Arthrex® AR-8933-20Recalls & reported problems
Arthrex, Inc. · LO-PRO SCRW,TI,3.0MMX 20MMCORT · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867049567), received between Jun 30, 2020 and Aug 16, 2023 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
- Detachment of Device or Device Component1
- Entrapment of Device1
- Material Separation1
- Patient Device Interaction Problem1
Patient problems
- Foreign Body In Patient2
- Device Embedded In Tissue or Plaque1
- Hypersensitivity/Allergic reaction1
- Insufficient Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryAug 16, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED, THAT THE PATIENT UNDERWENT SURGERY ON (B)(6) 2018. FOR AN OPEN REDUCTION INTERNAL FIXATION OF A LEFT TRIMALLEOLAR ANKLE FRACTURE, WITH FIXATION OF THE DISTAL FIBULA FRACTURE, POSTERIOR MALLEOLUS FRACTURE, AND SYNDESMOTIC FIXATION. DURING THE (B)(6) 2018 SURGERY, THE FOLLOWING PRODUCTS WERE USED AND/OR IMPLANTED: AR-8740-52, PTS QCKFIX SCRW, TI, CANN ST, CANC., 4.0X 52MM, LOT/SERIAL#: (B)(6); AR-8925T K, -LESS T-ROPE W/DRV, SYN REPR, TI, LOT/SERIAL#: (B)(6); AR-8933-20, LO-PRO SCRW, TI, 3.0MMX 20MMCORT, LOT/SERIAL#: (B)(6); AR-8933V-12, 3.0 X 12MM VAL SCREW, TI, LOT/SERIAL#: (B)(6); AR-8933V-14, 3.0 X 14MM VAL SCREW, TI, LOT/SERIAL#: (B)(6); AR-8933V-16, 3.0 X 16MM VAL SCREW, TI, LOT/SERIAL#: (B)(6); AR-8933V-18, 3.0 X 18MM VAL SCREW, TI, LOT/SERIAL#: (B)(6); AR-8933V-20, 3.0 X 20MM VAL SCREW, TI, LOT/SERIAL#: (B)(6); AR-8935-12, LO-PRO SCRW, TI, 3.5MMX 12MM, LOT/SERIAL#: (B)(6); AR-8935-14, LO-PRO SCRW, TI, 3.5MMX 14MM, LOT/SERIAL#: (B)(6); AR-8935-14, LO-PRO SCRW, TI, 3.5MMX 14MM, LOT/SERIAL#: (B)(6); AR-8935-26, LO-PRO SCRW, TI, 3.5MMX 26MM, LOT/SERIAL#: (B)(6); AR-8943-01, GUIDE WIRE, TROCAR TIP, 1.35 MM, LOT/SERIAL#: (B)(6); AR-8943-02, DRILL BIT, CANNULATED, 2
Patient problems: Hypersensitivity/Allergic reaction
MAUDE report 17561553 ↗MalfunctionApr 26, 2021Break
IT WAS REPORTED THAT THE HEAD OF THE SCREW BROKE OFF, LEAVING THE REMAINDER OF THE SCREW IN THE PATIENT. THIS OCCURRED DURING AN MTP FUSION CASE. THE SURGEON WAS ATTEMPTING TO REMOVE THE SCREW TO USE A SHORTER SCREW WHEN THE HEAD BROKE OFF. PATIENT IS MALE, 40S.
Patient problems: Foreign Body In Patient
MAUDE report 11724932 ↗MalfunctionFeb 11, 2021Detachment of Device or Device Component
IT WAS REPORTED THAT WHEN THE SURGEON DID THEIR FINAL TURN OF THE SCREW INTO CORTICAL BONE, THE HEAD OF THE AR-8933-20 SCREW POPPED OFF.
Patient problems: Insufficient Information
MAUDE report 11314841 ↗InjuryOct 6, 2020Break; Entrapment of Device
IT WAS REPORTED, THAT DURING A MTC FUSION, UPON INSERTION OF THE AR-8933-20, LO-PROFILE SCREW, AS THE SURGEON REACHED THE PLATE THE HEAD OF THE SCREW BROKE OFF. THE BROKEN FRAGMENT WAS REMOVED. THE SURGEON TRIED TO REMOVE THE BROKEN AR-8933-20 HOWEVER, WAS UNABLE TO REMOVE THE SCREW BODY. THE BODY OF THE SCREW WAS LEFT IN THE PATIENT. THE SURGEON DRILLED A SECOND INSERTION SITE AND THE CASE WAS COMPLETED USING A DIFFERENT AR-8933-20, LO-PROFILE SCREW. ADDITIONAL INFORMATION OBTAINED 09/21/2020: THE SECOND DRILL HOLE WAS NOT PUT INTO THE SAME HOLE AS THE BROKEN SCREW BODY, BUT WAS INSTEAD PUT IN A NEW HOLE IN THE PLATE DISTAL AND PROXIMAL TO THE BROKEN SCREW. THE DISTAL SCREW WAS AN AR-8933-20 AND PROXIMAL TO THE BROKEN SCREW WAS AN AR-8933-18.
Patient problems: Foreign Body In Patient
MAUDE report 10634452 ↗InjuryJun 30, 2020Material Separation
IT WAS REPORTED, THAT DURING AN ANKLE PROCEDURE, DURING INSERTION OF THE AR-8933-20, LO-PROFILE SCREW THE HEAD OF THE SCREW BROKE OFF. THE BODY OF THE SCREW WAS LEFT IN THE PATIENT AND IS FLUSH WITH THE FIBULA. THE BROKEN PIECE WAS REMOVED. THE SURGEON DRILLED A SECOND INSERTION SITE AND THE CASE WAS COMPLETED USING A DIFFERENT AR-8933-20, LO-PROFILE SCREW.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10213506 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page