Skip to content

Arthrex® AR-9301-02Recalls & reported problems

Arthrex, Inc. · ECLIPSE CAGE SCREW M, 35MM · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867059931), received between Feb 7, 2025 and Sep 29, 2026 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem3
  • Positioning Failure2
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Device Overstimulation of Tissue1
  • Erosion1
  • Failure of Implant1
  • Insufficient Information1
  • Metal Related Pathology1
  • Pain1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Patient Device Interaction Problem

    On 10/10/2024, a Clindex notification was received indicating that a patient listed on the Shoulder Arthroplasty Registry suffered a displaced lesser tuberosity osteotomy and subscapularis tendon tear. On (b)(6) 2024, the patient reported continuous pain after having experienced a jerk on the arm while walking the dog. The patient was treated with steroid injections. In September of 2024, a CT scan showed a displaced lesser tuberosity osteotomy and subscapularis tendon tear. The patient has been scheduled for an Arthroscopic subscapularis tendon tear repair procedure in October 2024. This event was initially indicated as unrelated to the Eclipse devices implanted on (b)(6) 2024.\\Additional information received on 10/17/2024:\During the original surgery, an AR-9106-03 Arthrex Univers VaultLock Glenoid, an AR-9349-18 Arthrex ECLIPSE Humeral Head, an AR-9301-49CPC Arthrex ECLIPSE Trunion, and an AR-9301-02 Arthrex ECLIPSE Cage Screw were implanted. After the patient received steroid injections, the patient's symptoms were persistent despite the injection, therefore the CT was ordered for further evaluation. Upon CT review, lesser tuberosity osteotomy displacement was seen, and it was

    Patient problems: Device Overstimulation of Tissue

    MAUDE report 50938391 ↗
  • InjuryMar 24, 2026Patient Device Interaction Problem

    ON 04-MAR-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT A PATIENT UNDERWENT A REVISION FROM ECLIPSE + VAULTLOCK TSA TO RTSA ON (B)(6) 2026. THE PATIENT¿S ORIGINAL SURGERY DATE WAS ON (B)(6) 2024. THE REVISION WAS PERFORMED DUE TO SUBSCAPULARIS FAILURE AND NOT DUE TO FAILURE OF THE ECLIPSE OR VAULTLOCK IMPLANTS. HOWEVER, SIGNIFICANT VAULTLOCK EROSION WAS NOTED AT LESS THAN TWO YEARS POST-IMPLANTATION. THE ARTHREX IMPLANTS EXPLANTED DURING THE REVISION INCLUDED: AR-9106-01 ARTHREX UNIVERS VAULTLOCK GLENOID, SMALL, AR-9301-02 ARTHREX ECLIPSE CAGE SCREW MEDIUM (35.0 MM), AR-9301-39CPC ARTHREX ECLIPSE TRUNION 39 MM, SLOTTED, TPS CAP, AND AN AR-9339-18 ARTHREX ECLIPSE HUMERAL HEAD, 39/18. ADDITIONAL INFORMATION WAS RECEIVED ON 10-MAR-2026: THE PATIENT WAS INITIALLY TREATED ON (B)(6) 2024 FOR A SUBSCAPULARIS TEAR SUSTAINED DURING A PHYSICAL ALTERCATION. DURING THE REVISION, ANTERIOR EROSION MEASURING APPROXIMATELY 2¿3 MM WAS OBSERVED. THE SURGEON DID NOT COMMENT ON WHETHER THE EROSION CONTRIBUTED TO THE NEED FOR REVISION OR IF IT WAS AN INCIDENTAL FINDING. THE REVISION PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO CASE DELAY AND NO ADDITIONAL ANESTHESIA REPORTED. NO FURTHER SURGERIE

