Arthrex® AR-8725-26HRecalls & reported problems

Arthrex, Inc. · COMPR FT SCRW,2.5 MICRO,26MM LGTH · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867126916), received between Sep 19, 2016 and Dec 27, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component1
  • Material Fragmentation1

Patient problems

  • Foreign Body In Patient2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 27, 2024Material Fragmentation; Material Fragmentation

    ON (B)(6) 2024 IT WAS REPORTED BY A SALES REPRESENTATIVE, VIA PHONE, THAT AN AR-8725-26H MICRO COMPRESSION FT DE-THREADED DURING INSERTION AND FRAGMENTS BROKE OFF THE SCREW. ONE OF THE METAL SHAVINGS COULD NOT BE REMOVED FROM THE PATIENT. THE CASE WAS COMPLETED BY REMOVING THE SCREW AND USING A NEW AR-8725-26H MICRO COMPRESSION FT. THIS WAS DISCOVERED DURING A PIP HAMMERTOE PROCEDURE ON (B)(6)2024. THE CASE WAS NOT DELAYED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21027203 ↗
  • InjuryJul 16, 2018Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DIPJ FUSION OF THE INDEX FINGER, THE SURGEON WAS ATTEMPTING TO INSERT A 2.5 MICRO COMPRESSION SCREW, (B)(4), ONE OF THE CUTTING FLUTES BROKE OFF OF THE DISTAL END OF THE SCREW. THE SURGEON USED A DOUBLE ENDED K-WIRE AND A PROFILE DRILL FOR MICRO HCS AND A 2.0 DRILL BIT TO A DEPTH OF 24MM TO PREPARE THE BONE. THE SMALL CUTTING FLUTE WAS LODGED IN THE CANAL OF THE PROXIMAL PHALANX OF THE INDEX FINGER AND COULD NOT BE RETRIEVED, HOWEVER, THE SURGEON WAS NOT CONCERNED OF THE PIECE FLOATING AROUND. ADDITIONAL INFORMATION OBTAINED. THE CASE WAS COMPLETED JUST AS PLANNED. THE SURGEON USED A SECOND 2.5 MICRO HCS AND PLACED IT OVER THE BROKEN HCS.

    Patient problems: Foreign Body In Patient

    MAUDE report 7691730 ↗
  • InjurySep 19, 2016Break

    IT WAS REPORTED THAT DURING A DISTAL INTERPHALANGEAL JOINT FINGER PROCEDURE THE COMPRESSION SCREW BROKE ON INSERTION. THE SURGEON BACKED OUT THE LOOSE BROKEN PIECE OF THE COMPRESSION SCREW AND REMOVED IT FROM THE PATIENT. THE SURGEON MADE THE DECISION THAT THE PARTIAL PART OF THE SCREW THAT WAS INSERTED WAS SECURE ENOUGH TO COMPLETE THE CASE WITHOUT ADDITIONAL IMPLANTATION. LOOSE PIECE THAT WAS RETRIEVED WAS DISCARDED BY THE FACILITY. SURGEON HAD DRILLED WITH A 2.0 CANNULATED DRILL, FOLLOWED BY THE MICRO COMPRESSION FT PROFILE DRILL.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5960132 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →