Skip to content

Arthrex® AR-9301-43CPCRecalls & reported problems

Arthrex, Inc. · ECLIPSE TRUNION 43MM, SLOTTED, TPS CaP · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867276147), received between Sep 21, 2020 and Feb 7, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device Dislodged or Dislocated1
  • Use of Device Problem1

Patient problems

  • Pain2
  • Failure of Implant1
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 7, 2025Break; Break

    ON 1/15/2025, AN FDA MEDWATCH NOTIFICATION WAS RECEIVED VIA EMAIL. A PATIENT REPORTED HAVING AN ARTHREX SHOULDER REPLACEMENT SURGERY ABOUT FOUR YEARS AGO. THE PATIENT REPORTED WORSENING SYMPTOMS AFTER THE SURGERY ON (B)(6) 2020. LESS THAN FOUR YEARS AFTER THE INITIAL SURGERY, THE PATIENT UNDERWENT REVISION SURGERY TO A REVERS TOTAL SHOULDER REPLACEMENT. THE PATIENT REPORTED THAT DURING THE REVISION SURGERY, IT WAS DISCOVERED THAT THE GLENOHUMERAL COMPONENT WAS DAMAGED, AND ONE OF THE PEGS WAS SHEERED COMPLETELY OFF, CAUSING PAIN, SWELLING, AND PIOUS AMOUNTS OF FLUID BUILDUP. NO FURTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION REQUESTED ON 1/23/2025. ADDITIONAL INFORMATION RECEIVED ON 1/23/025: DURING THE INITIAL SURGERY ON (B)(6) 2020, AN AR-9106-01 ARTHREX UNIVERS VAULTLOCK GLENOID, AN AR-9301-02 ARTHREX ECLIPSE CAGE SCREW, AND AN AR-9301-43CPC ARTHREX ECLIPSE TRUNION WERE IMPLANTED. THE PATIENT STARTED TO EXPERIENCE PAIN AND SOUGHT OUT MEDICAL TREATMENT. THE PATIENT UNDERWENT TWO DEBRIDEMENT PROCEDURES, IN WHICH THE MISSING PEG WAS DISCOVERED. THE THIRD PROCEDURE WAS THE REVERS TOTAL SHOULDER REPLACEMENT WHICH OCCURRED ON (B)(6)2024.

    Patient problems: Failure of Implant; Pain

    MAUDE report 21337004 ↗
  • MalfunctionFeb 8, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2022, A NOTIFICATION WAS RECEIVED INDICATING THAT A PATIENT, LISTED ON THE SHOULDER ARTHROPLASTY REGISTRY, IS EXPERIENCING WORSENING PAIN. AN X-RAY SHOWS A DISLODGED GLENOID COMPONENT. SURGEON HAS SCHEDULED THE PATIENT FOR A REVISION SURGERY TO AN RSA ON (B)(6) 2022. ORIGINAL SURGERY TOOK PLACE ON (B)(6) 2020. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2022: DURING SURGERY ON (B)(6) 2020, A TOTAL OF 4 ARTHREX DEVICES WERE IMPLANTED. AN AR-9301-03 ECLIPSE CAGE SCREW LARGE, LOT NUMBER 19.00501, AN AR-9301-43CPC ECLIPSE TRUNION, LOT NUMBER 18-01245, AN AR-9343-16 ECLIPSE HUMERAL HEAD, LOT NUMBER 1917001, AND AN AR-9106-01 UNIVERS VAULTLOCK GLENOID. AS OF NOW, NOTHING HAS SHOWN TO BE BROKEN INSIDE THE PATIENT BUT THERE'S AS DISLODGMENT OF GLENOID PROSTHESIS. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2022: ACCORDING TO FACILITY REPRESENTATIVE, PATIENT EXPERIENCE CARDIAC EVENTS DURING THE PRE-OP AREA AND REVISION SURGERY DID NOT TAKE PLACE. AT THIS TIME, IT IS UNKNOWN IF PATIENT WILL UNDERGO REVISION SURGERY. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2022: THE REVISION SURGERY THAT WAS TO TAKE PLACE ON (B)(6) 2022, WAS CANCELLED DUE TO CARDIAC CO-MORBIDITIES. NO FURTHER INFORMAT

    Patient problems: Pain

    MAUDE report 18666775 ↗
  • InjurySep 21, 2020Use of Device Problem

    IT WAS REPORTED THAT DURING AN ARTHROPLASTY TOTAL SHOULDER USING THE ECLIPSE, WHEN USING THE CENTRAL SCREW ON THE TRUNION, THE SCREW LOCKED UP, WHICH REMOVED THE CENTRAL CORE OF BONE. THEY USED THE APEX STEM TO COMPLETED THE CASE. ADDITIONAL INFORMATION OBTAINED 9/15/20: WHEN SCREWING THE CENTRAL SCREW IN THE TWO COMPONENTS LOCKED CAUSING THE TRUNION TO SPIN AND THE CORE OF BONE TO COME OFF IN THE CENTRAL SCREW. THE SURGEON THEN CHOSE TO ABORT USING THE PLANNED ECLIPSE AND INSTEAD IMPLANTED AN APEX STEM TO COMPLETE THE PROCEDURE.

    Patient problems: Insufficient Information

    MAUDE report 10563150 ↗

FDA clearance and classification

The FDA regulates product code QHQ (total shoulder arthroplasty system) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the Arthrex® AR-9301-43CPC

Has the Arthrex® AR-9301-43CPC been recalled?

No FDA recall record in our data names this exact device identifier (DI 00888867276147), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Arthrex® AR-9301-43CPC?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Sep 21, 2020 and Feb 7, 2025. In those reports, the most common specific device problems are “Break” (1), “Device Dislodged or Dislocated” (1) and “Use of Device Problem” (1); the most common patient problems are “Pain” (2) and “Failure of Implant” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Arthrex® AR-9301-43CPC?

None of the reports are classified as deaths, 2 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Arthrex® AR-9301-43CPC MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Arthrex® AR-9301-43CPC, and is it still sold?

The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-9301-43CPC). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Total shoulder arthroplasty system devices

All Total shoulder arthroplasty system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page