Arthrex® AR-8916VSR-05Recalls & reported problems
Arthrex, Inc. · VOLAR DISTAL RAD PLT TI STD RT 5H · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888867131156), received between Jun 9, 2020 and Sep 28, 2026 — classified as 4 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Positioning Failure2
- Defective Device1
- Device Damaged by Another Device1
- Inadequacy of Device Shape and/or Size1
- Use of Device Problem1
Patient problems
- Failure of Implant4
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- Bone Fracture(s)1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Device Damaged by Another Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50914725 ↗MalfunctionApr 21, 2025Positioning Failure; Defective Device; Positioning Failure; Defective Device
ON 3/25/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A VAL SCREW ADVANCE STRAIGHT THROUGH AN AR-8916VSR-05 VOLAR DISTAL RADIUS PLATE. THE SCREW WAS REMOVED, AND ANOTHER 2.4 VAL SCREW WAS INSERTED, BUT THE SAME THING HAPPENED. THIS WAS DISCOVERED DURING A DISTAL RADIUS FRACTURE PROCEDURE ON (B)(6) 2025.
Patient problems: Insufficient Information
MAUDE report 21875101 ↗MalfunctionApr 17, 2025Positioning Failure; Positioning Failure
ADDITIONAL INFORMATION WAS RECEIVED ON 4/15/2025: NOTHING BROKE INSIDE THE PATIENT. THE CASE WAS SUCCESSFULLY COMPLETED WITH A NEW PLATE AND SCREWS. HOWEVER, THE PROCEDURE WAS DELAYED BY APPROXIMATELY FORTY-FIVE MINUTES, REQUIRING THE ADMINISTRATION OF ADDITIONAL ANESTHESIA. ADDITIONAL INFORMATION WAS REQUESTED ON 5/6/2025.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21855556 ↗InjuryOct 6, 2023Break; Break
IT WAS REPORTED THAT A DISTAL RADIUS OSTEOTOMY REVISION SURGERY WAS PERFORMED ON (B)(6)2023, TO REPLACE A BROKEN-OFF VOLAR DISTAL RADIUS PLATE. A FEW WEEKS AFTER THE REVISION SURGERY THE PLATE BROKE AGAIN AND HAD TO BE REPLACED WITH THE SAME TYPE OF IMPLANT IN AN ADDITIONAL SURGERY ON (B)(6) 2023.
Patient problems: Failure of Implant
MAUDE report 17884648 ↗InjuryOct 3, 2023Break; Break
IT WAS REPORTED THAT A DISTAL RADIUS OSTEOTOMY REVISION SURGERY WAS PERFORMED ON (B)(6) 2023, TO REPLACE A BROKEN-OFF VOLAR DISTAL RADIUS PLATE. A FEW WEEKS AFTER THE REVISION SURGERY THE PLATE BROKE AGAIN A SECOND TIME AND HAD TO BE REPLACED AGAIN. A PLATE FROM A COMPETITOR WAS USED TO FINISH THE SURGERY SUCCESSFULLY AND THIS SURGERY WAS PERFORMED ON (B)(6) 2023. THE BROKEN IMPLANT WAS COMPLETELY RETRIEVED FROM PATIENT.
Patient problems: Failure of Implant
MAUDE report 17862862 ↗InjuryOct 3, 2023Break; Break
IT WAS REPORTED THAT THE VOLAR DISTAL RADIUS PLATE BROKE A FEW WEEKS POST-OP. NO FURTHER INFORMATION RECEIVED. UPDATE 07-SEP-2023 NROE: FURTHER INFORMATION REVEALED THAT DURING A DISTAL RADIUS OSTEOTOMY A VOLAR DISTAL RADIUS PLATE WAS IMPLANTED INTO THE PATIENT. A FEW WEEKS AFTER THE INITIAL SURGERY THE PLATE BROKE AND HAD TO BE REPLACED WITH THE SAME TYPE OF IMPLANT IN AN ADDITIONAL SURGERY ON (B)(6) 2023.
Patient problems: Failure of Implant
MAUDE report 17862702 ↗InjuryJun 9, 2020Inadequacy of Device Shape and/or Size; Use of Device Problem
AN ARTHREX SALES REP WAS NOT PRESENT DURING THE PROCEDURE. IT WAS REPORTED TO THE SALES REP THAT THE SURGEON WAS USING AN ARTHREX 5 HOLE STANDARD VOLAR PLATE (REP STATES THE PLATE WAS BELIEVED TO BE A RIGHT PLATE BASED ON THE X-RAY), AR-8916VSR-05, FROM A DRP SET. THE SET WAS A LOANER SET THAT HAD BEEN BROUGHT IN FOR A PREVIOUS CASE AND DUE TO COVID-19 IT WAS UNABLE TO BE RETRIEVED. THE SET WAS CALLED FOR AGAIN FOR THIS CASE. HE USED THE 1.7MM DRILL, AR-8916-14, THE DROP IN VAL GUIDE, AR-8916-21 (NOT AN EXTREME ANGLE), AND THEN PUT IN AN ARTRHERX 2.4MM GOLD VAL LOCKING SCREW, AR-8724V-XX (LENGTH UNKNOWN), INTO ONE OF THE DISTAL HOLES OF THE PLATE. SURGEON REPORTED TO THE SALES REP THAT THE PROCEDURE WAS ALL BY THE BOOK. THE LOCKING SCREW WENT INTO THE HOLE AND HAD JUST A TINY BIT OF BITE AND THEN IT JUST SPUN IN THE PLATE HOLE. IT IS REPORTED TO THE SALES REP THAT OVER TENSIONING DID NOT OCCUR. SURGEON TRIED A LOCKING SCREW, 2.4MM VAL LOCKING SCREW, AR-8724V-XX (SIZE UNKNOWN), IN ANOTHER PLATE TO TEST FEEL WHAT IT SHOULD BE LIKE INTRA OP AND IT HAD GOOD BITE. AFTER THE FIRST LOCKING SCREW SPUN IN THE HOLE, THE SURGEON TRIED ANOTHER SCREW IN THE SAME HOLE WITH THE SAME RESULT. SURGE
Patient problems: Bone Fracture(s); Failure of Implant
MAUDE report 10133205 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page