Arthrex® AR-8916VSR-05Recalls & reported problems

Arthrex, Inc. · VOLAR DISTAL RAD PLT TI STD RT 5H · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888867131156), received between Jun 9, 2020 and Sep 28, 2026 — classified as 4 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Positioning Failure2
  • Defective Device1
  • Device Damaged by Another Device1
  • Inadequacy of Device Shape and/or Size1
  • Use of Device Problem1

Patient problems

  • Failure of Implant4
  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2
  • Bone Fracture(s)1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Device Damaged by Another Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914725 ↗
  • MalfunctionApr 21, 2025Positioning Failure; Defective Device; Positioning Failure; Defective Device

    ON 3/25/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A VAL SCREW ADVANCE STRAIGHT THROUGH AN AR-8916VSR-05 VOLAR DISTAL RADIUS PLATE. THE SCREW WAS REMOVED, AND ANOTHER 2.4 VAL SCREW WAS INSERTED, BUT THE SAME THING HAPPENED. THIS WAS DISCOVERED DURING A DISTAL RADIUS FRACTURE PROCEDURE ON (B)(6) 2025.

    Patient problems: Insufficient Information

    MAUDE report 21875101 ↗
  • MalfunctionApr 17, 2025Positioning Failure; Positioning Failure

    ADDITIONAL INFORMATION WAS RECEIVED ON 4/15/2025: NOTHING BROKE INSIDE THE PATIENT. THE CASE WAS SUCCESSFULLY COMPLETED WITH A NEW PLATE AND SCREWS. HOWEVER, THE PROCEDURE WAS DELAYED BY APPROXIMATELY FORTY-FIVE MINUTES, REQUIRING THE ADMINISTRATION OF ADDITIONAL ANESTHESIA. ADDITIONAL INFORMATION WAS REQUESTED ON 5/6/2025.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21855556 ↗
  • InjuryOct 6, 2023Break; Break

    IT WAS REPORTED THAT A DISTAL RADIUS OSTEOTOMY REVISION SURGERY WAS PERFORMED ON (B)(6)2023, TO REPLACE A BROKEN-OFF VOLAR DISTAL RADIUS PLATE. A FEW WEEKS AFTER THE REVISION SURGERY THE PLATE BROKE AGAIN AND HAD TO BE REPLACED WITH THE SAME TYPE OF IMPLANT IN AN ADDITIONAL SURGERY ON (B)(6) 2023.

    Patient problems: Failure of Implant

    MAUDE report 17884648 ↗
  • InjuryOct 3, 2023Break; Break

    IT WAS REPORTED THAT A DISTAL RADIUS OSTEOTOMY REVISION SURGERY WAS PERFORMED ON (B)(6) 2023, TO REPLACE A BROKEN-OFF VOLAR DISTAL RADIUS PLATE. A FEW WEEKS AFTER THE REVISION SURGERY THE PLATE BROKE AGAIN A SECOND TIME AND HAD TO BE REPLACED AGAIN. A PLATE FROM A COMPETITOR WAS USED TO FINISH THE SURGERY SUCCESSFULLY AND THIS SURGERY WAS PERFORMED ON (B)(6) 2023. THE BROKEN IMPLANT WAS COMPLETELY RETRIEVED FROM PATIENT.

    Patient problems: Failure of Implant

    MAUDE report 17862862 ↗
  • InjuryOct 3, 2023Break; Break

    IT WAS REPORTED THAT THE VOLAR DISTAL RADIUS PLATE BROKE A FEW WEEKS POST-OP. NO FURTHER INFORMATION RECEIVED. UPDATE 07-SEP-2023 NROE: FURTHER INFORMATION REVEALED THAT DURING A DISTAL RADIUS OSTEOTOMY A VOLAR DISTAL RADIUS PLATE WAS IMPLANTED INTO THE PATIENT. A FEW WEEKS AFTER THE INITIAL SURGERY THE PLATE BROKE AND HAD TO BE REPLACED WITH THE SAME TYPE OF IMPLANT IN AN ADDITIONAL SURGERY ON (B)(6) 2023.

    Patient problems: Failure of Implant

    MAUDE report 17862702 ↗
  • InjuryJun 9, 2020Inadequacy of Device Shape and/or Size; Use of Device Problem

    AN ARTHREX SALES REP WAS NOT PRESENT DURING THE PROCEDURE. IT WAS REPORTED TO THE SALES REP THAT THE SURGEON WAS USING AN ARTHREX 5 HOLE STANDARD VOLAR PLATE (REP STATES THE PLATE WAS BELIEVED TO BE A RIGHT PLATE BASED ON THE X-RAY), AR-8916VSR-05, FROM A DRP SET. THE SET WAS A LOANER SET THAT HAD BEEN BROUGHT IN FOR A PREVIOUS CASE AND DUE TO COVID-19 IT WAS UNABLE TO BE RETRIEVED. THE SET WAS CALLED FOR AGAIN FOR THIS CASE. HE USED THE 1.7MM DRILL, AR-8916-14, THE DROP IN VAL GUIDE, AR-8916-21 (NOT AN EXTREME ANGLE), AND THEN PUT IN AN ARTRHERX 2.4MM GOLD VAL LOCKING SCREW, AR-8724V-XX (LENGTH UNKNOWN), INTO ONE OF THE DISTAL HOLES OF THE PLATE. SURGEON REPORTED TO THE SALES REP THAT THE PROCEDURE WAS ALL BY THE BOOK. THE LOCKING SCREW WENT INTO THE HOLE AND HAD JUST A TINY BIT OF BITE AND THEN IT JUST SPUN IN THE PLATE HOLE. IT IS REPORTED TO THE SALES REP THAT OVER TENSIONING DID NOT OCCUR. SURGEON TRIED A LOCKING SCREW, 2.4MM VAL LOCKING SCREW, AR-8724V-XX (SIZE UNKNOWN), IN ANOTHER PLATE TO TEST FEEL WHAT IT SHOULD BE LIKE INTRA OP AND IT HAD GOOD BITE. AFTER THE FIRST LOCKING SCREW SPUN IN THE HOLE, THE SURGEON TRIED ANOTHER SCREW IN THE SAME HOLE WITH THE SAME RESULT. SURGE

    Patient problems: Bone Fracture(s); Failure of Implant

    MAUDE report 10133205 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →