Skip to content

Arthrex® AR-8916VSR-03Recalls & reported problems

Arthrex, Inc. · VOLAR DISTAL RAD PLT TI STD RT 3H · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867131132), received between Sep 26, 2023 and Aug 31, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Defective Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Device1
  • Device Damaged by Another Device1
  • Physical Resistance/Sticking1
  • Positioning Failure1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 31, 2026Device Damaged by Another Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50518471 ↗
  • MalfunctionFeb 12, 2025Positioning Failure; Positioning Failure

    ON 01/22/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-8916VSR-03 VOLAR DISTAL RADIUS PLATE EXPERIENCED AN ISSUE WITH A SCREW NOT LOCKING. THIS OCCURRED DURING A DISTAL RADIUS FRACTURE ON (B)(6) 2025 WHEN THE DISTAL SCREW WENT STRAIGHT THROUGH THE HOLE IN THE PLATE. THE PLATE WAS REMOVED. THE CASE CONTINUED USING THE OTHER PLATE IN THE TRAY WHICH WORKED ACCORDINGLY. THERE WAS A SLIGHT DELAY, WHERE THE REPRESENTATIVE DID NOT BELIEVE ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: Insufficient Information

    MAUDE report 21364218 ↗
  • MalfunctionMay 7, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 03/10/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT WHEN SURGEON WENT TO INSERT THE VERY LAST AR-8724VCL-18 KREULOCK SCREW, IT SUNK BELOW THE DISTAL HOLE OF AN AR-8916VSR-03 VOLAR DISTAL RADIUS PLATE. THE SCREW BECAME STUCK IN THE PLATE AND WHEN SURGEON ATTEMPTED TO REMOVED THE SCREW, THE T8 SCREW DRIVER BENT. FINALLY, SURGEON HAS TO REMOVE ALL THE OTHER SCREWS AND FORCIBLY REMOVE THE LAST SCREW WHICH WAS STILL STUCK TO THE PLATE. A NEW PLATE WAS OPENED AND CASE WAS COMPLETED. THIS WAS DISCOVERED DURING A RADIUS FRACTURE PROCEDURE ON (B)(6) 2022.

    Patient problems: Insufficient Information

    MAUDE report 19257533 ↗
  • MalfunctionSep 26, 2023Defective Device; Defective Device

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8916VSR-03 VOLAR DISTAL RADIUS PLATE APPEARED NOT TO HAVE A STOPPING POINT, AND WHEN THE SCREW WAS INSERTER, IT WENT THROUGH THE PLATE. THE CASE WAS COMPLETED USING A NEW PLATE. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2023. ADDITIONAL INFORMATION RECEIVED ON 09/01/2023: THIS WAS DISCOVERED DURING A DISTAL RADIUS FRACTURE PROCEDURE. THE SURGEON REMOVED THE AR-8724V-18 LOW PROFILE VA LOCKING SCREW FROM THE PLATE AND TRIED TO INSERT A NEW ONE, WHICH HAD THE SAME RESULTS. THE SURGEON REMOVED AR-8724V-18 LOW PROFILE VA LOCKING SCREW THAT WENT THROUGH THE PLATED AND RE-IMPLANTED USING THE NEW AR-8916VSR-03 VOLAR DISTAL RADIUS PLATE TO COMPLETE THE CASE SUCCESSFULLY.

    Patient problems: Insufficient Information

    MAUDE report 17816848 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page