Skip to content

Arthrex® AR-9531Recalls & reported problems

Arthrex, Inc. · UNIVERS REVERS HUMERAL LINER IMPACTOR · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00888867062184), received between Apr 11, 2022 and Jun 10, 2024 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Naturally Worn2
  • Crack1
  • Defective Component1
  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Insufficient Information1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 10, 2024Break; Break

    ON 01/31/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT A AR-9531 HUMERAL LINER IMPACTOR BROKE. THIS OCCURRED DURING A CASE, WITH NO PATIENT EFFECT REPORTED. NO ADDITIONAL INFORMATION PROVIDED. ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED 01/30/2023: THE PROCEDURE PERFORMED WAS RSA ON (B)(6) 2023. THIS WAS DISCOVERED DURING THE IMPLANTATION OF HUMERAL CUP POLY WHICH IS THE LAST PART OF THE CASE. THE BONE QUALITY OF THE PATIENT WAS GOOD. NO FRAGMENTS BROKE INSIDE OF THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19498620 ↗
  • MalfunctionJun 7, 2024Crack; Crack

    IT WAS REPORTED DURING A REVERSE TOTAL SHOULDER PROCEDURE, WHEN THE SURGEON HIT THE AR-9531, IMPACTOR WITH THE MALLET THE TIP CRACKED. THE SURGEON USED A SECOND AR-9533, IMPACTOR TO COMPLETE THE PROCEDURE. WHEN THE CASE WAS COMPLETED THE AR-9592-2420, WAS STICKING TOGETHER NOT COMING APART FOR STERILIZATION.

    Patient problems: Insufficient Information

    MAUDE report 19490854 ↗
  • MalfunctionJun 5, 2024Break; Break

    IT WAS REPORTED ON 10/28/2022 BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9531 HUMERAL LINER IMPACTOR PLASTIC TIP BROKE OFF. CASE INVOLVEMENT, NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19471793 ↗
  • MalfunctionMay 20, 2024Break; Break

    IT WAS REPORTED ON 10/03/2022 BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-9531 HUMERAL LINER IMPACTOR CAP IS BROKEN OFF. THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19354299 ↗
  • MalfunctionFeb 13, 2024Naturally Worn; Naturally Worn

    ON 01/22/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (2) AR-9531 HUMERAL LINER IMPACTORS PLASTIC TIPS WORE OUT AND KEPT FALLING OFF THE HANDLE. THIS OCCURRED DURING A CASE, WITH NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18697159 ↗
  • MalfunctionFeb 13, 2024Naturally Worn; Naturally Worn

    ON 01/22/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (2) AR-9531 HUMERAL LINER IMPACTORS PLASTIC TIPS WORE OUT AND KEPT FALLING OFF THE HANDLE. THIS OCCURRED DURING A CASE, WITH NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18697106 ↗
  • MalfunctionMay 10, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-9531 HUMERAL LINER IMPACTOR HEAD FELL OFF. THIS OCCURRED DURING AN RTSA ON (B)(6) 2023 THE HEAD FELL OFF DUE TO THE THREADS BEING STRIPPED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONAL INFORMATION WAS PROVIDED ON (04/18/2023, IT WAS REPORTED THAT ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. THE CASE WAS COMPLETED, AND THERE WAS NO PATIENT EFFECT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16906325 ↗
  • MalfunctionApr 11, 2022Defective Component

    IT WAS REPORTED THAT DURING A SHOULDER PROSTHESIS SURGERY THE WHITE CAP OF THE DEVICE WAS DEFECT. THE THREAD COULDN`T BE SCREWED. NO PART OF THE DEVICE BROKE OFF. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE 28-MAR-2022: IT WAS CONFIRMED THAT THE WHITE CAP OF THE DEVICE BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14075386 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates