Skip to content

Arthrex® AR-8725-28HRecalls & reported problems

Arthrex, Inc. · COMPR FT SCRW,2.5 MICRO,28MM LGTH · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867126923), received between Sep 10, 2020 and Sep 10, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 10, 2020Break

    IT WAS REPORTED, THAT DURING A TALUS FRACTURE REPAIR, AFTER INSERTION OF THE SECOND SCREW THE TIP OF THE AR-8737-37 BROKE OFF INSIDE THE SCREW HEAD. ALL BROKEN PIECES WERE REMOVED AND AN AR-8737-61 (EASY OUT) WAS USED AND THE SCREW WAS STRIPPED. THE SURGEON COMPLETED THE CASE BY USING A BURR AND SHAVING THE AR-8725-28H SCREW (LOT 10061591) DOWN FLUSH WITH THE BONE.

    Patient problems: Failure of Implant

    MAUDE report 10513517 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page