Arthrex® AR-6435Recalls & reported problems
Arthrex, Inc. · DW OUTFLOW TUBE SET W/REDEUCE TUBING · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867118089), received between Oct 3, 2023 and Jun 14, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Flow or Under Infusion2
- Break1
- Detachment of Device or Device Component1
- Infusion or Flow Problem1
- Partial Blockage1
- Pumping Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Extravasation1
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 14, 2024Partial Blockage; Partial Blockage
ON 05/10/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-6435 OUTFLOW SET HAS A CLOG BEHIND THE OUTFLOW CASSETTE. THIS WAS THE END EFFECT, AS THE CLOG ALMOST WRAPPED THE ENTIRE CURVE AND COULD NOT BE RELIEVED. MINOR CASE DELAYS NO PATIENT ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19537867 ↗MalfunctionJun 8, 2024Break; Break
ON 05/18/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-6435 OUTFLOW SET IS BROKEN. THIS WAS DISCOVERED DURING AN UNSPECIFIED TIME WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19494682 ↗MalfunctionJun 6, 2024Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion
ON 09/23/2021 IT WAS REPORTED BY A SALE REPRESENTATIVE VIA SEMS THAT AN AR-6435 DW OUTFLOW TUBING WAS NOT WORKING PROPERLY, OUTFLOW WAS NOT DETECTED. THIS WAS DISCOVERED DURING A PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: Insufficient Information
MAUDE report 19477322 ↗MalfunctionFeb 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ON 2/8/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6435 DW OUTFLOW TUBE SET TUBING LOOKED LIKE IT WAS STILL ON THE SHAVER. HOWEVER, WHEN THEY WENT TO MOVE IT, THE PIECE OF TUBING THAT CONNECTS TO THE SHAVER WAS ON IT BUT THE WHOLE TUBING WAS COMPLETELY OUT AS IF THE SEALANT CAME APART. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18799201 ↗MalfunctionFeb 6, 2024Pumping Problem; Pumping Problem
IT WAS REPORTED THAT DURING A KNEE SURGERY THE PUMP SPEED INCREASES UNCONTROLLABLY. THE CUSTOMER HAS REPORTED THAT THE ISSUE COULD BE SOLVED WITH THE THIRD TUBING SET AND FINISHED THE SURGERY SUCCESSFULLY. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE 04-FEB-2021 DW IT WAS FURTHER REPORTED THAT THE PATIENT WAS ALREADY IN THE SURGERY ROOM BUT THE RELATED PUMP WAS NOT USED ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18653230 ↗MalfunctionOct 3, 2023Insufficient Flow or Under Infusion; Infusion or Flow Problem; Insufficient Flow or Under Infusion; Infusion or Flow Problem
ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06021994 THAT AN AR-6480 DW ARTHROSCOPY FLUID MANAGEMENT DEVICE HAD AN ISSUE. THE SURGEON NOTICED THE PUMP'S OUTFLOW WAS SLOW TO RESPOND, NOT PULLING FLUID FAST ENOUGH. THE PATIENT WAS NOT AFFECTED. THIS OCCURRED DURING USE IN AN UNSPECIFIED PROCEDURE WITH NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ON 9/7/2023, THE SALES REPRESENTATIVE PROVIDED THE FOLLOWING INFORMATION VIA EMAIL AND PHONE: THIS OCCURRED DURING A SHOULDER ARTHROSCOPY PROCEDURE ON (B)(6) 2023. AN AR-6411 REDEUCE PUMP TUBING AND AR-6435 DUALWAVE OUTFLOW TUBE SET WITH REDEUCE TUBING SYSTEM WITH UNKNOWN LOT NUMBERS WERE USED. THE PUMP SETTINGS WERE AT 45MMHG, AND THE PUMP WAS NOT KEEPING UP WITH THE PRESSURE. SLIGHT EXTRAVASATION OCCURRED, IT WAS LOCALIZED AND A MANAGEABLE ISSUE. THEY USED A RINSE FUNCTION W/ CLOSED INFLOW PORTAL TO REMOVE THE EXCESS FLUID. THE PUMP WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THE PATIENT WAS DISCHARGED AFTER THE PROCEDURE AND IS FINE TO DATE, THERE WAS NO HARM REPORTED AND NO CASE DELAY. THE TUBING SETS WERE DISCARDED AFTER THE PROCEDURE.
Patient problems: Extravasation
MAUDE report 17862871 ↗
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page