Arthrex® AR-8916VSR-03Recalls & reported problems
Arthrex, Inc. · VOLAR DISTAL RAD PLT TI STD RT 3H · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867131132), received between Sep 26, 2023 and Aug 31, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Defective Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Device1
- Device Damaged by Another Device1
- Physical Resistance/Sticking1
- Positioning Failure1
Patient problems
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2026Device Damaged by Another Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50518471 ↗MalfunctionFeb 12, 2025Positioning Failure; Positioning Failure
ON 01/22/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-8916VSR-03 VOLAR DISTAL RADIUS PLATE EXPERIENCED AN ISSUE WITH A SCREW NOT LOCKING. THIS OCCURRED DURING A DISTAL RADIUS FRACTURE ON (B)(6) 2025 WHEN THE DISTAL SCREW WENT STRAIGHT THROUGH THE HOLE IN THE PLATE. THE PLATE WAS REMOVED. THE CASE CONTINUED USING THE OTHER PLATE IN THE TRAY WHICH WORKED ACCORDINGLY. THERE WAS A SLIGHT DELAY, WHERE THE REPRESENTATIVE DID NOT BELIEVE ADDITIONAL ANESTHESIA WAS ADMINISTERED.
Patient problems: Insufficient Information
MAUDE report 21364218 ↗MalfunctionMay 7, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
ON 03/10/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT WHEN SURGEON WENT TO INSERT THE VERY LAST AR-8724VCL-18 KREULOCK SCREW, IT SUNK BELOW THE DISTAL HOLE OF AN AR-8916VSR-03 VOLAR DISTAL RADIUS PLATE. THE SCREW BECAME STUCK IN THE PLATE AND WHEN SURGEON ATTEMPTED TO REMOVED THE SCREW, THE T8 SCREW DRIVER BENT. FINALLY, SURGEON HAS TO REMOVE ALL THE OTHER SCREWS AND FORCIBLY REMOVE THE LAST SCREW WHICH WAS STILL STUCK TO THE PLATE. A NEW PLATE WAS OPENED AND CASE WAS COMPLETED. THIS WAS DISCOVERED DURING A RADIUS FRACTURE PROCEDURE ON (B)(6) 2022.
Patient problems: Insufficient Information
MAUDE report 19257533 ↗MalfunctionSep 26, 2023Defective Device; Defective Device
ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8916VSR-03 VOLAR DISTAL RADIUS PLATE APPEARED NOT TO HAVE A STOPPING POINT, AND WHEN THE SCREW WAS INSERTER, IT WENT THROUGH THE PLATE. THE CASE WAS COMPLETED USING A NEW PLATE. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2023. ADDITIONAL INFORMATION RECEIVED ON 09/01/2023: THIS WAS DISCOVERED DURING A DISTAL RADIUS FRACTURE PROCEDURE. THE SURGEON REMOVED THE AR-8724V-18 LOW PROFILE VA LOCKING SCREW FROM THE PLATE AND TRIED TO INSERT A NEW ONE, WHICH HAD THE SAME RESULTS. THE SURGEON REMOVED AR-8724V-18 LOW PROFILE VA LOCKING SCREW THAT WENT THROUGH THE PLATED AND RE-IMPLANTED USING THE NEW AR-8916VSR-03 VOLAR DISTAL RADIUS PLATE TO COMPLETE THE CASE SUCCESSFULLY.
Patient problems: Insufficient Information
MAUDE report 17816848 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page