Arthrex® AR-3210-0025Recalls & reported problems
Arthrex, Inc. · 4K SynergyUHD4 C-Mount Camera Head, Auto · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867197053), received between Dec 21, 2021 and Jun 20, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device3
- Display or Visual Feedback Problem2
- Melted1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 20, 2024Overheating of Device; Overheating of Device
ON 10/13/2021 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-3210-0025 4K HD C-MOUNT CAMERA, IS GETTING VERY WARM/HOT. THIS WAS DISCOVERED DURING USE IN A PROCEDURE. THE PATIENT WAS NOT AFFECTED. THE CASE WAS COMPLETED USING A DIFFERENT CAMERA. TS: CHECKED THE LIGHT SOURCE OUTPUT, CHANGED OUT THE LIGHT CORD THINKING IT WAS GOING BAD AND ALSO TESTED THE CAMERA AFTER THE CASE IN WHICH IT STARTED HEATING UP AGAIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19574205 ↗MalfunctionMay 23, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE DEVICE HAS A PROBLEN WITH OVERHEATING. NO ADDITIONAL INFORMATION REGARDING A SPECIFIC FAILURE MODE WAS PROVIDED. THE PROBLEM WAS NOTICED AFTER THE SURGERY. THERE WAS NO HARM FOR THE PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19380939 ↗MalfunctionJul 14, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT (2) AR-3210-0025 CAMERA HEADS ARE PRODUCING A YELLOW SCREEN AND ARE UNABLE TO WHITE BALANCE. THIS OCCURRED DURING A CASE, WITH NO PATIENT EFFECT REPORTED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED, ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17324927 ↗MalfunctionJul 14, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
ON 06/23/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT (2) AR-3210-0025 CAMERA HEADS ARE PRODUCING A YELLOW SCREEN AND ARE UNABLE TO WHITE BALANCE. THIS OCCURRED DURING A CASE, WITH NO PATIENT EFFECT REPORTED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED, ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17324912 ↗MalfunctionDec 21, 2021Melted; Overheating of Device
IT WAS REPORTED THAT THE DEVICE OVERHEATED. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED. UPDATE 30-NOV-2021: IT WAS CONFIRMED THAT THE ERROR OCCURRED DURING THE SURGERY. THE PATIENT WAS NOT INJURED BUT THE CAMERA HEAD WAS SO WARM THAT THE CAMERA DRAPE MELTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13053750 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page