Skip to content

Arthrex® AR-8943-08Recalls & reported problems

Arthrex, Inc. · SCREWDRIVER, T10 HEXALOBE · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867052185), received between Sep 14, 2021 and Feb 6, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2026Break

    ON 16-JAN-2026, A FACILITY REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-8943-08 T10 HEXALOBE SCREWDRIVER TIP BROKE DURING A HARDWARE REMOVAL CASE. THE TIP THAT BROKE WAS NOT RECOVERED. X-RAYS WERE TAKEN TO ENSURE IT WAS NOT LEFT IN THE PATIENT. THE DRIVER WAS REPLACED, AND THE CASE CONTINUED WITH NO ADVERSE EFFECTS TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED ON 20-JAN-2026. THE REP ADVISED THAT NO ATTACHMENT OR ADAPTER WAS USED, AND THAT THE 3.5 MM HEADLESS COMPRESSION SCREW WAS BEING REMOVED BY HAND VIA THE AR-8943-08.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24277856 ↗
  • MalfunctionDec 6, 2022Break

    ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT A AR-8770-01 DRIVER SHAFT, AND A AR-8943-08 SCREWDRIVER BROKE. THIS OCCURRED DURING A CASE, WITH NO PATIENT EFFECT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15923867 ↗
  • MalfunctionSep 14, 2021Break

    IT WAS REPORTED THE DOCTOR WAS TIGHTENING A 2.7 SCREW IN DISTAL FIBULA PLATE AND ON THE LAST QUARTER TURN THE DRIVER TIP BROKE IN THE SCREW HEAD. THE DOCTOR WAS ABLE TO REMOVE THE BROKE PART FROM THE SCREW WITH NO ISSUE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Insufficient Information

    MAUDE report 12466904 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page