Skip to content

Arthrex® AR-613-48Recalls & reported problems

Arthrex, Inc. · iBALANCE TKA, MODULAR T-HANDLE · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888867116498), received between Apr 13, 2021 and Dec 6, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Crack2
  • Device Damaged by Another Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information2
  • Failure of Implant1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 6, 2024Break; Break

    ON 11/08/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-613-48 IBALANCE¿ TKA, MODULAR T-HANDLE BROKE IN HALF. THE PATIENT WAS NOT AFFECTED. THIS WAS DISCOVERED DURING A KNEE PROCEDURE, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20864179 ↗
  • MalfunctionMay 23, 2024Break; Break

    ON 09/29/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT A AR-613-48 HANDLE, AND A AR-613-1 HANDLE BROKE. THIS OCCURRED DURING AN UNKNOWN CASE WITH NO PATIENT EFFECT REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19387635 ↗
  • MalfunctionDec 6, 2022Break

    IT WAS REPORTED ON (B)(6) 2022 BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-613-48 T-HANDLE WAS HIT WITH A MALLET AND BROKE INTO PIECES. CASE INVOLVEMENT, NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15922885 ↗
  • MalfunctionJul 5, 2022Crack

    ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT A AR-613-48 T-HANDLE CRACKED IN HALF. THIS OCCURRED DURING A TKA PROCEDURE, ALL FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED, WITH NO PATIENT EFFECT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14915134 ↗
  • MalfunctionJun 20, 2022Break

    ON (B)(6) 2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-613-48 T-HANDLE BROKE IN HALF WHEN SURGEON ATTEMPTED TO REMOVED THE IM ROD. THIS WAS DISCOVERED DURING A TKA PROCEDURE ON (B)(6) 2021. CASE WAS COMPLETED SUCCESSFULLY WITHOUT FURTHER ISSUES.

    Patient problems: Insufficient Information

    MAUDE report 14743877 ↗
  • MalfunctionJun 14, 2021Crack; Device Damaged by Another Device

    IT WAS REPORTED DURING AN TKA PROCEDURE THE AR-613-48, T-HANDLE CRACKED IN HALF AFTER, THE SURGEON STRUCK IT WITH A MALLET. THE HOSPITAL DISCARDED THE TOOL.

    Patient problems: Insufficient Information

    MAUDE report 11991347 ↗
  • MalfunctionApr 13, 2021Break

    IT WAS REPORTED THAT THE AR-613-48 IS SHATTERED.

    Patient problems: Failure of Implant

    MAUDE report 11655714 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices