Arthrex® AR-8827L-18Recalls & reported problems

Arthrex, Inc. · LO-PRO LOCK SCRW,SS 2.7X 18MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00888867047075), received between Feb 13, 2018 and Mar 4, 2026 — classified as 6 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 4 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Device Dislodged or Dislocated1
  • Fitting Problem1
  • Patient Device Interaction Problem1
  • Positioning Failure1
  • Use of Device Problem1

Patient problems

  • Implant Pain4
  • Insufficient Information3
  • Discomfort1
  • Pain1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 4, 2026Patient Device Interaction Problem

    ON (B)(6) 2025, A HEALTH CARE PROFESSIONAL REPORTED VIA EMAIL THAT A CLAVICLE PLATE IMPLANTED IN A PATIENT EXPERIENCED A FAILURE. ACCORDING TO THE REPORT, THE FAILURE OCCURRED APPROXIMATELY SIX MONTHS POST IMPLANTATION, AND NO TRAUMA WAS NOTED DURING THAT PERIOD. RADIOGRAPHIC IMAGES OF THE PLATE FAILURE WERE PROVIDED FOR REVIEW. ADDITIONAL INFORMATION WAS RECEIVED ON 24-OCT-2025: THE ORIGINAL SURGERY WAS PERFORMED ON (B)(6) 2025, INVOLVING AN ORIF OF THE RIGHT CLAVICLE. THERE WERE NO INTRAOPERATIVE COMPLICATIONS, INCLUDING NO CASE DELAYS, ADDED ANESTHESIA TIME, OR ADVERSE EVENTS. THE SPECIFIC ARTHREX PART NUMBERS USED DURING THE PROCEDURE WERE NOT PROVIDED. THE IMPLANT FAILURE WAS IDENTIFIED SIX MONTHS POSTOPERATIVELY, OCCURRING AT THE JUNCTION BETWEEN THE SCREW HOLES. A REVISION SURGERY HAS BEEN SCHEDULED. NO ADDITIONAL INFORMATION REGARDING THE EVENT IS AVAILABLE AT THIS TIME. ADDITIONAL INFORMATION WAS RECEIVED ON 10/30/2025: THE SPECIFIC ARTHREX PART NUMBERS USED DURING THE INITIAL PROCEDURE HAVE NOT BEEN PROVIDED AT THIS TIME. A REVISION SURGERY IS SCHEDULED TO TAKE PLACE IN TWO WEEKS. THE PATIENT DID NOT REPORT ANY PRIOR SYMPTOMS OR EXPERIENCE A POPPING SENSATION AT THE SITE.

    Patient problems: Insufficient Information

    MAUDE report 24506280 ↗
  • MalfunctionMay 9, 2024Use of Device Problem; Use of Device Problem

    ON 09/23/2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT (3) AR-8827L-18 LOW PROFILE LOCKING SCREW WOULD NOT ENGAGE INTO THE PLATE. THIS WAS DISCOVERED DURING USE IN A CLAVICLE FRACTURE ON (B)(6) 2021. THE CASE WAS COMPLETED BY PLACING ONLY 2 OUT OF THE 5 SCREWS INTO THE PLATE, LEAVING 3 OF THE HOLES EMPTY.

    Patient problems: Insufficient Information

    MAUDE report 19276962 ↗
  • InjuryFeb 16, 2024Break; Break

    ON (B)(6) 2024, AN FDA MEDWATCH NOTIFICATION WAS RECEIVED VIA EMAIL. A FACILITY REPRESENTATIVE REPORTED THAT AN AR-2685DR-10 DISTAL CLAVICLE PLATE BROKE EIGHT WEEKS POST-OPERATIVE. THE PATIENT UNDERWENT SURGERY ON (B)(6) 2023 FOR AN OPEN REDUCTION WITH INTERNAL FIXATION OF THE RIGHT COMMINUTED MID-SHIFT CLAVICLE FRACTURE. EIGHT WEEKS LATER, THE PATIENT RETURNED TO THE CLINIC, EXPERIENCING PAINFUL POPS IN THE SHOULDER. ON (B)(6) 2023, THE PATIENT UNDERWENT REVISION SURGERY TO REMOVE THE AR-2685DR-10 DISTAL CLAVICLE PLATE. THE REVISION SURGERY WAS A REVISION OPEN REDUCTION AND INTERNAL FIXATION OF THE RIGHT CLAVICLE FRACTURE WITH ILIAC BONE GRAFT AND RIGHT CLAVICLE HARDWARE REMOVAL. NO FURTHER INFORMATION WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON 1/23/2024: DURING THE ORIGINAL PROCEDURE ON (B)(6) 2023, AN AR-2685DR-10 DISTAL CLAVICLE PLATE, AN AR-8827L-20 LOW PROFILE LOCKING SCREW, (4) AR-8827L-18 LOW PROFILE LOCKING SCREW, (2) AR-8835-16 LOW PROFILE SCREW, AN AR-8835-18 LOW PROFILE SCREW, (2) AR-8835L-18 LOW PROFILE LOCKING SCREW, AND AN AR-8835L-16 LOW PROFILE LOCKING SCREW WERE IMPLANTED. DURING THE REVISION SURGERY ON (B)(6) 2023, ALL THE IMPLANTS WERE SUCCESSFULLY REMOVED

