Skip to content

Arthrex® AR-1927BCF-3-1Recalls & reported problems

Arthrex, Inc. · SUTURE ANCHOR, BIOCOMPOSITE CORKSCREW · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888867096790), received between May 10, 2021 and Feb 24, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Appropriate Device Problem Term/Code Not Available1
  • Crack1
  • Entrapment of Device1
  • Material Separation1

Patient problems

  • Foreign Body In Patient3
  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2025Break; Break

    ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-1927BCF-3-1 SUTURE ANCHOR, BIOCOMPOSITE CORKSCREW WAS REPORTED TO HAVE FAILED DURING USE. THIS OCCURRED ON (B)(6) 2025 IN ROYSTON HOSPITAL DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR. IT WAS REPORTED THAT A 5.5 CORKSCREW WAS IMPLANTED BUT THE INCORRECT PUNCH WAS USED TO PREPARE THE BONE PRIOR. THE TIP OF THE CORKSCREW CAME AWAY AND REMAINS WITHIN THE PATIENT. THE CASE WAS COMPLETED SUCCESSFULLY USING ANOTHER ANCHOR. THERE WAS CASE INVOLVEMENT.

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 21445852 ↗
  • MalfunctionOct 25, 2024Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE ANCHOR BROKE IN HALF. ALL PARTS WERE RETRIEVED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20538689 ↗
  • MalfunctionJul 11, 2022Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE ANCHOR OF THE DEVICE BROKE. ALL BROKEN PARTS WERE RETRIEVED FROM THE PATIENT. THIS RESULTED IN A PROLONGED SURGERY TIME. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 14976107 ↗
  • MalfunctionNov 2, 2021Entrapment of Device; Material Separation

    IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE DEVICE BROKE INSIDE THE PATIENT. THE BROKEN PIECE WASN`T RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 12737955 ↗
  • MalfunctionSep 14, 2021Break; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING A SURGERY THE ANCHOR BROKE OFF INSIDE THE PATIENT WHEN SCREWING IN THE HUMERUS TO SECURE THE SUTURES OF THE CAP REPAIR ASSEMBLY. NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY WAS REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 12467461 ↗
  • MalfunctionMay 10, 2021Break

    IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY FOR A ROTATOR CUFF SUTURE THE METAL IMPLANT HOLDER OF THE DEVICE BROKE IN THE HUMERAL HEAD WHEN THE SURGEON TRIED TO SCREW IT IN. THE ANCHOR HOLE HAD TO BE ENLARGED TO REMOVE THE DEVICE. ALL BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11798802 ↗
  • MalfunctionMay 10, 2021Break; Crack

    IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY FOR A ROTATOR CUFF SUTURE TWO DEVICES WERE TESTED WHICH CRACKED DURING INSERTION. ALL BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11798798 ↗

Other Fastener, fixation, biodegradable, soft tissue devices

All Fastener, fixation, biodegradable, soft tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page