    Patient problems: Erosion; Insufficient Information

    MAUDE report 24673781 ↗
  • MalfunctionJan 15, 2026Positioning Failure

    ON (B)(6) 2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-9301-02 ARTHREX ECLIPSE CAGE SCREW MEDIUM (35.0 MM) WOULD NOT BITE WHEN IT WAS BEING PROPERLY SIZED. THIS WAS DISCOVERED DURING A PROCEDURE, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 08-JAN-2026: THIS OCCURRED INSIDE THE PATIENT DURING A TOTAL SHOULDER ARTHROPLASTY PROCEDURE ON (B)(6) 2025. THE CASE WAS COMPLETED WITH AN AR-9301-03 ARTHREX ECLIPSE CAGE SCREW LARGE (40.0 MM). THE CASE WAS DELAYED FIVE MINUTES, AND IT IS UNKNOWN IF ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24078327 ↗
  • InjuryDec 30, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT THE PE INLAY AR-9121-02 OF THE UNIVERSAL GLENOID GOT WORN OFF, RESULTING IN METAL-ON-METAL CONTACT BETWEEN ECLIPSE HUMERAL HEAD AND THE BASEPLATE OF THE UNIVERSAL GLENOID. THIS RESULTED IN METALLOSE AND RELATED PROBLEMS REQUIRING A REVISION SURGERY WITH A CONVERSION TO A REVERSE PROSTEHSIS. DURING THE REVISION SURGERY THE ECLIPSE HUMERAL COMPONENTS AND THE REMAINING PARTS OF THE INLAY WERE REMOVED AND PROPER SEATING OF THE UNIVERSAL GLENOID BASEPLATE WAS CONFIRMED. THEN A GLENOSPHERE WAS INSERTED ONTO THE BASEPLATE AND THE UNIVERS REVERS HUMERAL COMPONENTS WERE IMPLANTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Metal Related Pathology

    MAUDE report 23927368 ↗
  • MalfunctionOct 16, 2025Positioning Failure; Positioning Failure

    ON (B)(6) 2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-9301-02 ARTHREX ECLIPSE CAGE SCREW (MEDIUM) FAILED TO ACHIEVE ADEQUATE FIXATION FOLLOWING IMPLANTATION. ALL FIXATION WAS LOST DURING THE INITIATION OF THE SPEEDSCAP REPAIR FOR THE SUBSCAPULARIS, AND THE IMPLANT BECAME LOOSE. SUBSEQUENTLY, A LARGER-SIZED CAGE SCREW WAS IMPLANTED, WHICH PROVIDED STABLE FIXATION DESPITE THE INITIAL SIZING AS MEDIUM. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY REPORTED PATIENT HARM. THIS ISSUE WAS IDENTIFIED DURING AN ECLIPSE TSA PROCEDURE PERFORMED ON (B)(6) 2025. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON 10/09/2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23311244 ↗
  • InjuryFeb 7, 2025Break; Break

    ON 1/15/2025, AN FDA MEDWATCH NOTIFICATION WAS RECEIVED VIA EMAIL. A PATIENT REPORTED HAVING AN ARTHREX SHOULDER REPLACEMENT SURGERY ABOUT FOUR YEARS AGO. THE PATIENT REPORTED WORSENING SYMPTOMS AFTER THE SURGERY ON (B)(6) 2020. LESS THAN FOUR YEARS AFTER THE INITIAL SURGERY, THE PATIENT UNDERWENT REVISION SURGERY TO A REVERS TOTAL SHOULDER REPLACEMENT. THE PATIENT REPORTED THAT DURING THE REVISION SURGERY, IT WAS DISCOVERED THAT THE GLENOHUMERAL COMPONENT WAS DAMAGED, AND ONE OF THE PEGS WAS SHEERED COMPLETELY OFF, CAUSING PAIN, SWELLING, AND PIOUS AMOUNTS OF FLUID BUILDUP. NO FURTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION REQUESTED ON 1/23/2025. ADDITIONAL INFORMATION RECEIVED ON 1/23/025: DURING THE INITIAL SURGERY ON (B)(6) 2020, AN AR-9106-01 ARTHREX UNIVERS VAULTLOCK GLENOID, AN AR-9301-02 ARTHREX ECLIPSE CAGE SCREW, AND AN AR-9301-43CPC ARTHREX ECLIPSE TRUNION WERE IMPLANTED. THE PATIENT STARTED TO EXPERIENCE PAIN AND SOUGHT OUT MEDICAL TREATMENT. THE PATIENT UNDERWENT TWO DEBRIDEMENT PROCEDURES, IN WHICH THE MISSING PEG WAS DISCOVERED. THE THIRD PROCEDURE WAS THE REVERS TOTAL SHOULDER REPLACEMENT WHICH OCCURRED ON (B)(6) 2024.

    Patient problems: Failure of Implant; Pain

    MAUDE report 21337006 ↗

FDA clearance and classification

The FDA regulates product code QHQ (total shoulder arthroplasty system) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the Arthrex® AR-9301-02

Has the Arthrex® AR-9301-02 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00888867059931), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Arthrex® AR-9301-02?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Feb 7, 2025 and Sep 29, 2026. In those reports, the most common specific device problems are “Patient Device Interaction Problem” (3), “Positioning Failure” (2) and “Break” (1); the most common patient problems are “Device Overstimulation of Tissue” (1), “Erosion” (1) and “Failure of Implant” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Arthrex® AR-9301-02?

None of the reports are classified as deaths, 4 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Arthrex® AR-9301-02 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Arthrex® AR-9301-02, and is it still sold?

The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-9301-02). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.