Skip to content

Arthrex® AR-1927BCFT-45Recalls & reported problems

Arthrex, Inc. · 4.5 BIOCOMPOSITE FT WITH TIGERTAIL · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867096844), received between Oct 23, 2020 and Feb 10, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 4 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Crack2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 10, 2025Break; Break

    ON (B)(6) 2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-1927BCFT-45, 4.5 BIOCOMPOSITE FT WITH TIGERTAIL THE IMPLANT BROKE. THIS WAS DETECTED DURING A LEFT SHOULDER ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2025. THE IMPLANT BROKE AS SURGEON WAS LIGHTLY GRAZING THE ANCHOR ON BONE SURFACE TO LOOK FOR THE HOLE. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO PATIENT HARM. THE BROKEN IMPLANT WAS REMOVED FROM THE OPERATION SITE FULLY INTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21346176 ↗
  • MalfunctionFeb 10, 2025Break; Break

    ON 28TH JANUARY 2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-1927BCFT-45, 4.5 BIOCOMPOSITE FT WITH TIGERTAIL THE IMPLANT BROKE. THIS WAS DETECTED DURING A LEFT SHOULDER ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2025. THE IMPLANT BROKE AS SURGEON WAS LIGHTLY GRAZING THE ANCHOR ON BONE SURFACE TO LOOK FOR THE HOLE. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO PATIENT HARM. THE BROKEN IMPLANT WAS REMOVED FROM THE OPERATION SITE FULLY INTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21346107 ↗
  • MalfunctionFeb 3, 2025Break; Crack; Break; Crack

    IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE DEVICE CRACKED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. NO FURTHER INFORMATION RECEIVED. UPDATE AVOE 22-JAN-2025: IT WAS CONFIRMED THAT THE ERROR OCCURRED DURING A SINGLE MEDIAL ROW ROTATOR CUFF REPAIR. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21294799 ↗
  • MalfunctionFeb 3, 2025Break; Crack; Break; Crack

    IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE DEVICE CRACKED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. NO FURTHER INFORMATION RECEIVED. UPDATE AVOE 22-JAN-2025. IT WAS CONFIRMED THAT THE ERROR OCCURRED DURING A SINGLE MEDIAL ROW ROTATOR CUFF REPAIR. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21294772 ↗
  • InjuryOct 23, 2020Break

    IT WAS REPORTED THAT DURING A REMPLISSAGE PROCEDURE THE AR-1927BCFT-45 (BIOCOMPOSITE FT WITH TIGERTAIL) WAS PLACED. WHEN THE SURGEON PULLED ON THE SUTURES TO TEST IT'S INTEGRITY THE EYELET BROKE AND THE REPAIR SUTURES BECAME UNUSABLE. THE SURGEON RAN OUT OF BONE AND THE CHOSE TO BAIL ON THE REMPLISSAGE. THE ANCHOR BODY REMAINED IN THE PATIENT. ADDITIONAL INFORMATION OBTAINED 10/8/2020: PATIENT BONE QUALITY IS REPORTED TO HAVE BEEN POOR. THE BONE HAD BEEN PREPARED WITH THE 3.5 PUSHLOCK PUNCH FOR THE 4.5 CORKSCREW. THE ANCHOR BODY WAS LEFT IN THE PATIENT. ALL FOUR SUTURE TAILS HAD BEEN PULLED ON TO TEST HE STRENGTH OF THE ANCHOR AND THE SULTURES PULLED OUT OF THE ANCHOR. REP WHO WAS PRESENT BELIEVES THE EYELET MAY HAVE BROKE. REP STATES THE EYELET IS NOT FLOATING OR IN THE PATIENT BUT IS UNSURE WHERE THE SUTURE EYELET IS. SUTURES WERE REMOVED AND DISCARDED AFTER THE CASE WAS OVER. A KNOTLESS SUTURETAK PULLED OUT OF THE BONE PRIOR TO THE CORKSCREW BEING INSERTED. THE SUTURETAK PULLING OUT IS REPORTED TO HAVE BEEN THE RESULT OF THE POOR BONE QUALITY AND, PER THE REP WHO WAS PRESENT, THE SURGEON BEING OVER ZEALOUS WITH HIS TUGS. THE SUTURETAK ANCHOR ITSELF HAD NO ISSUES. SINCE THE CORKS

    Patient problems: Foreign Body In Patient

    MAUDE report 10730344 ↗

Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices →

Sources and updates