Skip to content

Arthrex® AR-2324PCTTRecalls & reported problems

Arthrex, Inc. · PEEK, SWIVELOCK TT, 4.75x19.1MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867098121), received between Jul 10, 2023 and Sep 8, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1
  • Failure to Advance1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 8, 2026Detachment of Device or Device Component

    Patient problems: Insufficient Information

    MAUDE report 50670232 ↗
  • MalfunctionSep 19, 2025Failure to Advance; Failure to Advance

    ON 08/29/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-2324 PCTT PEEK SWIVELOCK SUTURE MALFUNCTIONED DURING A ROTATOR CUFF REPAIR PROCEDURE UTILIZING A SPEED BRIDGE TECHNIQUE. THE PATIENT'S BONE QUALITY WAS ASSESSED AS GOOD, SO THE SURGICAL TEAM OPTED TO USE THE SILVER PUNCH. AFTER 5 TO 6 TURNS OF THE DRIVER, THE ANCHOR DID NOT APPEAR TO BE ADVANCING. THE SURGEON REMOVED THE ANCHOR TO REPUNCH. AT THAT POINT, THE SCRUB TECHNICIAN ATTEMPTED TO TWIST THE PADDLE SO IT WOULD SIT FLUSH WITH THE KNOB, BUT WAS UNABLE TO ADVANCE OR RETRACT THE PADDLE. IT SIMPLY SPUN AND PRODUCED A CLICKING SOUND. A SECOND AR-2324 PCTT WAS OPENED AND IMPLANTED WITHOUT ISSUE. THIS WAS DISCOVERED DURING USE ON (B)(6) 2025, WITH NO PATIENT HARM REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON 09/08/2025. ADDITIONAL INFORMATION RECEIVED ON 9/10/2025: THE PROCEDURE WAS DELAYED ABOUT FIVE MINUTES AS THE SURGEON WAS TRYING TO REVERSE THE PADDLE AND SALVAGE THE ANCHOR. NO ADDITIONAL ANESTHESIA WAS NEEDED. THE CASE WAS COMPLETED USING ANOTHER AR-2324PCTT PEEK SWIVELOCK SUTURE FORM A DIFFERENT LOT NUMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23099531 ↗
  • MalfunctionJul 10, 2023Break; Break

    ADDITIONAL INFORMATION WAS PROVIDED ON 07/28/2023, IT WAS REPORTED THAT THIS OCCURRED IN A PROCEDURE FOR A LIGAMENT INJURY THAT WAS CORRECTED. ALL OF THE FRAGMENTS WERE RECOVERED, AND NO REMAINS WERE LEFT IN THE PATIENT, ANOTHER ANCHOR WAS OPENED, AND HE COMPLETED THE SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 17286708 ↗