Skip to content

Arthrex® AR-2923BCHRecalls & reported problems

Arthrex, Inc. · SUTR ANCH HIP BIO-COMP P-LCK 2.9X12.5MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867195844), received between Nov 4, 2020 and Jul 22, 2025 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 22, 2025Break; Break

    ON 07/01/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT (QTY. 3) OF AN AR-2923BCH SUTURE ANCHOR BIOCOMPOSITE HIP PUSHLOCK EYELETS CONSECUTIVELY BROKE OFF DURING IMPLANTATION. THE ISSUE OCCURRED DESPITE THE USE OF PROPER SURGICAL TECHNIQUE. THE THREADS ON THE AFFECTED ANCHORS WERE DAMAGED, PREVENTING REASSEMBLY. THE CASE WAS COMPLETED, AND NO REPORTED PIECES BROKE OFF INSIDE THE PATIENT. THIS WAS DISCOVERED DURING A HIP LABRAL REPAIR PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON 07/10/2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22569534 ↗
  • MalfunctionJul 22, 2025Break; Break

    ON 07/01/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT (QTY. 3) OF AN AR-2923BCH SUTURE ANCHOR BIOCOMPOSITE HIP PUSHLOCK EYELETS CONSECUTIVELY BROKE OFF DURING IMPLANTATION. THE ISSUE OCCURRED DESPITE THE USE OF PROPER SURGICAL TECHNIQUE. THE THREADS ON THE AFFECTED ANCHORS WERE DAMAGED, PREVENTING REASSEMBLY. THE CASE WAS COMPLETED, AND NO REPORTED PIECES BROKE OFF INSIDE THE PATIENT. THIS WAS DISCOVERED DURING A HIP LABRAL REPAIR PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON 07/10/2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22569532 ↗
  • MalfunctionJul 22, 2025Break; Break

    ON 07/01/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT (QTY. 3) OF AN AR-2923BCH SUTURE ANCHOR BIOCOMPOSITE HIP PUSHLOCK EYELETS CONSECUTIVELY BROKE OFF DURING IMPLANTATION. THE ISSUE OCCURRED DESPITE THE USE OF PROPER SURGICAL TECHNIQUE. THE THREADS ON THE AFFECTED ANCHORS WERE DAMAGED, PREVENTING REASSEMBLY. THE CASE WAS COMPLETED, AND NO REPORTED PIECES BROKE OFF INSIDE THE PATIENT. THIS WAS DISCOVERED DURING A HIP LABRAL REPAIR PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON 07/10/2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22569528 ↗
  • MalfunctionFeb 5, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON 1/17/2025, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT QTY. 2 OF AN AR-1938BCH SUTURE ANCHOR, BIOCOMPOSITE, KNOTLESS HIP SUTURETAK, AND AN AR-2923BCH SUTURE ANCHOR BIOCOMPOSITE HIP PUSHLOCK HAD AN ISSUE DURING A HIP ARTHROSCOPIC LABRAL REPAIR PROCEDURE ON (B)(6) 2025. WHEN USING THE FIRST AR-1938BCH SUTURE ANCHOR, UPON SHUTTLING THE FIRST REPAIR STITCH THROUGH, THE LEAD STITCH BROKE. ALL STITCHES WERE CUT OUT AND REMOVED, BUT THE ANCHOR REMAINED INSIDE THE PATIENT. WHEN USING THE SECOND AR-1938BCH SUTURE ANCHOR, UPON SHUTTLING THE LEAD STITCH, THE TOP ONE-THIRD OF THE ANCHOR BROKE OFF INSIDE THE PATIENT. THE SURGEON REMOVED ALL STITCHES AND THE BROKEN TOP OF THE ANCHOR, BUT THE REMAINING TWO-THIRDS OF THE ANCHOR STAYED INSIDE THE PATIENT. WHEN USING THE AR-2923BCH SUTURE ANCHOR BIOCOMPOSITE HIP PUSHLOCK, THAT ANCHOR PULLED OUT AFTER INSERTION IN ONE PIECE. EVERYTHING WAS REMOVED FROM THE PATIENT. THE THIRD ANCHOR WAS DISCARDED, AND NO PICTURES WERE TAKEN. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, USING A KNOTLESS FIBERTAK ANCHOR AND A PUSHLOCK ANCHOR WITH UNKNOWN PRODUCT PART NUMBERS. THE CASE WAS DELAYED FOR ABOUT 10 MINUTES, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. TH

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21315180 ↗
  • MalfunctionJan 16, 2023Break

    IT WAS REPORTED ON 12/29/2022 BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-2923BCH SUTURE ANCHOR EYELET BROKE AND WOULD NOT CONTAIN THE SUTURES. CASE INVOLVEMENT, DEVICE BROKE DURING USE. PER FACILITY: "DURING A LABRAL REPAIR OF THE HIP THE LABRAL TAPES WERE PASSED AROUND THE LABRUM AND THE PILOT HOLE WAS DRILLED WITH A 2923DHS. THE SUTURE WAS PUT THROUGH THE EYELET OF THE 2.9 PUSHLOCK AND INTRODUCED INTO THE HIP. THE EYELET WAS PUT INTO THE PILOT HOLE AND THE SLACK WAS PULLED OUT OF THE SUTURE AND THE EYELET POPPED OFF THE ANCHOR. WHEN THE EYELET AND ANCHOR WERE REMOVED FROM THE HIP WE NOTICED THERE WAS A SLIT IN THE PLASTIC EYELET AND THE SUTURES WOULDN'T STAY IN ANY LONGER. "

    Patient problems: Insufficient Information

    MAUDE report 16173458 ↗
  • InjuryNov 4, 2020Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A AFTER AN INITIAL HIP LABRUM STABILIZATION ON (B)(6) 2020 IT WAS FOUND DURING THE AFTERCARE THAT THE PATIENT HAS DEVELOPED LYSIS. THE SURGEON HAS FURTHER REPORTED THAT THE REPORTED EVENT HAS BEEN SEEN WITH OTHER PATIENTS AS WELL BUT WAS NOT ABLE TO PROVIDE ANY FURTHER INFORMATION REGARDING THESE EVENTS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 10786048 ↗

Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page