Arthrex® AR-9202Recalls & reported problems
Arthrex, Inc. · UNIVERS/ECLIPSE ORIEN PIN FOR RESECT GD · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00888867058224), received between Apr 8, 2021 and Mar 17, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Material Twisted/Bent3
- Physical Resistance/Sticking2
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 17, 2026Break
ON (B)(6), IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-(B)(4) THAT AN AR-9202 PIN WAS STRUCK WITH A MALLET WHEN REMOVING AND SNAPPED. NOTICED DURING A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24618063 ↗MalfunctionMar 17, 2026Break
ON 24-FEB-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9202 PIN WAS STRUCK WITH A MALLET WHEN REMOVING AND SNAPPED. NOTICED DURING A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24617919 ↗MalfunctionNov 25, 2025Material Twisted/Bent; Material Twisted/Bent
ON 11/05/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (QTY. 2) OF AN AR-9202 UNIVERS/ECLIPSE ORIEN PINS WERE BENT AND UNUSABLE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23645155 ↗MalfunctionNov 25, 2025Material Twisted/Bent; Material Twisted/Bent
ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (QTY. 2) OF AN AR-9202 UNIVERS/ECLIPSE ORIEN PINS WERE BENT AND UNUSABLE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23645039 ↗MalfunctionJul 15, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON 06/18/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-9202 ARTHREX ECLIPSE¿ AND UNIVERS¿ ORIENTATION PIN FOR RESECTION GUIDES SHAFT FUSES INSIDE THE REAMER HANDLE. THE T HANDLE WILL NOT ACCEPT THE SMALLER REAMERS. THE PATIENT AND PHYSICIAN WERE NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22508994 ↗MalfunctionDec 31, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SHOULDER SURGERY THE DEVICE WAS BENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21048012 ↗MalfunctionJun 6, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED ON 09/09/2022 BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-9202 VERSION RODS ARE WORN. THIS WAS DISCOVERED DURING AN UNSPECIFIED TIME WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19478216 ↗MalfunctionJun 14, 2021Break; Physical Resistance/Sticking
IT WAS REPORTED DURING DURING A HEMI SHOULDER ARTHROPLASTY PROCEDURE, THE VERSION ROD BROKE OFF THE AR-9401-13 HEAD CUT GUIDE. THE THREADED PORTION OF THE VERSION ROD IS NOW STUCK INSIDE THE GUIDE. THE CASE WAS STILL SUCCESSFUL AND NO HARM WAS DONE. SEE ATTACHED IMAGES. ADDITIONAL INFORMATION RECEIVED ON 5/26/2021: THE REP PROVIDED THE PART NUMBER OF THE VERSION ROD THAT BROKE DURING THE PROCEDURE (AR-9202). THE VERSION ROD BROKE WHILE THE GUIDE WAS IN USE. THE RODS ARE THREADED INTO THE TOP OF THE GUIDE, AND ARE NOT DIRECTLY IN THE PATIENT BUT ABOVE THE HUMERUS.
Patient problems: Insufficient Information
MAUDE report 11990877 ↗MalfunctionApr 8, 2021Break
IT WAS REPORTED THAT THE PIN FROM AR-9207S KIT (LOT:11536563) WAS INTRODUCED INTO THE AR-9401-13 CUT GUIDE. PIN BOUND UP IN CANNULATION CAUSING GUIDE TO SPIN AND BREAKING OFF TWO AR-9202 VERSION RODS. PIN WAS NOT ABLE TO BE REMOVED FROM GUIDE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11633754 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page