Skip to content

Arthrex® AR-9121-06Recalls & reported problems

Arthrex, Inc. · UNIVERSAL GLENOID - INLAY LARGE PLUS · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867057463), received between Aug 10, 2021 and Jun 20, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Material Erosion” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Material Erosion1
  • Positioning Failure1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 20, 2024Positioning Failure; Positioning Failure

    ADDITIONAL INFORMATION RECEIVED ON 7/23/2024: THE INTENDED PROCEDURE WAS AN ANATOMIC SHOULDER WITH A UNIVERSAL CONVERTIBLE BASEPLATE BUT SINCE THE ARTHREX UNIVERSAL GLENOID WOULD NOT PRESS ONTO THE BASEPLATE, THE SURGEON COMPLETED THE CASE AS A REVERSE PROCEDURE. THE CASE WAS NOT DELAYED, AND ADDITIONAL ANESTHESIA WAS NOT NEEDED. THE PATIENT SUFFERED NO NEGATIVE EFFECTS ON OR AFTER THE PROCEDURE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19579466 ↗
  • MalfunctionAug 25, 2021Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION SHOULDER PROCEDURE ON (B)(6) 2021 WHERE AN ARTHREX IMPLANT WAS REMOVED. NO FURTHER DETAILS, ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONAL INFORMATION PROVIDED 08/19/2021: THE DATE OF THE PRIMARY PROCEDURE WAS ON (B)(6) 2021 FOR A LEFT TOTAL SHOULDER ARTHROPLASTY. THIS PROCEDURE TOOK PLACE AT (B)(6) SURGERY CENTER, ARTHREX ACCOUNT # (B)(4). REVISION FOR A LEFT ANATOMIC TOTAL SHOULDER ARTHROPLASTY TO REVERSE SHOULDER ARTHROPLASTY TOOK PLACE AT ARTHREX ACCOUNT # (B)(4) ON (B)(6) 2021 WHERE THE FOLLOWING DEVICES WERE EXPLANTED: AR-9301-02 (20.00960); AR-9301-47CPC (19.04096); AR-9347-20 (2797, 47/20); AR-9121-06 (17.01402).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12364477 ↗
  • InjuryAug 10, 2021Material Erosion

    IT WAS REPORTED THAT A REVISION SURGERY WAS NECESSARY AS THE GLENOID COMPONENT ABRADED. THE SHOULDER PROSTHESIS WAS CHANGED TO AN INVERSE PROSTHESIS. NO FURTHER INFORMATION RECEIVED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12300879 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page