Arthrex® AR-9345-17Recalls & reported problems

Arthrex, Inc. · ARTHREX ECLIPSE HUMERAL HEAD, 45/17 · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867060029), received between Jun 28, 2017 and Mar 12, 2025 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem3
  • Adverse Event Without Identified Device or Use Problem2
  • Degraded1
  • Device Dislodged or Dislocated1
  • Failure To Adhere Or Bond1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Failure of Implant2
  • Foreign Body Reaction2
  • Hypersensitivity/Allergic reaction1
  • Metal Related Pathology1
  • Osteolysis1
  • Purulent Discharge1
  • Reaction1
  • Rupture1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 12, 2025Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem

    ON 2/17/2025, A CLINDEX NOTIFICATION WAS RECEIVED INDICATING THAT A PATIENT LISTED ON THE SHOULDER ARTHROPLASTY REGISTRY SUFFERED A RUPTURED SUBSCAPULARIS DUE TO A TRAUMATIC FALL. THE PATIENT UNDERWENT REVISION SURGERY OF THE ANATOMIC SHOULDER REPLACEMENT TO A REVERSE SHOULDER ARTHROPLASTY ON (B)(6) 2023. THE EVENT WAS INDICATED AS UNRELATED TO THE ECLIPSE SYSTEM IMPLANTED ON (B)(6) 2023. ADDITIONAL INFORMATION RECEIVED ON 2/27/2025: ON (B)(6) 2023, THE PATIENT UNDERWENT REVISION SURGERY TO HAVE AN AR-9301-45CPC ARTHREX ECLIPSE TRUNION, AN AR-9106-02 ARTHREX UNIVERS VAULTLOCK GLENOID, AN AR-9301-03 ARTHREX ECLIPSE CAGE SCREW LARGE, AND AN AR-9345-17 ARTHREX ECLIPSE HUMERAL HEAD REMOVED. AFTER THE SURGERY, THE PATIENT IS DOING WELL WITH AN EIGHTY-FIVE TO NINETY PERCENT IMPROVEMENT IN SYMPTOMS.

    Patient problems: Failure of Implant; Rupture

    MAUDE report 21586172 ↗
  • InjuryJan 8, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    ON (B)(6) 2024, IT WAS REPORTED BY THE PATIENT VIA PHONE THAT THE PATIENT HAD THE AR-9345-17 ARTHREX ECLIPSE¿ HUMERAL HEAD, 45/17 EXPLANTED FROM THE PATIENT'S RIGHT SHOULDER FOR A REVISION PROCEDURE ON (B)(6) 2024 DUE TO A BAD ALLERGIC REACTION WITH THE PRODUCT. THE PATIENT HAD VISIBLE ROUND RASHES SPREADING FROM THE EARS, FACE, INNER ELBOWS, AND UPPER CHEST AND THEN WENT TO THE PATIENT'S RIGHT SHOULDER, STOMACH, AND INNER EAR CANAL AND BECAME A SYSTEMIC RASH CAUSING BOWEL ISSUES. THE ORIGINAL PROCEDURE FOR THE RIGHT SHOULDER ARTHROSCOPIC STEMLESS REPLACEMENT WAS ON (B)(6) 2022. THE ORIGINAL AND REVISION PROCEDURES WERE PERFORMED WITH THE SAME SURGEON IN THE SAME FACILITY. THE PATIENT THEN INQUIRED ABOUT THE MATERIAL COMPOSITION OF THE AR-9345-17T ARTHREX ECLIPSE¿ TITANIUM HUMERAL HEAD 45/17 IMPLANTED IN THE REVISION SURGERY. THE PATIENT WOULD LIKE TO KNOW IF THE TITANIUM HAS NICKEL OR OTHER METALS. THE PATIENT HAS A NICKEL ALLERGY AND IS CONCERNED ABOUT HAVING ANOTHER REACTION LIKE BEFORE.

    Patient problems: Foreign Body Reaction; Hypersensitivity/Allergic reaction

    MAUDE report 21102517 ↗
  • InjuryAug 9, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A LOOSENING OF THE PEGGED GLENOID LED TO AN EXPLANTATION OF THE ANATOMIC ECLIPSE PROSTHESIS AND THE PEGGED GLENOID. NO FURTHER INFORMATION WAS PROVIDED. UPDATE AVOE (B)(6) 2023: IT WAS CONFIRMED THAT THE INITIAL SURGERY TOOK PLACE IN 2015 IN THE PRIVATKLINIK BETHANIEN, 8044 ZÜRICH. AFTER THE ECLIPSE AND THE PEGGED GLENOID WERE EXPLANTED, THESE DEVICES WERE REPLACED WITH DEVICES FROM ANOTHER MANUFACTURER.