    Patient problems: Implant Pain

    MAUDE report 18725285 ↗
  • InjuryFeb 16, 2024Break; Break

    ON 1/18/2024, AN FDA MEDWATCH NOTIFICATION WAS RECEIVED VIA EMAIL. A FACILITY REPRESENTATIVE REPORTED THAT AN AR-2685DR-10 DISTAL CLAVICLE PLATE BROKE EIGHT WEEKS POST-OPERATIVE. THE PATIENT UNDERWENT SURGERY ON (B)(6) 2023 FOR AN OPEN REDUCTION WITH INTERNAL FIXATION OF THE RIGHT COMMINUTED MID-SHIFT CLAVICLE FRACTURE. EIGHT WEEKS LATER, THE PATIENT RETURNED TO THE CLINIC, EXPERIENCING PAINFUL POPS IN THE SHOULDER. ON (B)(6) 2023, THE PATIENT UNDERWENT REVISION SURGERY TO REMOVE THE AR-2685DR-10 DISTAL CLAVICLE PLATE. THE REVISION SURGERY WAS A REVISION OPEN REDUCTION AND INTERNAL FIXATION OF THE RIGHT CLAVICLE FRACTURE WITH ILIAC BONE GRAFT AND RIGHT CLAVICLE HARDWARE REMOVAL. NO FURTHER INFORMATION WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2024: DURING THE ORIGINAL PROCEDURE ON (B)(6) 2023, AN AR-2685DR-10 DISTAL CLAVICLE PLATE, AN AR-8827L-20 LOW PROFILE LOCKING SCREW, (4) AR-8827L-18 LOW PROFILE LOCKING SCREW, (2) AR-8835-16 LOW PROFILE SCREW, AN AR-8835-18 LOW PROFILE SCREW, (2) AR-8835L-18 LOW PROFILE LOCKING SCREW, AND AN AR-8835L-16 LOW PROFILE LOCKING SCREW WERE IMPLANTED. DURING THE REVISION SURGERY ON 6/30/2023, ALL THE IMPLANTS WERE SUCCESSFULLY REMOVED.

    Patient problems: Implant Pain

    MAUDE report 18725224 ↗
  • InjuryFeb 16, 2024Break; Break

    ON 1/18/2024, AN FDA MEDWATCH NOTIFICATION WAS RECEIVED VIA EMAIL. A FACILITY REPRESENTATIVE REPORTED THAT AN AR-2685DR-10 DISTAL CLAVICLE PLATE BROKE EIGHT WEEKS POST-OPERATIVE. THE PATIENT UNDERWENT SURGERY ON (B)(6) 2023 FOR AN OPEN REDUCTION WITH INTERNAL FIXATION OF THE RIGHT COMMINUTED MID-SHIFT CLAVICLE FRACTURE. EIGHT WEEKS LATER, THE PATIENT RETURNED TO THE CLINIC, EXPERIENCING PAINFUL POPS IN THE SHOULDER. ON (B)(6) 2023, THE PATIENT UNDERWENT REVISION SURGERY TO REMOVE THE AR-2685DR-10 DISTAL CLAVICLE PLATE. THE REVISION SURGERY WAS A REVISION OPEN REDUCTION AND INTERNAL FIXATION OF THE RIGHT CLAVICLE FRACTURE WITH ILIAC BONE GRAFT AND RIGHT CLAVICLE HARDWARE REMOVAL. NO FURTHER INFORMATION WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON 1/23/2024: DURING THE ORIGINAL PROCEDURE ON (B)(6) 2023, AN AR-2685DR-10 DISTAL CLAVICLE PLATE, AN AR-8827L-20 LOW PROFILE LOCKING SCREW, (4) AR-8827L-18 LOW PROFILE LOCKING SCREW, (2) AR-8835-16 LOW PROFILE SCREW, AN AR-8835-18 LOW PROFILE SCREW, (2) AR-8835L-18 LOW PROFILE LOCKING SCREW, AND AN AR-8835L-16 LOW PROFILE LOCKING SCREW WERE IMPLANTED. DURING THE REVISION SURGERY ON (B)(6) 2023, ALL THE IMPLANTS WERE SUCCESSFULLY REMOVED.

