Skip to content

Arthrex® AR-8740-34HRecalls & reported problems

Arthrex, Inc. · COMPR FT SCRW,4.0 STD,34MM LGTH · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867127722), received between Jun 27, 2019 and Jun 27, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Material Twisted/Bent1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 27, 2019Break; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A PROCEDURE, THE SURGEON WAS ATTEMPTING TO REMOVE A 4.0 HEADLESS SCREW FROM THE PROXIMAL TIBIA WITH A T10 HEXALOBE DRIVER, AR-8737-38, AND THE TIP BROKE OFF INTO THE HEAD OF THE SCREW. THE TIP WAS ABLE TO BE REMOVED FROM THE HEAD WITH NO ISSUE. THE SURGEON THEN USED A CANNULATED T10 HEXALOBE DRIVER, AR-8737-26, TO FINISH THE REMOVAL OF THE TIP OF THE DRIVER WHICH BECAME TWISTED. THE SCREW WAS ABLE TO BE REMOVED WITH THE SECOND DRIVER WITH NO ISSUE TO THE PATIENT. ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED (B)(6) 2019: THE PATIENT UNDERWENT A PROCEDURE ON (B)(6) 2019 WHERE A 4.0 HEADLESS SCREW, AR-8740-34H, WAS IMPLANTED. SHORTLY AFTER, THE PATIENT BEGAN EXPERIENCING PAIN AND HAD TO HAVE A SECOND SURGERY TO HAVE THE SCREW REMOVED ON (B)(6) 2019. PAIN IS THE ONLY SYMPTOM THE PATIENT WAS EXPERIENCING AND IT WAS THE ONLY DEVICE REMOVED DURING THE REVISION PROCEDURE. NO HARDWARE WAS PUT IN ITS PLACE.

    Patient problems: Pain

    MAUDE report 8739448 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page