Arthrex® AR-9202Recalls & reported problems

Arthrex, Inc. · UNIVERS/ECLIPSE ORIEN PIN FOR RESECT GD · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00888867058224), received between Apr 8, 2021 and Mar 17, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Material Twisted/Bent3
  • Physical Resistance/Sticking2
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Insufficient Information1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 17, 2026Break

    ON (B)(6), IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-(B)(4) THAT AN AR-9202 PIN WAS STRUCK WITH A MALLET WHEN REMOVING AND SNAPPED. NOTICED DURING A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24618063 ↗
  • MalfunctionMar 17, 2026Break

    ON 24-FEB-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9202 PIN WAS STRUCK WITH A MALLET WHEN REMOVING AND SNAPPED. NOTICED DURING A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24617919 ↗
  • MalfunctionNov 25, 2025Material Twisted/Bent; Material Twisted/Bent

    ON 11/05/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (QTY. 2) OF AN AR-9202 UNIVERS/ECLIPSE ORIEN PINS WERE BENT AND UNUSABLE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645155 ↗
  • MalfunctionNov 25, 2025Material Twisted/Bent; Material Twisted/Bent

    ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (QTY. 2) OF AN AR-9202 UNIVERS/ECLIPSE ORIEN PINS WERE BENT AND UNUSABLE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645039 ↗
  • MalfunctionJul 15, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 06/18/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-9202 ARTHREX ECLIPSE¿ AND UNIVERS¿ ORIENTATION PIN FOR RESECTION GUIDES SHAFT FUSES INSIDE THE REAMER HANDLE. THE T HANDLE WILL NOT ACCEPT THE SMALLER REAMERS. THE PATIENT AND PHYSICIAN WERE NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22508994 ↗
  • MalfunctionDec 31, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SHOULDER SURGERY THE DEVICE WAS BENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21048012 ↗
  • MalfunctionJun 6, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED ON 09/09/2022 BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-9202 VERSION RODS ARE WORN. THIS WAS DISCOVERED DURING AN UNSPECIFIED TIME WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19478216 ↗
  • MalfunctionJun 14, 2021Break; Physical Resistance/Sticking

    IT WAS REPORTED DURING DURING A HEMI SHOULDER ARTHROPLASTY PROCEDURE, THE VERSION ROD BROKE OFF THE AR-9401-13 HEAD CUT GUIDE. THE THREADED PORTION OF THE VERSION ROD IS NOW STUCK INSIDE THE GUIDE. THE CASE WAS STILL SUCCESSFUL AND NO HARM WAS DONE. SEE ATTACHED IMAGES. ADDITIONAL INFORMATION RECEIVED ON 5/26/2021: THE REP PROVIDED THE PART NUMBER OF THE VERSION ROD THAT BROKE DURING THE PROCEDURE (AR-9202). THE VERSION ROD BROKE WHILE THE GUIDE WAS IN USE. THE RODS ARE THREADED INTO THE TOP OF THE GUIDE, AND ARE NOT DIRECTLY IN THE PATIENT BUT ABOVE THE HUMERUS.

    Patient problems: Insufficient Information

    MAUDE report 11990877 ↗
  • MalfunctionApr 8, 2021Break

    IT WAS REPORTED THAT THE PIN FROM AR-9207S KIT (LOT:11536563) WAS INTRODUCED INTO THE AR-9401-13 CUT GUIDE. PIN BOUND UP IN CANNULATION CAUSING GUIDE TO SPIN AND BREAKING OFF TWO AR-9202 VERSION RODS. PIN WAS NOT ABLE TO BE REMOVED FROM GUIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11633754 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates