Arthrex® AR-8930-12Recalls & reported problems
Arthrex, Inc. · QUICKFIX SCREW, 2 X 12 MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867049185), received between Mar 22, 2017 and Jul 31, 2026 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Device Dislodged or Dislocated1
- Difficult to Remove1
- Entrapment of Device1
- Material Twisted/Bent1
- Output Problem1
Patient problems
- Foreign Body In Patient2
- Bone Shedding Debris1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- Pain1
- Swelling1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 31, 2026Material Twisted/Bent
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50101548 ↗MalfunctionJan 25, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON 01/09/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA E-MAIL THAT AN AR-8930-12 ARTHREX QUICKFIX¿ SCREW PULLED OUT. THIS OCCURRED ON 12/14/2023. NO ADDITIONAL INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION HAS BEEN ASKED.
Patient problems: Insufficient Information
MAUDE report 18581850 ↗MalfunctionAug 24, 2021Output Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SURGERY AFTER IMPLANTING THE SCREW THE INSERTER DID NOT BREAK OFF. DUE TO THIS FAILURE THE BONE SHATTERED. THE BONE WAS THEN FIXATED WITH A COMPRESSION SCREW AND KIRSCHNER-WIRE TO TREAT THE PATIENT.
Patient problems: Bone Shedding Debris
MAUDE report 12362132 ↗InjuryJul 7, 2017Break; Entrapment of Device
IT WAS REPORTED VIA FDA MEDWATCH (B)(4) WHICH WAS SUBMITTED BY THE FACILITY: "PER THE OPERATIVE REPORT...ANOTHER 2.0 ARTHREX 5 SCREW WAS DETERMINED TO BE PLACED PRODUCING SCLEROSIS. THE SCREW DID BREAK-OFF AND WAS UNABLE TO BE REMOVED DUE TO THE IMPLANTATION OF THE BONE. MULTIPLE ATTEMPTS WERE MADE TO REMOVE AND IT WAS TRIMMED-OFF AS SHORT AS POSSIBLE." (THIS DESCRIPTION IN QUOTES IS EXACTLY AS ENTERED BY THE FACILITY MEDWATCH REPORTER) INTENDED PROCEDURE WAS A LEFT 3RD HAMMERTOE CORRECTION WITH PIP FUSION. UPDATE 6/22/2017: BASED ON THE DEVICE RETURNED AND THE EVENT DESCRIPTION PROVIDED THE MOST LIKELY PART NUMBER WAS DETERMINED TO BE AR-8930-12. FACILITY MEDWATCH REPORTER HAS RETIRED SINCE SUBMITTING MEDWATCH (B)(4). CURRENT RISK MANAGER WAS ABLE TO LOCATE THE REMAINING PIECE OF THE BROKEN SCREW AND IT HAS BEEN RETURNED. CURRENT RISK MANAGER HAS BEEN UNABLE TO OBTAIN OR PROVIDE ANY ADDITIONAL INFORMATION.
Patient problems: Foreign Body In Patient
MAUDE report 6692699 ↗InjuryMar 22, 2017Break; Difficult to Remove
IT WAS REPORTED ON 9/15/2016 BY A PATIENT VIA THE ARTHREX WEBSITE THAT THE PATIENT HAD UNDERGONE FOOT SURGERY (B)(6) 2015. SURGERY WAS LEFT PROXIMAL BUNIONECTOMY IN CONJUNCTION WITH 2ND MTP RECONSTRUCTION FOLLOWING DISLOCATION AND PRIOR PODIATRIC MANAGEMENT. DURING THE SURGERY THE FOLLOWING PARTS WERE IMPLANTED: AR13200M-055R ((B)(4)), AR-8724-16 ((B)(4)), AR-8724-12 ((B)(4)), AR-8724-14 ((B)(4)), AR-8724-18 ((B)(4)), AR-8930-12 ((B)(4)) AND AR-8724-12PT ((B)(4)). ON (B)(6) 2016 PATIENT FOLLOWED UP WITH SURGEON DUE TO RECENT ONSET OF LEFT MID-FOOT PAIN AND SWELLING. MEDICAL RECORDS NOTED THAT UPON EXAMINATION OF THE LEFT FOOT THERE WAS PAIN AT THE 1ST TMT AND PROXIMAL OSTEOTOMY SITE WITH VISIBLE SWELLING. THERE WAS NO ERYTHEMA, DRAINAGE OR OTHER EVIDENCE OF INFECTION. X-RAY RESULTS NOTED THERE IS BROKEN HARDWARE (PROXIMAL BUNION PLATE - AR-13200M-055R) WITH QUESTION OF NON-UNION. SURGEON HAS RECOMMENDED A PROXIMAL LEFT BUNION HARDWARE REMOVAL WITH A PLAN TO PROCEED WITH A LAPIDUS FUSION IF IN FACT THE OSTEOTOMY IS INCOMPLETE. PATIENT REVISION IS TENTATIVELY SCHEDULED FOR EARLY 2017 DUE TO PATIENT LACK OF AVAILABLE TIME OFF FROM WORK. ADDITIONAL INFORMATION RECEIVED 2/28/2017: THE R
Patient problems: Foreign Body In Patient; Pain; Swelling
MAUDE report 6424850 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page