Skip to content

Arthrex® AR-8970TTCRecalls & reported problems

Arthrex, Inc. · ANKLE FUSION PLATE, LATERAL TTC · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867134010), received between Jul 22, 2020 and Jul 22, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Failure of Implant1
  • Pain1
  • Swelling1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 22, 2020Break

    IT WAS REPORTED THE PATIENT HAD AN ANKLE FUSION ON (B)(6) 2019 AND THE PATIENT EXPERIENCED PAIN AND SWELLING HAD A FOLLOW UP X-RAY AT THE END OF (B)(6) THAT REVEALED THE AR-8970TTC, ANKLE FUSION PLATE, WAS BROKEN. LOT NUMBER OF THE PLATE WAS UNKNOWN AT THE TIME OF INITIAL REPORT. THE PATIENT HAD TO WAIT UNTIL JUNE TO HAVE A REVISION SURGERY TO REMOVE THE BROKEN PLATE AND SCREWS. THE DOCTOR USED A REVISION ANKLE SUBTALAR FUSION TO REMEDY THE SITUATION. THE PATIENT IS CURRENTLY IN A CAST AND IN A HEALING STATE. ADDITIONAL INFORMATION OBTAINED 7/20/2020: THE ORIGINAL PROCEDURE WAS A TTC FUSION ON (B)(6) 2019. THE REVISION SURGERY TOOK PLACE ON (B)(6) 2020. THE REVISION WAS COMPLETED USING A NAIL. THE SURGEON HAS THE PLATE AND SCREWS AND WILL RETURN THEM TO ARTHREX FOR EVALUATION. THE PATIENT SUFFERED NO KNOWN TRAUMA SINCE THE ORIGINAL PROCEDURE. BOTH ORIGINAL AND REVISION PROCEDURES WERE PERFORMED BY THE SAME SURGEON AT THE SAME FACILITY. SCREW PART NUMBERS ARE NOT KNOWN AT THIS TIME. ADDITIONAL INFORMATION OBTAINED 7/21/2020: THE FOLLOWING ARE THE DEVICES ORIGINALLY IMPLANTED (B)(6) 2019. THE SURGEON CHOSE TO REMOVE ALL OF THE BELOW DEVICES DURING THE REVISION SURGERY: AR-8970TTC, AN

    Patient problems: Failure of Implant; Pain; Swelling

    MAUDE report 10309706 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page