Skip to content

Arthrex® AR-8930GRecalls & reported problems

Arthrex, Inc. · DEPTH GUIDE FOR 3.0 MM QUICKFIX SCREW · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867049260), received between Dec 20, 2023 and Apr 9, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect Measurement2
  • Output Problem1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 9, 2024Output Problem; Output Problem

    ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS (B)(4) THAT DURING A PROCEDURE THE SURGEON WENT TO START IMPLANTING SCREWS INTO THE PATIENT WHEN THE AR-8970RH HANDLE WAS NOT RATCHETING AND AN AR-8930G DEPTH GAUGE SHOWS AN INCORRECT NUMBER. AFTER THE SURGEON IMPLANTED FOUR SCREWS, HE WENT TO CHECK THE X-RAY, AND EVERY SCREW WAS TOO LONG. HE CHECKED THE DEVICE AND WHEN THE DEVICE WAS PULLED ALL THE WAY BACK TO A LENGTH OF 0MM, IT SHOWED 10MM.

    Patient problems: Insufficient Information

    MAUDE report 19075500 ↗
  • MalfunctionMar 7, 2024Incorrect Measurement; Incorrect Measurement

    ON (B)(6) 2024, IT WAS REPORTED BY A DISTRIBUTOR VIA SEMS (B)(4) THAT AN AR-8930G ARTHREX QUICKFIX¿ DEPTH GUIDE IS 10 MM OFF TO GIVE READINGS. THERE WAS NO EFFECT ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18856622 ↗
  • MalfunctionDec 20, 2023Incorrect Measurement; Incorrect Measurement

    ON 11/20/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06406153 THAT AN AR-8930G ARTHREX QUICKFIX¿ DEPTH GUIDE WAS NOT MEASURING PROPERLY. IT WAS NOTICED DURING A CASE AND THE SURGEON HAD TO GUESS ON ALL OF HIS SCREW MEASUREMENTS. EVEN WHEN THE DEPTH GAGE IS "ZEROED OUT", IT IS STILL MEASURING OVER 10MM. THERE WAS NO ADDITIONAL INFORMATION PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 18367071 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page