Arthrex® AR-8956-01Recalls & reported problems
Arthrex, Inc. · UNIVERSAL 5TH METARSAL HOOK PLATE · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867055636), received between Jan 28, 2020 and Jan 28, 2020 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Break1
Patient problems
- Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 28, 2020Break; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT THAT SHE HAD UNDERGONE A RIGHT FOOT METATARSAL FRACTURE PROCEDURE ON (B)(6) 2019. THE PATIENT BEGAN HAVING SEVERE PAIN AND ON OR ABOUT (B)(6) 2019 IT WAS DETERMINED THAT THE PLATE IMPLANT HAD BROKEN. THE PATIENT UNDERWENT A REVISION SURGERY BY THE SAME SURGEON, AT THE SAME FACILITY, ON (B)(6) 2019. DURING THE REVISION PROCEDURE, THE PLATE AND CORRESPONDING SCREWS WERE EXPLANTED. AT THE TIME OF THE PATIENT'S INITIAL CALL TO ARTHREX ON 9/25/2019 THE PATIENT HAD NO INFORMATION ON THE PRODUCTS USED OR THE FACILITY INVOLVED. THE PATIENT INDICATED THAT THE SURGEON INFORMED HER THAT HE HAD REPORTED THE INCIDENT TO HIS SALES REP. THE PATIENT ALSO REPORTED THAT THE SURGEON HAD INFORMED HER THAT THE EXPLANTED DEVICES HAD BEEN TURNED OVER TO THE REP FOR RETURN FOR EVALUATION. ON 12/30/19 THE PATIENT PROVIDED THE ADDRESS OF THE ORIGINAL AND REVISION SURGERIES. USING THE ADDRESS ARTHREX WAS ABLE TO OBTAIN THE ORIGINAL SALES ORDER FROM THE (B)(6) 2019 PROCEDURE WHICH IDENTIFIED THE DEVICES USED IN SURGERY. IT WAS ALSO DETERMINED AT THIS TIME THAT THE INCIDENT HAD NOT BEEN PREVIOUSLY REPORTED TO ARTHREX BY THE SURGEON OR SALES REPRESENTATIVE. THE FOLLOWING WERE THE
Patient problems: Pain
MAUDE report 9635990 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page