Arthrex® AR-8610D-65Recalls & reported problems
Arthrex, Inc. · DRIVER SHAFT, T25 HEXALOBE, CANNULATED · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00888867197299), received between Nov 30, 2021 and Apr 28, 2026 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Break” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break13
- Material Twisted/Bent2
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Insufficient Information1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 28, 2026Break
IT WAS REPORTED THAT DURING THE ANKLE ARTHRODESIS, THE TIP OF THE SCREWDRIVER TWISTED AND BROKE OFF WHILE THE SCREW (6.5 MM HEADLESS COMPRESSION SCREW) WAS BEING INSERTED. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. IN ADDITION, A GUIDE WIRE (NOT AN ARTHREX PRODUCT) BECAME LODGED IN THE SCREWDRIVER. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25020946 ↗MalfunctionApr 28, 2026Break
IT WAS REPORTED THAT DURING THE ANKLE ARTHRODESIS, THE SCREWDRIVER BROKE WHILE THE SCREW (6.5 MM HEADLESS COMPRESSION SCREW) WAS BEING INSERTED. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25020748 ↗MalfunctionDec 23, 2025Break; Break
IT WAS REPORTED THAT DURING A FOOT SURGERY THE TIP OF THE DEVICE BROKE OFF WHILE USING THE DEVICE IN THE PATIENT. ALL FRAGMENTS WERE RETRIEVED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23880060 ↗MalfunctionJul 2, 2025Break; Break
ON 06/06/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8610D-65 DRIVE SHAFT BROKE DURING INSERTION. ISSUE DISCOVERED DURING A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22386285 ↗MalfunctionNov 10, 2023Break; Break
IT WAS REPORTED THAT DURING A TIBIAL TUBERCLE TRANSFER THE SURGEON DRILLED & COUNTERSUNK APPROPRIATELY. THE SURGEON BEGAN INPUTTING THE SCREW WITH A POWER TOOL AND THEN THEN CHANGED FOR THE FINAL TENSIONING TO A SCREWDRIVER. DURING THE TENSIONING THE DEVICE BROKE OFF INSIDE THE SCREW. THE BROKEN PIECES WERE RETRIEVED OUT OF THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18110927 ↗MalfunctionJun 19, 2023Break; Break
ON 05/23/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-8610D-65 DRIVER SHAFT BROKE. THIS OCCURRED DURING A TRIPLE ARTHRODESIS ON (B)(6) 2023 WHILE TRYING TO ADVANCE A 6.5MM COMP PT SCREW BEING USED FOR A SUBTALAR FUSION. THE CASE WAS COMPLETED USING ANOTHER DRIVER.
Patient problems: Insufficient Information
MAUDE report 17152488 ↗MalfunctionSep 6, 2022Break
SCREW DRIVER HANDPIECE BROKE. NO FURTHER INFORMATION AVAILABLE. MORE INFORMATION REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15362506 ↗MalfunctionJan 31, 2022Break
IT WAS REPORTED THAT DURING A SCREW REMOVAL THE DRIVER SHAFT BECAME TWISTED AND SNAPPED OFF DURING THE REMOVAL. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13397855 ↗MalfunctionJan 31, 2022Break
IT WAS REPORTED THAT DURING A FOOT FIXATION SURGERY THE TIP OF THE DRIVER TWISTED AND BROKE OFF DURING THE SCREW INSERTION. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13397424 ↗MalfunctionJan 31, 2022Break; Material Twisted/Bent
IT WAS REPORTED THAT DURING A FOOT FIXATION SURGERY THE TIP OF THE DEVICE GOT TWISTED AND BROKE OFF. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13397411 ↗MalfunctionJan 31, 2022Break; Material Twisted/Bent
IT WAS REPORTED THAT DURING A FOOT FIXATION SURGERY THE TIP OF THE DRIVER TWISTED AND BROKE DURING THE INSERTION OF THE SCREW. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13397407 ↗MalfunctionJan 31, 2022Break
IT WAS REPORTED THAT DURING AN ANKLE FUSION SURGERY THE TIPS OF TWO DRIVER BROKE OFF WHILE THE SURGEON TRIED TO INSERT THE SCREWS. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13397321 ↗MalfunctionNov 30, 2021Break
IT WAS REPORTED THAT DURING A CALCANEAL SURGERY THE SURGEON DIDN`T USE POWER TOOLS WITH THE DEVICE BUT HAND TIGHTENING AND THE PICTURES SHOW, THAT THE TIP OF THE DEVICE BROKE OFF. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12897038 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page