Arthrex® AR-8943BL-04Recalls & reported problems
Arthrex, Inc. · LOCK DISTAL FIBULA PLT,SS LFT,4H · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867052512), received between Jan 27, 2020 and Apr 20, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Break1
- Difficult to Remove1
Patient problems
- Dizziness1
- Fatigue1
- Headache1
- Insufficient Information1
- Tinnitus1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 20, 2021Break; Difficult to Remove
IT WAS REPORTED THAT DURING A PROCEDURE, THE SURGEON/SCRUB TECH DID NOT REALIZE THEY WERE USING A CANNULATED DRIVER AND CROSS THREADED THE LOCKING SCREW INTO THE PLATE. THE TORQUE ON THE DRIVER BROKE THE DRIVER AND THE SCREW AND THEY WERE UNABLE TO GET THE SCREW OUT OF THE PLATE. ADDITIONAL INFORMATION 4/9/2021: IT WAS REPORTED THAT DURING AN ANKLE FRACTURE REPAIR PROCEDURE, THE SURGEON/SCRUB TECH DID NOT REALIZE THEY WERE USING A CANNULATED DRIVER AND CROSS THREADED THE LOCKING SCREW INTO THE PLATE. THE IMPLANT SCREW WAS CROSS THREADED INTO THE PLATE AND THE WHOLE THING PULLED OUT OF THE PATIENT¿S BONE. THE TORQUE ON THE DRIVER BROKE THE DRIVER AND THE SCREW AND THEY WERE UNABLE TO GET THE SCREW OUT OF THE PLATE. THE CASE WAS COMPLETED USING A NEW PLATE, SCREW AND DRIVER.
Patient problems: Insufficient Information
MAUDE report 11692724 ↗InjuryJan 27, 2020Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT THAT SHE HAD AN ORIGINAL ANKLE PROCEDURE (B)(6) 2017. DURING THE SURGERY SIX ARTHREX PRODUCTS WERE IMPLANTED. THE PATIENT DEVELOPED ALLERGIC LIKE REACTION SYMPTOMS AND RECENTLY HAD TO UNDERGO A SECOND SURGERY ON (B)(6) 2019 TO EXPLANT THE ARTHREX DEVICES. PATIENT DID NOT HAVE AN FURTHER DETAILS AT THE TIME OF REPORT BUT WILL PROVIDE SOON. ADDITIONAL INFORMATION 1/8/2020: ORIGINAL PROCEDURE WAS ON (B)(6) 2017 AND WAS A LEFT ANKLE ORIF. FOLLOWING SURGERY THE PATIENT WAS IN THE HOSPITAL FOR THREE NIGHTS. SHE STARTED HAVING RINGING IN HER EARS AND DIZZINESS 3 MONTHS AFTER SURGERY. THIS WAS FOLLOWED BY TIREDNESS, MIGRAINES, AND TMJ ON HER RIGHT SIDE. SHE ALSO DEVELOPED NOTICEABLY DRIER, FLAKY SKIN AND MORE ANXIETY AFTER ABOUT 6 MONTHS. SHE DEVELOPED FOOD INTOLERANCE TO DAIRY AND DEVELOPED A THYROID NODULE AS WELL. PATIENT HAD A SECOND SURGERY ON (B)(6) 2019 TO EXPLANT ALL OF THE ORIGINAL IMPLANT DEVICES. THE SECOND SURGERY WAS PERFORMED BY A DIFFERENT SURGEON AT A DIFFERENT FACILITY. THE FOLLOWING DEVICES WERE REPORTED TO HAVE BEEN IMPLANTED ON (B)(6) 2017 AND EXPLANTED ON (B)(6) 2019 (SPECIFIC ARTHREX PART NUMBERS WERE NOT PROVIDED): AR-8926SS, KNOTLESS T
Patient problems: Dizziness; Fatigue; Headache; Tinnitus
MAUDE report 9631382 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page