Bone Fracture(s)1
  • Failure of Implant1
  • Foreign Body Reaction1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 22, 2026Patient Device Interaction Problem

    ON 02-JAN-2025, A PATIENT CALLED IN REPORTING THAT AFTER UNDERGOING AN ANKLE PROCEDURE ON (B)(6) 2025, SHE HAD EXPERIENCED A REACTION, ALONG WITH EXPERIENCING HER SKIN CHANGING, AND PAIN NEAR HER ANKLE. ADDITIONAL INFORMATION WAS PROVIDED ON 24-MAR-2026 WHERE THE PATIENT REPORTED THAT SHE UNDERWENT A LEFT ANKLE PROCEDURE FOR A DISPLACED BIMALLEOLAR FRACTURE OF THE LOWER LEG ON (B)(6) 2025 WHERE AN AR-8943DL-04 LOCKING DISTAL FIBULA PLATE, AR-8827-18 LOW PROFILE SCREW¿, 2.7 X 18 MM, (3) AR-8827CL-12 KREULOCK¿ COMPRESSION SCREW, SS, 2.7 MM X 12 MM, AR-8827CL-14 KREULOCK¿ COMPRESSION SCREW, SS, 2.7 MM X 14 MM, (2) AR-8827CL-16 KREULOCK¿ COMPRESSION SCREW, SS, 2.7 MM X 16 MM, (3) AR-8835CL-12 KREULOCK¿ COMPRESSION SCREW, SS, 3.5 MM X 12 MM, AR-8835-14 LOW PROFILE SCREW¿, SS, 3.5 X 14 MM, CORTICAL, AND AN AR-8925SS SYNDESMOSIS TIGHTROPE® XP, STAINLESS STEEL WERE IMPLANTED. THE PATIENT BEGAN TO EXPERIENCE THE FOLLOWING SYMPTOMS ON (B)(6) 2025; RUNNY NOSE, DIARRHEA, BLURRED VISION, SENSITIVITY TO THE ANKLE, BOTTOM OF THE TOES ACHE, CHANGES IN SKIN, BLOOD PRESSURE GOING UP AND DOWN, BACK OF HEELS ACHE, ITCHY ANKLE; THAT IS BEING TREATED WITH A CREAM, SHARP PAIN IN THE ANKLE, RIGHT EAR BLOC

    Patient problems: Foreign Body Reaction; Pain

    MAUDE report 24956989 ↗
  • InjuryJul 8, 2025Defective Device; Adverse Event Without Identified Device or Use Problem; Defective Device; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION WAS RECEIVED ON 07/08/2025: THE AR-8973DS FIBULOCK IMPLANT SYSTEM HAD THE END CAP IN THE KIT INSERTED INTO THE FIBULOCK NAIL DURING THE INITIAL PROCEDURE. DURING THE ROH/TTC REVISION, THE END CAP WAS REMOVED BEFORE THE EXTRACTION OF THE BROKEN FIBULAR NAIL. THE END CAP FROM THE AR-8973DS FIBULOCK IMPLANT SYSTEM DID NOT BREAK DURING THE PROCEDURE.

    Patient problems: Bone Fracture(s); Failure of Implant

    MAUDE report 22439189 ↗
  • InjurySep 6, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A PATIENT UNDERWENT REVISION SURGERY DUE TO PAIN. ON (B)(6) 2023, THE PATIENT UNDERWENT SURGERY FOR AN ANKLE FRACTURE AND SYNDESMOSIS REPAIR. DURING THE PROCEDURE, TWO AR-8730-36PT LOW PROFILE SCREW, TI, 3.0 MM X 36 MM, CANCELLOUS, PARTIALLY THREADED, AN AR-8827-18 LOW PROFILE SCREW, 2.7 X 18 MM, CORTEX, AN AR-8827-22 LOW PROFILE SCREW, 2.7 X 22 MM, CORTEX, AN AR-8973DS FIBULOCK IMPLANT SYSTEM, AN AR-8973L-38-180 ARTHREX FIBULOCK NAIL, 3.8 X 180 MM LEFT, AND AN AR-8925SS SYNDESMOSIS TIGHTROPE XP, STAINLESS STEEL WERE IMPLANTED. POST-OPERATIVE, THE PATIENT STARTED TO EXPERIENCE PAIN. ON (B)(6) 2024, REVISION SURGERY WAS PERFORMED TO REMOVE ALL THE IMPLANTS EXCEPT FOR THE AR-8973L-38-180 ARTHREX FIBULOCK NAIL. DURING THE REMOVAL OF THE IMPLANTS, THE DISTAL END OF THE EXTRACTION TOOL OF AN AR-8973RK FIBULOCK REMOVAL KIT BROKE OFF AFTER ROUGHLY 8 MALLET STRIKES. ANOTHER AR-8973RK FIBULOCK REMOVAL KIT WAS OPENED, AND THE SAME THING OCCURRED. THE EXTRACTION TOOLS BROKE, BUT NO PIECES WERE IN THE PATIENT. THE SURGEON ELECTED TO CLOSE AND FINISH THE CASE WITH THE NAIL STILL IMPLANTED AND FELT CONFIDENT THAT THE REMOVE

    Patient problems: Pain

    MAUDE report 20156936 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page