Arthrex® AR-2922BCRecalls & reported problems
Arthrex, Inc. · SUTR ANCH, BIO-COMP P-LCK, 2.4 X 11.3 · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867197664), received between Sep 6, 2017 and Apr 19, 2024 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Device Dislodged or Dislocated2
- Component Missing1
- Entrapment of Device1
Patient problems
- Insufficient Information2
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 19, 2024Break; Break
ON 06/28/2022, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS THAT AN AR-2922BC BIOCOMPOSITE PUSHLOCK IS BROKEN. THIS WAS DISCOVERED DURING AN UNSPECIFIED TIME.
Patient problems: Insufficient Information
MAUDE report 19147127 ↗MalfunctionOct 19, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON 9/26/2023, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA EMAIL THAT AR-2922BC BIOCOMPOSITE PUSHLOCK AND AN AR-1923BC BIOCOMPOSITE PUSHLOCK ANCHORS PULLED OUT. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2023. ADDITIONAL INFORMATION REQUESTED.
Patient problems: Insufficient Information
MAUDE report 17966823 ↗MalfunctionJan 18, 2022Break
ON 12/27/2021, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA EMAIL THAT AN AR-2922BC BIOCOMPOSITE-CORKSCREW BROKE OFF UPON INSERTION. THIS WAS DISCOVERED DURING AN ARTHROSCOPIC PROCEDURE ON (B)(6) 2021. SURGEON COULD NOT LOCATE THE BROKEN ANCHOR INSIDE THE PATIENT, OUTSIDE IN THE OR FIELD AND COULD NOT DETECT THE ANCHOR VIA X-RAY.
Patient problems: Foreign Body In Patient
MAUDE report 13281741 ↗MalfunctionMar 12, 2021Break; Entrapment of Device
IT WAS REPORTED THAT DURING A LABRAL REPAIR, SIX AR-2922BCS WERE USED. ONE IMPLANTED FINE, THE OTHER 5 BROKE DURING INSERTION. IN ALL 5 ATTEMPTS, THE EYELET AND AN ANCHOR FRAGMENT REMAINS IN THE PATIENT. PATIENT IS 16 YEAR OLD FEMALE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 11475192 ↗InjuryNov 6, 2019Component Missing; Device Dislodged or Dislocated
IT WAS DISCOVERED VIA FACILITY MEDWATCH REPORT ((B)(4)) THAT THE FOLLOWING INCIDENT OCCURRED: WHILE ATTEMPTING TO INSERT THE ANCHOR FOR THE LABRAL REPAIR IT DISLODGED UPON INSERTION AND COULD NOT BE FOUND VISUALLY WITH THE ARTHROSCOPE. ANOTHER ANCHOR WAS OPEN AND UTILIZED CORRECTLY TO COMPLETE THE PROCEDURE. THE INITIAL ANCHOR THAT WAS DISLODGED WAS NOT FOUND AND REMAINED IN THE PATIENT. ARTHREX ANCHOR WAS A BIO-COMPOSITE PUSHLOCK 2.4 X 11.3MM (EXPIRATION DATE 04/30/2021, REFERENCE NUMBER AR-2922BC. SURGERY WAS COMPLETED WITH NO FURTHER COMPLICATIONS, DRESSINGS APPLIED AND PATIENT WAS TAKEN TO PACU. ADDITIONAL INFORMATION OBTAINED FROM THE FACILITY CLINICAL RISK SPECIALIST: THE DATE OF THE PROCEDURE WAS (B)(6) 2019. IT IS POSSIBLE THE DEVICE MAY HAVE BEEN SUCTIONED OUT BY THE NEPTUNE DEVICE HOWEVER THEY WERE UNABLE TO CONFIRM THIS. THE DEVICE COULD STILL BE IN THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9287380 ↗InjurySep 6, 2017Break
IT WAS REPORTED BY A SALES REP THAT DURING A LABRAL REPAIR PROCEDURE, THE 2.4 BIO-COMPOSITE PUSHLOCK WAS INSERTED AND WHILE MALLETTING, THE ANCHOR BROKE. A SECOND ANCHOR WAS OPENED AND AGAIN, THE BIO-COMPOSITE PUSHLOCK BROKE ON INSERTION. THIS HAPPENED THREE MORE TIMES FOR A TOTAL OF 5 DEVICES ALL WITH THE SAME LOT NUMBER (LOT: 10143236). ALL BROKEN PIECES WERE RETRIEVED AND THE ADEQUATE FIXATION WAS FINALLY OBTAINED WITH THE 5 ANCHORS IMPLANTED. FOLLOW UP INVESTIGATION: ALL PIECES OF THE BROKEN IMPLANTS WERE RETRIEVED WITH GRASPERS AND A SHAVER. THE 5 PARTIAL IMPLANTS (OF THE SAME LOT NUMBER) REMAINED IN PATIENT WITH ADEQUATE FIXATION.
Patient problems: No Code Available
MAUDE report 6845921 ↗
Other Fastener, fixation, biodegradable, soft tissue devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.