Arthrex® AR-3200-0021Recalls & reported problems
Arthrex, Inc. · CCU, ARTHREX SYNERGY UHD4 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Arthrex Synergy UHD4 camera control unit is the console that powers the light and camera used in arthroscopy and general laparoscopic surgery. Most of the 58 reports describe the unit running too hot, cutting out mid-procedure, giving off a burning smell, or blackening the light cord where it plugs in; a few mention smoke or a spark. Eight injuries were reported, including burns and an electric shock, and no deaths. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 57
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 57 adverse event reports that list this exact device (DI 00888867197022), received between Feb 27, 2019 and Apr 8, 2026 — classified as 8 injuries, 49 malfunctions.
The most frequently reported specific device problem is “Display or Visual Feedback Problem” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Display or Visual Feedback Problem12
- Overheating of Device11
- Defective Device7
- Device Emits Odor4
- Intermittent Loss of Power3
- Break2
- Computer Software Problem2
- Excessive Heating2
Patient problems
- No Clinical Signs, Symptoms or Conditions45
- Insufficient Information4
- Burn(s)2
- Complaint, Ill-Defined2
- Electric Shock2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Burn, Thermal1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 57 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2026Defective Device
IT WAS REPORTED THAT DURING AN ARTHROSCOPY SURGERY THE LIGHT SOURCE OF THE DEVICE FAILED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24818279 ↗MalfunctionFeb 6, 2025Overheating of Device; Overheating of Device
ON (B)(6) 2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-3200-0021 SYNERGYUHD4 CAMERA CONTROL UNIT HAD A BURNING SMELL AND WAS PHYSICALLY DAMAGED WHERE THE LIGHT CORD WAS INSERTED INTO THE UNIT. THE END OF THE LIGHT CORD SEEMED BURNED AND WAS BLACK AT THE END, AS WELL AS THE CONSOLE'S PLUG-IN. THE CASE WAS COMPLETED USING ANOTHER LIGHT CORD, AND AT THE END OF THE CASE, THE SAME ISSUE OCCURRED. IT IS UNKNOWN IF THE TWO LIGHT CORDS USED IN THE CASE WERE ARTHREX PRODUCTS. THERE WAS NO INJURY TO THE PATIENT, SURGEON, OR FACILITY STAFF. THE CASE WAS DELAYED FOR A COUPLE OF MINUTES, AND THERE WAS NO INFORMATION ON ADDED ANESTHESIA ADMINISTERED TO THE PATIENT. THERE WAS NO REP PRESENT AT THE CASE. THIS WAS DISCOVERED DURING A UROLOGY PROCEDURE ON (B)(6) 2025, WITH NO REPORT OF PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 01/13/2025: THE TWO UNKNOWN LIGHT CORDS REPORTED DURING THIS PROCEDURE ARE NON-ARTHREX PRODUCTS AND WERE DISCARDED. NO ADDITIONAL ANESTHESIA WAS ADMINISTERED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21320896 ↗MalfunctionJun 19, 2024Sparking; Sparking
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06577322 THAT AN AR-3200-0021 CCU CONSOLE HAD A SPARK COME FROM THE UNIT. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19567679 ↗MalfunctionMay 17, 2024Overheating of Device; Overheating of Device
ON 4/19/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4)THAT AN AR-3200-0021 SYNERGYUHD4 CAMERA CONTROL UNIT WAS OVERHEATING AND GOT VERY HOT AND BURNED THROUGH THE LIGHT GUIDES AND A STAFF MEMBER'S CLOTHING. IT WAS STATED THAT THE STAFF MEMBER WAS NOT HURT, AND THE LIGHT GUIDES WERE A NON-ARTHREX PRODUCT. THIS WAS DISCOVERED DURING A PROCEDURE, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 4/29/2024: THIS WAS DISCOVERED DURING A URETEROSCOPY PROCEDURE ON (B)(6) 2024. IT WAS STATED THAT THE EVENT WAS A "NEAR MISS" EVENT. THE OR TECH'S GOWN HAD TWO BURN HOLES CAUSED BY THE END OF THE LIGHT CABLE AFTER IT WAS REMOVED FROM THE CCU. NOTHING ELSE WAS PLUGGED INTO THE CCU AT THE TIME OF THE EVENT. THE END OF THE LIGHT SOURCE WAS TOO HOT TO TOUCH AND LOOKED SCORCHED. ONLY A BURNING SMELL CAME FROM THE OR TECH'S GOWN. THE CASE WAS COMPLETED AS PLANNED BEFORE IDENTIFYING THE LIGHT CABLE ISSUE (THE END OF THE LIGHT SOURCE HAD SCORCH MARKS THAT COULD NOT BE REMOVED DURING PROCESSING IN SPD; THEREFORE, THERE ARE CONCERNS ABOUT OVERHEATING AND THE CCU). THERE WAS NO DELAY IN THE CASE OR ADVERSE EFFECTS DUE TO THIS BEING A "NEAR MISS" EVENT. NO PHOTOS WERE TAKEN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19343403 ↗MalfunctionApr 30, 2024Intermittent Loss of Power; Intermittent Loss of Power