    Patient problems: Implant Pain

    MAUDE report 18724882 ↗
  • InjuryFeb 16, 2024Break; Break

    ON 1/18/2024, AN FDA MEDWATCH NOTIFICATION WAS RECEIVED VIA EMAIL. A FACILITY REPRESENTATIVE REPORTED THAT AN AR-2685DR-10 DISTAL CLAVICLE PLATE BROKE EIGHT WEEKS POST-OPERATIVE. THE PATIENT UNDERWENT SURGERY ON (B)(6) 2023 FOR AN OPEN REDUCTION WITH INTERNAL FIXATION OF THE RIGHT COMMINUTED MID-SHIFT CLAVICLE FRACTURE. EIGHT WEEKS LATER, THE PATIENT RETURNED TO THE CLINIC, EXPERIENCING PAINFUL POPS IN THE SHOULDER. ON (B)(6) 2023, THE PATIENT UNDERWENT REVISION SURGERY TO REMOVE THE AR-2685DR-10 DISTAL CLAVICLE PLATE. THE REVISION SURGERY WAS A REVISION OPEN REDUCTION AND INTERNAL FIXATION OF THE RIGHT CLAVICLE FRACTURE WITH ILIAC BONE GRAFT AND RIGHT CLAVICLE HARDWARE REMOVAL. NO FURTHER INFORMATION WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON 1/23/2024: DURING THE ORIGINAL PROCEDURE ON (B)(6) 2023, AN AR-2685DR-10 DISTAL CLAVICLE PLATE, AN AR-8827L-20 LOW PROFILE LOCKING SCREW, (4) AR-8827L-18 LOW PROFILE LOCKING SCREW, (2) AR-8835-16 LOW PROFILE SCREW, AN AR-8835-18 LOW PROFILE SCREW, (2) AR-8835L-18 LOW PROFILE LOCKING SCREW, AND AN AR-8835L-16 LOW PROFILE LOCKING SCREW WERE IMPLANTED. DURING THE REVISION SURGERY ON (B)(6) 2023, ALL THE IMPLANTS WERE SUCCESSFULLY REMOVED.

    Patient problems: Implant Pain

    MAUDE report 18723327 ↗
  • MalfunctionDec 19, 2023Positioning Failure; Fitting Problem; Positioning Failure; Fitting Problem

    ON 11/22/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8827L-18 LOW PROFILE LOCKING SCREW SS, 2.7 X 18 MM DID NOT ENGAGE. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2023. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 18360240 ↗
  • InjuryFeb 13, 2018Device Dislodged or Dislocated

    IT WAS REPORTED BY A FACILITY, WHO SENT A COPY OF A MEDWATCH ((B)(4)) WHICH THEY HAD FILED, THAT A PATIENT HAD ORIGINAL PLACEMENT OF HARDWARE THAT INCLUDED A PLATE AND SCREWS ON (B)(6) 2017 FOR A RIGHT DISPLACED LATERAL MALLEOLAR FRACTURE BY A HOSPITAL ORTHOPEDIC SURGEON. NO COMPLICATIONS WERE REPORTED. ON (B)(6) 2017 THE PATIENT WAS SEEN IN THE SURGEON'S OFFICE AND HAD COMPLAINTS OF INCREASING PAIN FOR THE PAST TWO MONTHS. THE PATIENT REPORTED TENDERNESS OVER PROMINENCE OF ONE OF THE SCREWS AS WELL AS THE ADDITIONAL AREA OF THE PLATE. THE SURGEON REVIEWED THE X-RAYS AND INDICATED NO CHANGE IN THE HARDWARE POSITION WITH NONE OF THE SCREWS BACKING OUT. ON (B)(6) 2017 THE PATIENT PRESENTED TO THE HOSPITAL WHERE THE ORTHOPEDIC SURGEON REPORTED THE FRACTURE AS HEALED BUT THE PATIENT REPORTED PAIN AND DYSFUNCTION WITH THE HARDWARE AND HAD WISHES FOR IT TO BE REMOVED. THE PATIENT HAD PLATE (AR-8943BR-04 LOCKING DISTAL FIBULA PLATE - (B)(4), LINE (B)(6)) AND SCREW REMOVAL FROM HER RIGHT ANKLE UNDER GENERAL PLUS LOCAL ANESTHESIA. NO COMPLICATIONS WERE REPORTED FOLLOW-UP INVESTIGATION: PER FACILITY MEDWATCH REPORTER, THE SURGEON'S OFFICE NOTES INDICATED THAT THERE WAS PAIN AND "PROMINENCE A

    Patient problems: Discomfort; Pain

    MAUDE report 7266949 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →