Skip to content

Arthrex® AR-3210-0025Recalls & reported problems

Arthrex, Inc. · 4K SynergyUHD4 C-Mount Camera Head, Auto · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867197053), received between Dec 21, 2021 and Jun 20, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device3
  • Display or Visual Feedback Problem2
  • Melted1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 20, 2024Overheating of Device; Overheating of Device

    ON 10/13/2021 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-3210-0025 4K HD C-MOUNT CAMERA, IS GETTING VERY WARM/HOT. THIS WAS DISCOVERED DURING USE IN A PROCEDURE. THE PATIENT WAS NOT AFFECTED. THE CASE WAS COMPLETED USING A DIFFERENT CAMERA. TS: CHECKED THE LIGHT SOURCE OUTPUT, CHANGED OUT THE LIGHT CORD THINKING IT WAS GOING BAD AND ALSO TESTED THE CAMERA AFTER THE CASE IN WHICH IT STARTED HEATING UP AGAIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19574205 ↗
  • MalfunctionMay 23, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE HAS A PROBLEN WITH OVERHEATING. NO ADDITIONAL INFORMATION REGARDING A SPECIFIC FAILURE MODE WAS PROVIDED. THE PROBLEM WAS NOTICED AFTER THE SURGERY. THERE WAS NO HARM FOR THE PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19380939 ↗
  • MalfunctionJul 14, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT (2) AR-3210-0025 CAMERA HEADS ARE PRODUCING A YELLOW SCREEN AND ARE UNABLE TO WHITE BALANCE. THIS OCCURRED DURING A CASE, WITH NO PATIENT EFFECT REPORTED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED, ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17324927 ↗
  • MalfunctionJul 14, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem

    ON 06/23/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT (2) AR-3210-0025 CAMERA HEADS ARE PRODUCING A YELLOW SCREEN AND ARE UNABLE TO WHITE BALANCE. THIS OCCURRED DURING A CASE, WITH NO PATIENT EFFECT REPORTED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED, ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17324912 ↗
  • MalfunctionDec 21, 2021Melted; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE OVERHEATED. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED. UPDATE 30-NOV-2021: IT WAS CONFIRMED THAT THE ERROR OCCURRED DURING THE SURGERY. THE PATIENT WAS NOT INJURED BUT THE CAMERA HEAD WAS SO WARM THAT THE CAMERA DRAPE MELTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13053750 ↗

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page