ON 04/03/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3200-0021 CONSOLE LIGHT CORD WOULD GO BLACK FOR A FEW SECONDS A FEW TIMES. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM. TECH SUPPORT RESOLVED ISSUE BY NAVIGATING TO LIGHT GUIDE AUTO SELECTION AND SWITCH IT TO OFF TO KEEP LIGHT GUIDE FROM TURNING ON AND OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19214918 ↗MalfunctionApr 23, 2024Intermittent Loss of Power; Intermittent Loss of Power
IT WAS REPORTED THAT THE LIGHT SOURCE OF THE DEVICE IS SPORADICALLY WITHOUT FUNCTION. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. UPDATE AHA 08-APR-2024: THE FAILURE OCCURRED DURING A SURGERY BUT THERE WAS NO PATIENT HARM. THE LED ALSO CAME ON AGAIN, THERE WERE BRIEF INTERRUPTIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19166424 ↗MalfunctionApr 18, 2024Intermittent Loss of Power; Intermittent Loss of Power
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS (B)(4) THAT AN AR-3200-0021 ARTHREX SYNERGY UHD4 IS CAUSING LIGHT CORDS TO TURN OFF INTERMITTENTLY. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCEDURE, WITH NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19140206 ↗MalfunctionApr 17, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE HEAT FROM LIGHT SOURCE IS TOO MUCH, ITS BURNS THE LIGHT CABLE. THE PROBLEM WAS NOTICED AFTER THE SURGERY. THERE WAS NO HARM FOR THE PATIENT, OPERATOR OR THIRD PARTY REPORTED. ALSO THERE WAS NO CASE INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19123703 ↗MalfunctionMar 28, 2024Device Emits Odor; Device Emits Odor
ON (B)(6) 2024, IT WAS REPORTED BY A DISTRIBUTOR VIA SEMS (B)(4) THAT THE CONNECTION END THAT GOES INTO AN AR-3200-0021 CCU OF A LIGHT GUIDE HAD A BURNING SMELL. WHEN IT WAS REMOVED, IT WAS SMOKING AND HAD VISIBLE BURNS. THIS WAS DISCOVERED DURING SET UP, BEFORE THE PROCEDURE AND HAD NO EFFECT ON THE PATIENT. THE LIGHT CORD PRODUCT WAS FROM ANOTHER MANUFACTURER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18994955 ↗MalfunctionMar 22, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT THE LIGHT CABLE ADAPTER IS STUCK. IT CAN NOT BE TURNED AROUND AS USUAL. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. ALSO THERE WAS NO CASE INVOLVEMENT REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18961693 ↗InjuryMar 21, 2024Unintended Electrical Shock; Unintended Electrical Shock
IT WAS REPORTED THAT THE CAMERA SHOCKED THE SURGEON WHEN HE TRIED TO FOCUS. ADDITIONAL INFORMATION OBTAINED 11/16/20: THE SALES REP REPORTED THAT THE SHOCK WAS LIKE A JOLT WHEN THE SURGEON WEN TO FOCUS. THE JOLT WAS FELT ON THE PALM OF HIS HAND. THE SURGEON FINISHED UP WITH THE CASE. THE SURGEON RECEIVED NO BURNS AND THE JOLT OCCURRED THROUGH HIS SURGICAL GLOVE.
Patient problems: Insufficient Information
MAUDE report 18956124 ↗MalfunctionJan 30, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE FIBER OPTIC CABLE BECOMES VERY HOT DURING OPTICAL CONNECTION. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THERE WAS NO CASE INVOLVEMENT REPORTED. NO FURTHER INFORMATION RECEIVED. UPDATE DW 22-DEC-2023: THE REPORTED ISSUE WAS NOTICED DURING A SURGERY. UPDATE SWIT 02-JAN-2024: A SHOULDER ARTHROSCOPY WAS PERFORMED. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE CABLE DID NOT COME IN CONTACT WITH THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME LIGHT GUIDE CABLE. UPDATE DW 10-JAN-2024: FURTHER INFORMATION WERE PROVIDED THAT THE EVENT IS NOT RELATED TO THE USED CABLES AND NOT THE CCU WITH THE PART NUMBER AR-3240-5027. UPDATE DW 29-JAN-2024: THE NUMBER OF AFFECTED DEVICES AND LOT NUMBER OF THE DEFECTIVE CABLES WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18607925 ↗MalfunctionJan 5, 2024Device Emits Odor; Device Emits Odor
ON 12/8/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3200-0021 CCU HAD AN ISSUE. THE FAN STOPPED WORKING, AND A BURNING SMELL CAME OUT. THIS WAS DISCOVERED AFTER USE. NO CASE INVOLVEMENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ON 1/2/2024, THE SALES REPRESENTATIVE PROVIDED ADDITIONAL INFORMATION VIA EMAIL: THIS WAS NOT REPORTED DURING A CASE. THEY NOTICED A BURNT SMELL COMING FROM THE CCU AND, AS A RESULT, CLOSED THE OR DOWN FOR THAT DAY AND DID NOT DO ANY CASES IN THAT OR UNTIL THE CCU WAS REPLACED THE NEXT DAY. NO SMOKE WAS NOTICED, AND NO ADVERSE EFFECTS TO THE STAFF OR SURGEON WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 18453601 ↗MalfunctionDec 28, 2023Defective Device; Defective Device