    Patient problems: Failure of Implant

    MAUDE report 17486646 ↗
  • MalfunctionMar 21, 2022Degraded; Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A REVISION SURGERY WAS NECESSARY DUE TO A PROGRESSIVE POLYETHYLENE DEGRADATION WITH METALLOSIS. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Metal Related Pathology

    MAUDE report 13839585 ↗
  • InjuryOct 23, 2019Adverse Event Without Identified Device or Use Problem

    IT WAS DISCOVERED DURING SOCIAL SAFEGUARD REVIEW THAT THE FOLLOWING INCIDENT WAS MENTIONED IN AN ARTHREX CHATTER COMMUNITY: ON (B)(6) 2019 A (B)(6) YEAR OLD PATIENT UNDERWENT A REVISION PROCEDURE WHERE AN ARTHREX ECLIPSE AR-9345-17 (LOT 1295172500) WAS EXPLANTED. THE PATIENT HAD A LONG SURGICAL HISTORY INVOLVING THIS SHOULDER. INCLUDING AN OPEN GLENOHUMERAL OPERATION THAT REQUIRED HIS SUBSCAPULARIS BEING TAKEN DOWN. THE SUBSCAP REPAIR DID NOT INVOLVE ARTHREX PRODUCTS AND WAS REPORTED TO HAVE FAILED. THE ARTHREX ECLIPSE DEVICE HAD BEEN IMPLANTED DURING A 2018 PROCEDURE WHICH WAS PERFORMED IN KUWAIT. THERE IS NO ADDITIONAL INFORMATION AVAILABLE WITH REGARD TO THE ORIGINAL ARTHREX IMPLANT PROCEDURE. DURING THE (B)(6) 2019 PROCEDURE THE ARTHREX ECLIPSE DEVICE WAS FOUND TO BE HAVE A SOLID FIXATION IN THE PATIENT BONE. THE DEVICE WAS ABLE TO BE EXPLANTED WITH ONLY THE MINIMAL BONE LOSS EQUAL TO THE SPACE OF THE 18MM CAGE SCREW. DURING THE PROCEDURE THE CAGE SCREW, REUNION AND GLENOID WERE ALSO REMOVED HOWEVER THERE IS NO INFORMATION AVAILABLE WITH REGARD TO THE PART NUMBERS OF THESE DEVICES. THE DEVICES WILL NOT BE RETURNING TO ARTHREX FOR EVALUATION BECAUSE THEY WERE REPORTED TO HAVE BE

    Patient problems: Osteolysis; Tissue Damage

    MAUDE report 9225661 ↗
  • InjuryJun 28, 2017Failure To Adhere Or Bond; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT RECEIVED AN ECLIPSE IMPLANT INTO THE LEFT SHOULDER WITH THE FOLLOWING DEVICES USED DURING PROCEDURE: AR-9345-17, LOT 1611019 ((B)(4)); AR-9300-45CPC, LOT 150083212((B)(4)); AR-9301-02, LOT 150073614 ((B)(4)); AND AR-1928SF-3, LOT 10020391 ((B)(4)); ALONG WITH A NON ARTHREX GLENOID. THE IMPLANTS GOT DISLOCATED DUE TO A TENDON THAT DID NOT ADHERE IT. THE DECISION WAS MADE TO REVISE IT WITH A REVERSE IMPLANT IN 2017. DURING THE REVISION SURGERY WHEN THE SHOULDER WAS OPENED THE SURGEON FOUND PUS AND THE ECLIPSE IMPLANTS WERE LOOSE IN THE JOINT. PATHOLOGICAL INVESTIGATION DID NOT FIND ANY LAB CULTURE BUT A FOREIGN BODY REACTION /ALLERGIC REACTION WAS DIAGNOSED. ADDITIONAL INFORMATION RECEIVED ON 13 JUN 2017: INFORMATION RECEIVED THAT THE COMPLAINT DEVICES WILL NOT BE RETURNED AS THEY WERE DISCARDED AT THE FACILITY. DATE OF THE INITIAL SURGERY WAS ON (B)(6) 2017, DATE OF SECOND SURGERY/REVISION WAS ON (B)(6) 2017.

    Patient problems: Foreign Body Reaction; Purulent Discharge; Reaction

    MAUDE report 6672682 ↗

FDA clearance and classification

The FDA regulates product code QHQ (total shoulder arthroplasty system) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the Arthrex® AR-9345-17

Has the Arthrex® AR-9345-17 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00888867060029), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Arthrex® AR-9345-17?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jun 28, 2017 and Mar 12, 2025. In those reports, the most common specific device problems are “Patient Device Interaction Problem” (3), “Degraded” (1) and “Device Dislodged or Dislocated” (1); the most common patient problems are “Failure of Implant” (2), “Foreign Body Reaction” (2) and “Hypersensitivity/Allergic reaction” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Arthrex® AR-9345-17?

None of the reports are classified as deaths, 5 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Arthrex® AR-9345-17 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Arthrex® AR-9345-17, and is it still sold?

The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-9345-17). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Total shoulder arthroplasty system devices