ON 11/30/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3200-0021 SYNERGY CONSOLE HAD A FEW OF THE CONNECTOR TEETH DAMAGED MAKING IT HARD TO PLUG IN CAMERAS. THIS WAS DISCOVERED AFTER THE CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18415283 ↗MalfunctionDec 8, 2023Computer Software Problem; Computer Software Problem
ON 11/14/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3200-0021 CCU HAD ISSUES AFTER APPLYING THE 4.17.5 UPDATE. THERE WAS GREEN STATIC WHILE USING MULTIPLE CAMERA HEADS (THESE CAMERA HEADS WERE THEN TESTED ON A DIFFERENT CONSOLE AND HAD NO ISSUES), CASES ENDING UNEXPECTEDLY, FOLLOWED BY A NEW CASE AUTO-STARTING WITH PATIENT INFORMATION FROM A SEPARATE ROOM, DUPLICATED CASES, AND NO PHOTOS SAVING FROM THE CASES. BEFORE THE UPGRADE, THE ACCOUNT HAD NO ISSUES WITH THEIR CCU. THIS WAS DISCOVERED DURING THEIR FIRST CASE SINCE THE UPDATE ON 11/14/2023, WITH NO REPORTED ADVERSE EVENT OR PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18292567 ↗MalfunctionDec 8, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
ON 11/14/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06401201 THAT AN AR-3200-0021 CCU HAD A DARK SCREEN/IMAGE AND COULD NOT VISUALLY SEE THE PICTURE AS WELL AS PRIOR USES. TO TROUBLESHOOT, TRIED DIFFERENT F/O CABLES AND CHECKED TABLET CONTROLS TO MAKE SURE THE LIGHT SOURCE WAS TURNED UP, AND IT WAS. THE CASE WAS COMPLETED BY SWITCHING OUT TOWERS AND USING A DIFFERENT CCU BOX WITH NO ISSUES WITH THE LIGHT SOURCE. THERE WAS NO REPORTED DELAY IN THE CASE. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCEDURE, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18292367 ↗MalfunctionDec 1, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
ON 11/07/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06393791 THAT AN AR-3200-0021 CCU IS SHOWING LINES ON THE SCREEN. THIS WAS DISCOVERED DURING USE IN A CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18252541 ↗MalfunctionNov 28, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
ON (B)(6)/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06261780 THAT AN AR-3200-0021 CCU WAS PRODUCING LINES ACROSS THE SCREEN. THIS OCCURRED DURING A CASE WHERE ANOTHER DEVICE WAS USED TO COMPLETE THE CASE.
Patient problems: Insufficient Information
MAUDE report 18223273 ↗MalfunctionNov 15, 2023Computer Software Problem; Computer Software Problem
ON 10/20/2023, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA (B)(4) THAT AN AR-3200-0021 CCU WAS EXPERIENCING AN ISSUE WITH (OVERSIZED LOG FILES IN /VAR/LOG DIRECTORY) AFTER BEING UPDATED TO V4.15.0. THIS OCCURRED DURING A CASE WITH NO PATIENT EFFECT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18141846 ↗MalfunctionNov 15, 2023Poor Quality Image; Poor Quality Image
ON 10/20/2023 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3200-0021 CCU HAS A GRAINY IMAGE. THIS WAS DISCOVERED DURING USE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18141811 ↗MalfunctionNov 10, 2023Break; Break
ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3200-0021 CAMERA PORT IS BROKEN. THIS WAS DISCOVERED DURING USE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18111775 ↗MalfunctionNov 10, 2023Defective Device; Defective Device
ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3200-0021 CCU PRODUCED A "SYSTEM ERROR" ALARM. THIS OCCURRED DURING A PROCEDURE, WHEN THE DEVICE WAS RESTARTED, THE MESSAGE WAS GONE. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18111036 ↗MalfunctionNov 6, 2023Break; Break
IT WAS REPORTED THAT THE LIGHT IS BROKEN. THERE WAS NO HARM FOR THE PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18074973 ↗MalfunctionNov 3, 2023Optical Problem; Optical Problem
ON 10/19/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(6) THAT AN AR-3200-0021 CCU WAS PRODUCING LIGHT SOURCES THAT KEEP GOING OUT. THE CASE WAS COMPLETED USING THE SAME CONSOLE WITH NO REPORTED ADVERSE EFFECTS ON THE PATIENT. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18067497 ↗MalfunctionOct 25, 2023Display or Visual Feedback Problem; Image Display Error/Artifact; Display or Visual Feedback Problem; Image Display Error/Artifact
IT WAS REPORTED THAT THE DATA TRANSFER/DATA PLAYBACK IS POOR. THE PICTURE APPEARS TO BE NOISY. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION WAS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18003035 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page